Faraz Amzajerdian1, Hooman Hamedani1, Ryan Baron1, Yi Xin1, Tahmina Achekzai1, Luis Loza1, Mostafa Ismail1, Ian Duncan1, Stephen Kadlecek1, Kai Ruppert1, and Rahim Rizi1
1University of Pennsylvania, Philadelphia, PA, United States
Synopsis
Hyperpolarized
xenon-129 (HXe) imaging is capable of quantifying lung function through
measurements of ventilation and gas exchange. However, traditional HXe approaches
rely on long breath-holds for imaging, which may not be representative of
steady-state behavior and are limited by the volume of gas that can be
delivered. In this work, we imaged gas- and dissolved-phase xenon continuously
over approximately 50 breaths and retrospectively binned the images, generating
dynamic maps encompassing the entire respiratory cycle.
Introduction
In order to
quantify gas exchange and assess regional changes in lung function, hyperpolarized
xenon-129 (HXe) MRI takes advantage of the two distinct chemical shifts of
xenon dissolved in the tissue/plasma (TP) and red bloods cells (RBC),
respectively. Traditional HXe imaging approaches have typically involved long
breath-holds at end-inspiration/expiration, producing static snapshots which
may not fully characterize lung behavior and may be insensitive to the heterogeneous
replacement of air with xenon1. In order to fully quantify the
dynamics of ventilation and gas exchange, information from all phases of the
breathing cycle is required. Due to the rapid depolarization of HXe in response
to repeated RF excitations, acquiring this information during one breathing
cycle is impractical. However, imaging over the course of several breaths with
smaller doses of HXe continuously replenishes available polarization while
maintaining sufficient signal-to-noise (SNR) in each individual image. This
approach introduces additional SNR benefits for dissolved xenon in particular,
as the total xenon dose is no longer limited by safety/anesthetic concerns or
the patient’s tidal volume. Here, we explored the feasibility of continuously
acquiring gas- (GP) and dissolved-phase (DP) xenon images over several minutes
while the subject freely breathed a low-dose xenon mixture. The resulting
images were retrospectively binned using diaphragm position to produce high-resolution
maps of ventilation and gas exchange at distinct respiratory stages. Methods
One healthy
subject was imaged using an Institutional Review Board (IRB)-approved protocol
in both prone and supine positions in a 1.5T scanner (Magnetom Avanto, Siemens)
using an 8-channel 129Xe coil (Stark Contrast, Germany). A prototype
commercial system (XeBox-E10, Xemed LLC, NH) was used to polarize 87% enriched
xenon-129.
Gas delivery,
automated via a custom device, was administered through a nasal canula. The
subject received a 50 mL dose of HXe with every inhalation while freely
breathing room air. Imaging was performed continuously via a 3D stack-of-spirals
sequence over the course of approximately four minutes, delivering 2.5 L of HXe
in total. To quantify both ventilation and gas exchange, the frequencies of
each RF excitation pulse were alternated between the GP (0 ppm) and RBC (218
ppm) resonances. The TE was chosen such that, for each RBC excitation, the
phase of the RBC and TP resonances differed by precisely 90° at the center of
k-space, allowing separation of these signals via the 1-point Dixon method2.
The global RBC/TP ratio was determined from an acquired spectrum and used for
phase correction to align the RBC and TP signals into real and imaginary
compartments, respectively. The GP phase maps were subsequently used to correct
B0 inhomogeneities. Other imaging parameters included: GP/RBC flip
angles of 3°/30°, TR/TE = 8.13/0.81 ms, matrix size = 80x80x5, and FOV of
350x350x175 mm3. Approximately 500 GP and 500 DP images
(undersampled by factor of 2) were acquired. From the GP images, the position
of the diaphragm was calculated from the signal of the lower right lung, as
indicated in Figure 1A. An exponential curve was fit to the sum of signals
along the y-axis in this region, with the diaphragm position assumed to be at
20% of the maximum signal (Figure 1B). The diaphragm position over time (Figure
1C) was used to bin the GP and DP images into 8 distinct phases of the
breathing cycle, after which the GP, RBC/GP, TP/GP, and RBC/TP maps were
generated. Results and Discussion
Figure 2
shows the retrospectively binned ratio maps at end-inspiration for both prone
and supine positions in a healthy subject. The anterior-to-posterior gravitational
gradient, consistent with supine HXe imaging, is clearly apparent in the
distributions of the TP/GP, RBC/GP, and RBC/TP maps, in which the greater
pleural pressure in the posterior lung increases tissue density, resulting in
increased uptake and exchange. This gradient is partially reversed in the prone
position due to reduced compression of the posterior lung. Figure 3 shows the
middle slice across all 8 phases, illustrating changes in gas uptake and
exchange over the breathing cycle. There are notable inaccuracies in the
calculated ratio maps, however, as the phase corrections were performed based
on spectra acquired during an end-inhale breath-hold, so the resulting RBC/TP
ratios will not be particularly accurate for other stages of respiration. This source
of error can be ameliorated in the future by interspersing spectra acquisitions
throughout the imaging protocol. Additionally, the large number of acquired
images creates an opportunity for further improvements in both SNR and
resolution via the implementation of undersampling and keyhole techniques, with
the latter showing particular promise due to the dependence of the 1-point
Dixon method on the center of k-space. Conclusion
Combining retrospective gating
with continuous, low-dose HXe imaging can quantify ventilation and gas exchange
dynamics throughout the respiratory cycle, providing a more comprehensive evaluation
of lung function during natural breathing.Acknowledgements
No acknowledgement found.References
[1] Hamedani,
Hooman, et al (2021). Ventilation heterogeneity imaged by multibreath wash-ins
of hyperpolarized 3He and 129Xe in healthy rabbits.
The Journal of Physiology 599(17), 4197–4223.
[2] Kaushik,
Sivaram S., et al (2016). Single-breath clinical imaging of hyperpolarized
129xe in the airspaces, barrier, and red blood cells using an interleaved 3D radial
1-point Dixon acquisition. MRM 75(4), 1434–43.