Tal Finkelman1, Edna Furman-Haran2, Rony Paz3, and Assaf Tal1
1Chemical and Biological Physics, Weizmann, Rehovot, Israel, 2Life Sciences Core Facilities, Weizmann, Rehovot, Israel, 3Brain Sciences, Weizmann, Rehovot, Israel
Synopsis
Glutamate
(Glu) and γ‑aminobutyric-acid (GABA) are central
neurotransmitters participating in many cognitive processes, and their combined
measurement is of great relevance. Proton Magnetic Resonance Spectroscopy (1H-MRS)
allows the no–invasive, in-vivo, measurement of GABA and Glu. At magnetic fields
below 7T, GABA is detected by spectral editing strategies, while Glu is detected
by non-editing strategies. We compared both edited and non‑edited approaches at
7T for simultaneously quantifying Glu and GABA from 21 volunteers, and found
that non-edited sequences at TE=80ms provide better combined-reproducibility
than either edited sequences at the same TE, or non-edited sequence at a
shorter TE (42ms).
Background
The
ratio of glutamate (Glu) to γ‑aminobutyric-acid
(GABA), also known as the excitatory-inhibitory balance (E/I balance),
modulates a wide range of cognitive and behavioral processes1.Proton
Magnetic Resonance Spectroscopy (1H-MRS) remains the only method
capable of safely and non-invasively measuring the concentrations of Glu and
GABA in‑vivo. At 3T, MRS relies on spectral Mescher-Garwood (MEGA)2 editing techniques to distinguish GABA from its
overlapping resonances. However, at ultrahigh field strengths (≥7T), and the
increases in spectral resolution they afford, non-edited spectroscopic
techniques become potential viable alternatives to MEGA based approaches, and
address some of their shortcomings. We present a comprehensive comparison of
edited and non‑edited strategies at 7T for simultaneously quantifying Glu and
GABA, and evaluate their reproducibility and relative bias. Methods
21 volunteers were scanned on a 7T scanner,
using three single‑voxel MRS sequences (Fig. 1): SemiLASER (TE=42 ms), and
symmetric and asymmetric MEGA‑SemiLASER (TE=80 ms). Once completed, each
volunteer was removed from the scanner, repositioned, and rescanned for
reproducibility estimation.
We compared eight strategies (Fig. 1C) for simultaneously
quantifying both GABA and Glu at 7T: LCModel‑fitted non edited sLASER at TE=42
ms without (sLASER) and with the inclusion of a measured macromolecule (MM)
baseline3,4 (sLASER‑MM). LCModel‑fitted
MEGA‑based sequences, using only the OFF data (Sym‑msLASER‑OFF; Asym‑msLASER‑OFF)
(Fig. 2A-B), and double Gaussian fitted MEGA‑based sequences, which quantify
only GABA, using the difference spectra (Sym‑msLASER-DIFF; Asym‑msLASER-DIFF)
(Fig. 2C-D). To enable a fair comparison with the same signal to noise (SNR)
between the sLASER and msLASER‑OFF datasets, the last two strategies repeated
the spectral LCModel fitting for the sLASER acquisitions, with half the number
of excitations (sLASER‑32, sLASER‑32‑MM). Metabolites concentrations were
corrected for tissue fractions within the voxel and relaxation times. The
intra-subject CV for each metabolite was calculated using the first and second
scan of each volunteer, and averaging across volunteers. The combined CV of
GABA and Glu was calculated as CVE/I=√(CVGABA2+CVGlu2).Results
The first and second scans were unbiased (Fig. 3), meaning
that the magnet, sequences and volunteers all remained stable between the two
scans. Fig. 4 present GABA and Glu concentrations, CRLBs and CVs in the
different quantification strategies. The best intra‑subject CV for GABA was
observed for the Asym-msLASER‑OFF data. The reproducibility of Glu was better
under sLASER, but not substantially. The msLASER‑OFF data exhibit the best CVE/I:
13.3% (Asym‑msLASER‑OFF) and 13.8% (Sym‑msLASER‑OFF), compared to much higher
CVs for sLASER: 24.9% (sLASER), 29.5% (sLASER‑MM), 25.3% (sLASER‑32), 32.6%
(sLASER‑32‑MM), 20.1% (Asym‑msLASER‑DIFF), and 21.0% (Sym‑msLASER‑DIFF).
Inclusion of a MM baseline in sLASER slightly reduced the
quantified concentration of Glu (11.5 mM vs. 11.0 mM), and increased GABA
estimates from 0.64 mM to 0.92 mM. LCModel fitting of the Sym‑msLASER‑OFF and
Asym‑msLASER‑OFF data resulted in similar estimates of Glu (11.1 mM and 11.2 mM),
but a higher estimate of GABA: 1.30 mM and 1.35 mM. GABA quantification based on the edited
spectra resulted in 1.13 mM for the Asym‑msLASER‑DIFF, and 1.06 mM for the Ssym‑msLASER‑DIFF.
Significant change (P<0.01) was found between the symmetric and asymmetric
approaches in the OFF based quantification, and between the OFF and DIFF approaches,
yet non-significant change was found between the symmetric and asymmetric approaches
in the DIFF based quantification.Conclusions
The MM
elimination in the edited strategy did not affect the quantification, likely due to the narrow bandwidth of the spectral
editing pulses (Fig. 1B). Non-edited sequences at TE=80 ms provide better reproducibility
than either edited sequences at the same TE, or non-edited sequence at a
shorter TE (42 ms); This is due to the judicious choice of echo time leads to a
pseudo-singlet appearance for the GABA multiplets at 1.89 and 2.29 ppm
(Fig. 5). Our results motivate a transition to non-edited MRS of GABA and
glutamate at ultrahigh fields, provided the echo time is set appropriately. Acknowledgements
Assaf Tal acknowledges the
support of the Monroy-Marks Career Development Fund, the historic generosity of
the Harold Perlman Family, and the support of NIH grant R21NS112853 and Israeli
Science Foundation personal grant 416/20. Dr. E. Furman-Haran holds the
Calin and Elaine Rovinescu Research Fellow Chair for Brain Research. We
would like to acknowledge Edward J. Auerbach, Ph.D. and Małgorzata Marjańska,
Ph.D. (Center for Magnetic Resonance Research and Department of Radiology,
University of Minnesota, USA) for the development of the pulse sequences for
the Siemens platform which were provided by the University of Minnesota under a
C2P agreement. References
1. Bezalel, V.,
Paz, R. & Tal, A. Inhibitory and excitatory mechanisms in the human
cingulate-cortex support reinforcement learning: A functional Proton Magnetic
Resonance Spectroscopy study. Neuroimage 184, 25–35 (2019).
2. Mescher, M., Merkle, H., Kirsch, J.,
Garwood, M. & Gruetter, R. Simultaneous in vivo spectral editing and water
suppression. NMR Biomed. 11, 266–272 (1998).
3. Snoussi, K. et al. Comparison
of brain gray and white matter macromolecule resonances at 3 and 7 Tesla. Magn.
Reson. Med. 74, 607–613 (2015).
4. Schaller, B., Xin, L. & Gruetter,
R. Is the macromolecule signal tissue-specific in healthy human brain? A 1
H MRS study at 7 tesla in the occipital lobe. Magn. Reson. Med. 72,
934–940 (2014).