Michael Carl1, Dina Moazamian2, Alecio F. Lombardi2, Amir Masoud Afsahi2, Yajun Ma2, and Jiang Du2
1GE Healthcare, San Diego, CA, United States, 2UCSD, San Diego, CA, United States
Synopsis
We compared two different T1rho sequence approaches and assessed their performance
for quantitative musculoskeletal imaging. We found that using either a MAPSS or
UTE acquisition after T1rho preparation resulted in similar T1rho decay values
for tissues of moderate to long T2s (e.g., cartilage or meniscus). The UTE
acquisition sequence, however, was also readily able to obtain T1rho decay
values for tissues with short T2s (e.g., tendons), thereby facilitating whole
knee assessment.
Introduction:
T1rho is a
noninvasive biomarker that is sensitive to the dipolar interactions of slow macromolecular
motions within proteoglycan. These macromolecular motions are governed by a
correlation time tc. In
tissues such as cartilage, slow macromolecular tumbling motions have
correlation times on the order of ms, which can be partially refocused by
spin-locking. Previous studies have suggested that T1rho may therefore be correlated
with proteoglycan depletion in osteoarthritis (OA) [1]. Most common T1rho
techniques have relied on either gradient-echo or spin-echo acquisition
sequences that are geared toward tissues with transverse relaxation times (T2s) of several tens of ms
or longer (i.e., “long T2 tissues”). Tissues with short T2s such as tendons
have only recently become accessible to direct magnetic resonance (MR) imaging
through ultrashort echo time (UTE) sequences [2-4]. Combining UTE sequences with T1rho preparation therefore
also allows for the assessment of T1rho value in short T2 tissues. In this
study, we implemented a UTE-T1rho sequence for assessment of the major tissue components
in knee and compared its performance against MAPSS-T1rho.Methods:
The pulse sequences are shown in Figure 1. After a magnetization reset module
(which sets the total magnetization vector to zero) and a fat saturation pulse,
the T1rho preparation uses a 90° excitation pulse followed by a spin-lock pulse
in the orthogonal direction (along the transverse magnetization vector), then a
second 90° pulse to tip the magnetization back to the z-axis. This is then
followed by either a gradient-echo (MAPSS) or UTE multi-spoke imaging sequence
with varying flip angles. The same k-space line is acquired twice with the radiofrequency
(RF) amplitudes shown in red either positive or negative to compensate for T1
recovery during the acquisition train [1]. The resulting T1rho decay parameters
within signal regions of interest (ROIs) can then be fitted to a
mono-exponential decay. We scanned 4 knee joint specimens (donors between 36 and
70 years of age) on a GE 3T MR750. These knee specimens were frozen at -80° F and thawed for ~30 hours prior to scanning. The
pulse sequences shown in Figure 1 were used to scan these specimens at four spin
locks times (TSLs) to allow decay curve fitting (see Figure 3). The acquisition
parameters were as follows: freq matrix=256, field of view (FOV)=15cm, slice thickness=4mm,
number of slices=32, bandwidth=62.5kHz, TE=28µs (UTE), and 2.3ms (MAPSS),
magnetization recovery time=150ms, spin-lock
amplitude=500Hz, 64 spokes-per-prep, average
scan time (for 4 TSLs)=7.5 minutes. The resulting decay curves were then fitted
using least-squares fitting in MATLAB. Results and Discussion:
Representative images
of both MAPSS- and UTE-based T1rho imaging are shown in Figure 2 with different
TSL values. Some of the corresponding decay and fit curves are shown in Figure 3,
as well. Table 1 shows the measured T1rho values for several tissues within the
knee. For most tissues, including the cartilage, and meniscus, there was reasonable
agreement between the MAPSS and UTE imaging acquisitions. However, for short T2
tissues such as the patellar tendon, the UTE acquisition yielded noticeably
lower T1rho values, likely due to the pool selections inherent to the MAPSS and
UTE imaging acquisitions, where only UTE can access appreciable signals from
the short T2 pools that also contribute to lower T1rho values. This can also be
seen in Figure 2 where the patellar
tendon appears as dark signals for MAPSS even for the lowest
TSL values (upper blue arrow), resulting in noisy data points (see Figure 3). In
comparison, the UTE based signals inside the patellar tendon are noticeably stronger (lower
blue arrow in Figure 2) and result in less noisy ROI signals in Figure 3, demonstrating
that UTE-T1rho is a promising tool in assessment of the whole knee.Conclusion
We have examined two T1rho sequences. As expected, we found that using either
a T1rho-prepared MAPSS or UTE imaging protocol yielded similar results for
longer T2 tissues such as cartilage and meniscus. Significant deviations in
T1rho were only observed in short T2 tissues, where UTE showed lower values
than an equivalent MAPSS-based scan. Acknowledgements
No acknowledgement found.References
[1] Li et al, Magn Reson Med 2008 59:298–307
[2] Rahmer et al, Magn Reson Med, 2006. 55(5): p.1075-82
[3] Du et al, Magn Reson Imag 2011:29:470–482
[4] Weiger et al, NMR
Biomed. 2015 28(2):247-54