Iman Khodarahmi1, Inge M Brinkmann2, Mary Bruno1, Jerzy Walczyk1, Ryan Brown1, and Jan Fritz1
1Department of Radiology, New York University School of Medicine, New York, NY, United States, 2Siemens Medical Solutions USA Inc, Malvern, PA, United States
Synopsis
New generation low-field MRI systems can be ideal for
MRI-guided interventions due to wide bore diameters and lower susceptibility
artifacts improving patient access and needle visualization, respectively. We
evaluated a set of pulse sequences for their suitability for needle
visualization using a new-generation 0.55T system and compared the results with
those of a clinical 3T system. Interventional needles can be adequately
visualized at 0.55T. Using TSE pulse sequences, needle artifacts display more
favorable at 0.55T than 3T. SEMAC and HASTE pulse sequences may not facilitate
diagnostic needle visualization at 0.55T but are not needed for clinical MR
interventions.
Introduction
The new generation of low-field MRI systems is equipped with
the state-of-the-art hardware and software developed for high-field systems and
is projected to provide broader world-wide access due to its lower cost. Such
systems can be an ideal platform for MRI-guided procedures and interventions
for two reasons: First, the lower magnetic field, gradient strength and slew
rate have enabled the development of wide bore diameters of 80 cm that provide
space for interventional apparatus. Second, lower susceptibility artifacts at
low field strengths are expected to improve the visualization of the metallic
devices used in interventional procedures (1,2). In this study, we aimed to propose
a set of pulse sequence parameters suitable for needle visualization at a new generation
0.55T system and compare the results with those of a 3T system used in clinical
practice.Methods
A commercially available MR-conditional 20-gauge
cobalt-chromium injection needle (Cook Medical, Bloomington, IN, USA) was
placed in the central slot of a phantom designed to assess the metal artifacts
(Figure 1). The phantom was made of eight rows of six parallel plastic strips with
different thicknesses and spacings, and was filled with coconut oil. Similar to
human interventions, this setup was placed in the axial plane at the iso-center
of a commercial MRI system (1.5T MAGNETOM Aera; Siemens Healthcare GmbH,
Erlangen, Germany) modified to operate as a prototype system at 0.55T field
strength with a maximum gradient strength
of 25 mT/m and a maximum slew rate of 40 T/m/s. An 11 cm loop coil and an 18-channel spine
array tuned to operate at 0.55T were used for signal reception.
Various pulse sequences, each with different parameters,
including turbo spin-echo (TSE), true fast imaging with steady-state free precession
(TRUFI), volumetric interpolated breath-hold examination gradient-echo (VIBE), slice
encoding for metal artifact correction (SEMAC), and half-Fourier acquisition
single-shot TSE (HASTE) were evaluated for optimal needle visualization. The
extent of the metal artifacts and overall image quality was qualitatively assessed
by two readers by noting acquisitions that provided clear visualization of the
strips adjacent to the needle.
For comparison with the standard 3T field strength, the same
experimental setup was imaged with a clinical system (MAGNETOM Skyra, Siemens
Healthcare) using the clinical pulse sequences and signal reception by 3-channel
surface and 32-channel spine arrays.Results and Discussion
For the TSE sequence, the effect of various signal averages,
parallel imaging acceleration factors (GRAPPA), receiver bandwidths, and turbo
factors were assessed (Figure 2) with the optimized parameters being a signal
average of 3, no parallel imaging, a receiver bandwidth of 300 Hz/pixel, and no
turbo factor dependence. Findings of other pulse sequences are summarized in
Table 1.
Figure 3 compares the images acquired at 0.55T using the
optimized parameters and 3T using the clinical protocols. For the TSE pulse
sequence, metal artifacts were substantially smaller at 0.55T than 3.0T,
although there is a time-penalty cost to reach diagnostic signal-to-noise ratios.
TRUFI images, considered not diagnostic at 3T, can be rapidly acquired at 0.55T
and used for fluid visualization during MRI-guided injections. Metal artifacts
at 0.55T are smaller with the VIBE pulse sequence with comparable acquisition
times. SEMAC is not needed at 0.55T because the through-plane artifacts
associated with the needle in the axial plane are minimal and images are degraded
by the blurring that is caused by view-angle-tilting. Within the range of
tested parameters, HASTE images were considered non-diagnostic for needle
visualization at 0.55T.Conclusion
Interventional needles can be adequately visualized at
0.55T, suggesting that the performance of MRI-guided procedures is feasible at
the new generation of low-field systems. The TSE pulse sequence, with a longer
acquisition time, can result in smaller metal artifacts at 0.55T than 3T. Within
the range of tested parameters, SEMAC and HASTE pulse sequences were non-diagnostic,
but also not needed for needle visualization at 0.55T, whereas images acquired
with TRUFI can be used for rapid tracking of the injected fluid.Acknowledgements
The authors would like to acknowledge the assistance of
Siemens Healthcare in the modification of the MRI system for operation at 0.55T
under an existing research agreement between NYU and Siemens Healthcare.References
- I. Khodarahmi, I.M.
Brinkmann, D. Lin, M. Bruno, P. Johnson, F. Knoll, M.B. Keerthivasan, H.
Chandarana , J. Fritz. Advanced Low-Field MRI of Hip Arthroplasty Implants:
First Experience at 0.55 T. International Society for Magnetic Resonance in
Medicine (ISMRM). May 2021.
- Campbell-Washburn AE,
Ramasawmy R, Restivo MC, et al. Opportunities in Interventional and Diagnostic
Imaging by Using High-Performance Low-Field-Strength MRI. Radiology.
2019;293(2):384-393.