Mathias Davids1,2,3, Livia Vendramini1, Natalie Ferris4,5, Valerie Klein1,3, Bastien Guerin1,2, and Lawrence L. Wald1,2,5
1Martinos Center for Biomedical Imaging, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Heidelberg, Germany, 4Harvard Graduate Program in Biophysics, Harvard University, Cambridge, MA, United States, 5Harvard-MIT Division of Health Sciences and Technology, Boston, MA, United States
Synopsis
We describe the process of designing and analyzing two body gradient
coils with and without PNS optimization suitable for prototype construction and
experimental validation. The optimized coil achieves a 51% increase in PNS
thresholds at a 15% inductance penalty. Both coils are construction-ready
(single continuous wire path) and have realistic and matched design
characteristics (actively shielded, torque/force balanced, high field linearity
in 40 cm ROL). We are in the process of constructing coil prototypes, with the
ultimate goal of experimentally validating their PNS differences.
Target audience
MRI gradient
designers, safety researchers and imaging scientistsPurpose
Peripheral Nerve Stimulation (PNS) limits the usable image encoding
performance in state-of-the-art body and head gradient coils [1-3]. We recently
developed an approach to model and incorporate PNS metrics in the numeric coil
winding optimization to generate PNS optimized coils [4]. This optimization can
increase PNS thresholds by up to 90% while satisfying traditional engineering
constraints (linearity, FOV, shielding, torque/force balancing). In this work,
we compare two body gradient coils (PNS constrained and unconstrained)
designed for construction and experimental validation.Methods
Choice of coil
designs:
We design two
optimized Y-axis gradient coils (YG1 and YG2) using the framework recently presented
[4]: YG2 is designed with a PNS constraint and 15% inductance penalty while YG1
is a conventional design without PNS optimization. The designs have otherwise
identical constraints, and the windings are construction-ready with the goal of
ultimately measuring and comparing PNS thresholds. Both coils are smaller
diameter (lower inductance) than typical designs to ensure stimulation can be
achieved in each case.
The two coil designs
were chosen by generating a family of L-curves analyzing the tradeoff between PNS
oracle (reciprocal threshold) and coil inductance for a range of different coil
characteristics (varying coil lengths, inner diameters, sizes of linear
volumes, and field linearities). Other engineering constraints were kept
constant: active shielding (≤ 0.5 μT/A on the cryostat
surface), sensitivity (0.08 mT/m/A), linearity (≤ 5% deviation), FOV (40 cm diameter), force and torque
balancing (≤ 0.4 N/A and ≤ 0.3 Nm/A) and wire spacing (≥ 8mm). The PNS
optimization and analysis followed the previously described electromagnetic and
neurodynamic modeling using a male and female body model [5-7]. All coils are
designed for a polygonal rather than cylindrical coil geometry to simplify prototype construction.
Generation of
buildable wire pattern:
Conversion of the
stream function solution into a single continuous buildable wire path is
considerably more complex for PNS optimized coils due to the twist in the
winding pattern. We developed an algorithm to automatically generate buildable
wire paths from the individual closed loops (Fig. 3). The windings are designed
to be milled into ABS sheets with a CNC router, populated with AWG 8 wire,
assembled to form the 16-sided polygonal coil former and covered with
fiberglass/epoxy. Note that these coils are meant for PNS threshold
measurements outside the magnet at low duty-cycle rather than for imaging.
Heating test:
We anticipate a heat generation of approx. 1.5
kW for typical trapezoidal waveforms used in threshold experiments. Cooling will be performed by wrapping the coil in water cooling blankets. We tested the
feasibility of this approach on a smaller test coil (one fourth of the
unoptimized coil YG1 primary). The coil was driven with a continuous 1 kHz
sinusoid at 80A, which corresponds to 50 ms pulses at ~700A every 5 seconds
(mimicking the stimulation experiments). The cooling blankets were placed on
the wire side (i.e., at minimum distance to the wire), while we filmed the
patient-facing side with an infra-red camera.Results
Figure 1 shows L-curve trade-offs between
PNS oracle (reciprocal threshold) and inductance. The shaded area of each
L-curve denotes the span of the female/male PNS oracle, while the solid line
(with markers) corresponds to the average of the two models. We found the
average to be a good proxy for the population averages in previous studies [5,6].
The dashed gray shaded area corresponds to the performance space inaccessible
due to the amplifier’s current/voltage limits (900A, 2200V). The different
panels correspond to variations of a single coil design parameter. Coil
solutions YG1 and YG2 achieve good balance between the four design parameters
with a 51% increase in PNS thresholds (35% oracle reduction). Note that the PNS
thresholds of the two coils and both models lie within the operational region
which is critical for experimental validation.
Figure 2 shows the PNS threshold curves of
YG1 (black) and YG2 (red) for both model sexes (and their average). The
gray dashed area corresponds to the performance space inaccessible to the
amplifier. The blue dashed area corresponds to the performance gain resulting from
51% greater PNS thresholds of coil YG2; the yellow area shows the lost performance
from the 15% inductance increase. Figure 2 also shows the final winding
patterns and predicted nerve activation maps for YG1 and YG2.
Figure 3 illustrates the workflow of
converting the raw winding pattern (iso-contours of the stream function) into a
single buildable coil pattern. Figure 4 shows the experimental setup (CAD design rendering).
Figure 5A/B show the test coil used in the
heating experiment and an infra-red image of Ohmic heating. Figure 5C shows the
temperature over time with and without cooling which reduced the max.
temperature from 72 to 46 °C.Conclusion
PNS
constrained coil design enables studying trade-offs between nerve thresholds and
conventional engineering metrics. This allowed us to identify two optimized designs
YG1 and YG2, with the latter increasing PNS thresholds by 51% at a 15%
inductance penalty. The coils were designed such that the PNS thresholds of
both coils lie well within the operational parameter space to allow for
experimental validation. Construction and experimental validation of these prototypes is on-going.Acknowledgements
The authors would like to acknowledge the help of past and present members of the gradient coil group at Siemens Healthineers, including Peter Dietz, Gudrun Ruyters, Axel vom Endt, Ralph Kimmlingen, Franz Hebrank, and Eva Eberlein. Research reported in this publication
was supported by the National Institute of Biomedical Imaging and
Bioengineering, and the National Institute for Mental Health of the National
Institutes of Health under award numbers R24MH106053, U01EB026996, U01EB025162,
U01EB025121, R01EB028250. The content is solely the responsibility of the
authors and does not necessarily represent the official views of the National
Institutes of Health.References
[1] Setsompop
et al., “Pushing the limits of in vivo diffusion MRI for the human connectome
project”, NeuroImage, vol. 80, pp. 220 – 233, 2013
[2] McNab et
al., “The human connectome project and beyond: Initial applications of 300 mT/m
gradients”, NeuroImage, vol. 80, pp. 234 – 245, 2013
[3] Tan et al.,
“Peripheral nerve stimulation limits of a high amplitude and slew rate magnetic
field gradient coil for neuroimaging”, Magnetic Resonance
in Medicine, vol. 83, no. 1, pp.
352–366, 2020.
[4] Davids et
al., “Optimization of MRI Gradient Coils with Explicit Peripheral Nerve
Stimulation Constraints”, IEEE Transactions on Medical Imaging, 2020, 40, 129-142
[5] Davids et al. “Predicting
magnetostimulation thresholds in the peripheral nervous system using realistic
body models", Scientific Reports, vol. 7,
no. 1, p. 5316, 2017.
[6] Davids et al. "Prediction of peripheral
nerve stimulation thresholds of MRI gradient coils using coupled
electromagnetic and neurodynamic simulations", Magnetic resonance in
medicine 81.1 (2019): 686-701.
[7] Davids et al., “A Huygens’ surface approach
to rapid characterization of Peripheral Nerve Stimulation”, Magnetic Resonance
in Medicine, 2021, in press