Nicolas Boulant1, Cécile Lerman1, Lionel Quettier2, Olivier Dubois2, Frédéric Molinié2, Peter Dietz3, and Guy Aubert2
1University Paris-Saclay, CEA, CNRS, BAOBAB, NeuroSpin, Gif sur Yvette, France, 2University Paris-Saclay, CEA, Irfu, Gif sur Yvette, France, 3Siemens Healthcare GmbH, Erlangen, Germany
Synopsis
The whole-body Iseult
11.7T CEA magnet has delivered its first images after nearly 20 years of
research and development. Before reaching this long-waited step, a gradient
coil–magnet interaction test campaign was run for several months at 3T, 7T,
10.2T and 11.7T on the same identical system. It included acoustics,
vibrations, magnet safety system voltage and power deposition in the He bath
measurements. Some vibration results versus field strength are presented here.
Vibration amplitudes are shown to increase less than with B0 field
strength.
Introduction
MRI at UHF offers great promises in SNR and CNR.
In the quest of reaching ever higher magnetic fields, one crucial aspect
beforehand remains the design of the magnet, the gradient coil and their
interaction. The time-varying fields generated by the gradient cause
eddy-currents in the structure which, under the influence of a strong B0
field, generate vibrations, which engender themselves electric fields and power
deposition in the He bath. In addition to these power losses which can be
critical for the integrity of the magnet, mechanical resonances also can be
detrimental for imaging performance (e.g. ghosting). In this context, we report
in this work vibration measurements on the SC72 Siemens Healthineers whole-body
gradient in the bore of the Iseult magnet at 3T, 7T, 10.2T and 11.7T. Methods
The Iseult magnet1 is 5 m long/wide
and weighs 132 tons. The main superconducting coil is immerged into a He bath
of 7000 liters cooled down to 1.8 K, where He is superfluid. The gradient coil
is surrounded by a lead tube to screen gradient-magnet interactions at some
frequencies2. Six mono-axial accelerometers (Hottinger Bruel &
Kjaer, Naerum, Denmark) were glued on the gradient, the lead tube and the
cryostat to measure accelerations along the X and Y directions (Fig. 1). Vibrations
were measured by applying a 1-min chirp excitation on the 3 gradient axis
separately with G=1 mT/m over the 0-3200 Hz range. Close to a well-defined natural
frequency ω0, the response of the
accelerometers on the gradient was temptingly modeled with a damped harmonic
oscillator: $$$\ddot{u}+\eta \dot{u} + \omega_0^2u=F(t)/m$$$, where m is an effective mass and where the
damping coefficient η shall
vary with B0. Because of the Lorentz force, the force should be
proportional to the current (thus gradient strength) and the static field, i.e. $$$F \propto I B_0$$$. For an oscillating force at angular frequency ω, assuming
a large quality factor, solving the differential equation in the steady state
yields approximately for the peak displacement $$$u_{max} \propto IB_0/\eta$$$. To verify this simple model, measurements were performed
at 0.1, 1 and 5 mT/m to confirm the linearity of the response versus I, for B0
constant. They were finally repeated on the same identical system at 3T,
7T, 10.2T and 11.7T to study the response and the dependence of η on B0
over a broad frequency range. Results
Fig.2.a. shows the measured
temporal displacement along the Y axis for a sine excitation of GY=20 mT/m at
550 Hz and at 7T, with its Fourier transform. Fig.2.b shows acceleration spectra
acquired at 11.7T for the three gradient axis and normalized by the different gradient
strengths. The plots being nearly identical and the system responding at the
frequency of the excitation indicate a behavior consistent with a harmonic
oscillator. Fig. 3 shows the acceleration spectra under a G=1 mT/m excitation
at different B0 field strengths and for the three gradient axis,
showing qualitatively the same behavior aside from expected increased
amplitudes versus field strength. The amplitudes of the peaks, normalized to the
ones at 3T, are shown in Fig. 4 and reveal a response that grows slower than the
B0 field. Discussion and conclusion
To a good
approximation, the experimental results show that the vibration response at one
location of the gradient coil is consistent with a damped harmonic oscillator.
The data reveal a less severe dependence for the response than with B0,
which could explain moderate increases of acoustic noise versus field
strength3 for a same set up. Having $$$\eta \propto B_0$$$ would
yield a nearly constant amplitude response versus B0, which is not
the case. Interestingly, some resonances also exhibit a plateau which could be
consistent with a dependence of the damping coefficient as $$$\eta(\omega) \propto \eta_0(\omega) + \alpha B_0$$$ (with α being a scalar). The
data, at least up to 11.7T, on the other hand cannot verify a dependence as $$$\eta(\omega) \propto \eta_0(\omega) + \alpha B_0^2$$$ as suggested in the literature4 which would imply an increase of the
response until a maximum is reached, followed by a drop versus field strength. Acknowledgements
NB is funded by the
European Union’s Horizon 2020 research and innovation program under grant
agreement No 885876 (AROMA project). We thank Hermann Landes for valuable
discussions. François Nunio and Loris Scola are also thanked for the fruitful
discussions about the mechanics during the tests.References
[1] D. Le Bihan, T.
Schild, Human Brain MRI at 500MHz, scientific perspectives and technological
challenges, Supercond. Sci. Techno. 2017;30:1-19.
[2] G. Aubert. NMR
imaging system with reduced cryogenic losses and reduced acoustic noise. US
patent no 8,410,777 B2.
[3] G. Metzger.
Future Directions in UHF MR Safety from the Perspective of 10.5T. Proceedings
of MR safety workshop of the ISMRM, 2019, Utrecht, The Netherlands.
[4] S. Winkler et al.
On the Accurate Analysis of
Vibroacoustics in Head Insert Gradient Coils. MRM 78:1635-1645
(2017).