Mareike Alicja Buck1,2, Simon Konstandin1,3, Nora-Josefin Breutigam1, and Matthias Günther1,2,3
1Fraunhofer MEVIS, Bremen, Germany, 2University Bremen, Bremen, Germany, 3mediri GmbH, Heidelberg, Germany
Synopsis
In this
work, the application of velocity-selective inversion ASL in healthy breast
tissue is demonstrated for the first time. The measurements show a low measured
signal in breast tissue. However, the prominent fat artifacts that appear in
the images will need to be addressed in future work. The technique presented
here has nonetheless been shown to hold promise in the field of non-invasive
breast cancer diagnostics with further development, particularly the inclusion
of background suppression to get clear perfusion signal.
Introduction
Breast
Cancer is the most common form of cancer among women. Breast MRI has been shown
to be the best tool for detecting breast cancer [1]. Generally, a dynamic
contrast-enhanced (DCE) MRI scan with injection of contrast agent is used. A
completely non-invasive alternative to this is Arterial Spin Labeling (ASL), which
uses the blood itself as an endogenous contrast agent. The most commonly used
techniques are pulsed ASL (PASL) and pseudo-continuous ASL (pCASL) [2]. In both
cases, the labeling is performed outside the imaging region. For imaging, the
so-called arterial transit time (ATT) must be waited until the labeled blood reaches
the spatially distant imaging plane. An alternative to this is
velocity-selective ASL, which labels blood above a certain velocity (called
cut-off velocity vc) within the imaging region [3] and is thus independent
of the feeding arteries and ATT. Especially for organs like breast tissue, without
a clear and straight feeding artery, this method is of interest.
In this
work, velocity-selective inversion ASL (VSI-ASL) was used for the first time in
healthy breast tissue to investigate whether this method is also suitable for
perfusion imaging, especially for the eventual detection of breast tumors, where
higher perfusion is expected in the tumor region compared to healthy tissue.Methods
Image acquisition
of breast tissue of three female volunteers (30-31y) using the VSI-ASL sequence
described by Qin et al. [4] was performed on a 3T MR system (MAGNETOM Skyra,
Siemens Healthineers, Erlangen, Germany) implemented in the vendor independent
sequence development framework gammaSTAR [5, 6] with following parameters: TVS=44.3 ms,
Tstep=5.52 ms, Tref=0.84 ms, TG=0.7 ms, Tramp=0.35 ms,
Gmax=27.5 mT/m, Tgap=0.3 ms, Δ=2.5 ms, vc=2.5 cm/s
with velocity-encoding in foot-head direction (sequence diagram in Figure 1).The
post labeling delay (PLD) was 1000 ms. Additionally, a spoiling gradient was
used after the pulse train and PLD to disrupt the remaining transverse
magnetization. Only one VSASL module was used to measure just the signal above
vc. One transverse slice along the isocenter was acquired with 8 mm
thickness and 2.0x2.0 mm² resolution, FoV=256x256 mm², TE=25 ms, TR=2000 ms. An EPI
readout [7] was used with anterior-posterior phase encoding and additional fat
saturation, but without any further fat or background suppression. For
averaging, 20 measurements were taken including three control-label pairs (120
ASL measurements in total). This resulted in a total scan time of 3:59 mins.
For better visualization of the perfusion-weighted images, contours of the
breast were created with MeVisLab (MeVis Medical Solutions, Bremen, Germany) using
the localizer image of every volunteer (Figure 2). Results
Figure 3 shows
the subtracted images of the averaged control and label images for every
volunteer. It can be seen that the images are mainly influenced by fat artifacts,
resulting in a superposition of the measured signal in the breast in coronal
direction. This is indicated illustratively by red arrows. Additionally, the blood
signal inside the heart is visible (green arrow). Furthermore, motion artifacts
of the blood inside the heart are visible (yellow arrow). In Figure 4, the
created contours of the breast are overlaid using the world coordinates in the
perfusion-weighted images to get a better focus on the organ of interest. Here,
a low signal can be seen within the breast tissue in all cases.Discussion and Conclusion
This work
shows the first application of velocity-selective inversion ASL in the breast. The
perfusion-weighted images show a low signal in the breast, which could be
interpreted as perfusion. However, the strong fat artifacts make a clear
assignment difficult. Therefore, for the next step, it is important to include
background suppression to reduce the strong influence of fat signal in the
images to acquire the pure perfusion signal. In addition, a smaller FoV can be selected
to drop the cardiac signal with phase encoding in right-left direction to
minimize fold ins.
Despite the
visible artifacts in these initial proof-of-principle measurements, this work demonstrates
that VSI-ASL imaging could be a viable technique for perfusion measurements in
the breast. Hence, VSI-ASL imaging could be an important step in bringing non-invasive
breast tumor detection to the clinic, with all the advantages that contrast agent
free measurement brings to the examination workflow and patients.Acknowledgements
No acknowledgement found.References
[1] Mann et al. Contrast-enhanced MRI for
breast cancer screening. J Magn Reson Imaging 2019; 50: 377-390
[2] Alsop et al. Recommended Implementation of
Arterial Spin Labeled Perfusion MRI for Clinical Applications: A consensus of
the ISMRM Perfusion Study Group and the European Consortium for ASL in Dementia,
Magn Reson Med. 2015; 73(1): 102–116.
[3] Wong et al. Velocity selective arterial
spin labeling. Magn Reson Med 2006; 55(6): 1334-1341.
[4] Qin et al. Velocity-Selective-Inversion
Prepared Arterial Spin Labeling, Magn Reson Med. 2016; 76(4): 1136–1148.
[5] Cordes et al. Portable and
platform-independent MR pulse sequence programs, Magn Reson Med. 2020; 83(4): 1277-1290
[6] https://gamma-star.mevis.fraunhofer.de/
[7] Mansfield. Multi-planar image formation
using NMR spin echoes, J Phys C 1977; 10(3): 55-58