Guodong Weng1,2, Piotr Radojewski1,2, Philippe Schucht3, Roland Wiest1,2, and Johannes Slotboom1,2
1Institute for Diagnostic and Interventional Neuroradiology, Support Center for Advanced Neuroimaging (SCAN), INSEL Gruppe AG, University of Bern, Bern, Switzerland, 2Translational Imaging Center, sitem-insel AG, Bern, Switzerland, 3Department of Neurosurgery, Inselspital Bern and University Hospital, Bern, Switzerland
Synopsis
Purpose: to detect phosphoethanolamine (PEtn or PE) in vivo for whole-brain MRSI, overcoming challenges of B1+-inhomogeneities and CSDE at 7T.
Methods: An EPSI-based B1+ robust SLOW-editing was applied to three healthy subjects. TE = 90 ms, TR = 1500 ms, and TA = 9:04 mins.
Results: the editing PE signal at 3.22 ppm was clearly present, and the B1+-homogenous result/map was obtained.
Conclusions: whole-brain in vivo PE-editing can be performed
in around 9 minutes measurement time and is therefore clinically
applicable.
INTRODUCTION
Increased phosphoethanolamine (PEtn
or PE) is frequently observed in 31P-MRS studies of various cancers
in humans. The role of PE in cancer cell survival and the molecular causes of
this increase are, however, relatively underexplored1. Recently Osawa et al 2 presented evidence that glutamine
deprivation leads to phosphoethanolamine (PE) accumulation in cancer cells via
the downregulation of PE-cytidylyltransferase, a rate-limiting enzyme of
phosphatidylethanolamine biosynthesis. Spatially-resolved in vivo PE-mapping
could be useful for studying the heterogeneity in metabolism within tumors. Up
till now, such mapping has been done using 31P-MRSI with the
drawback of low spatial resolution. Recently, however, Hui et al 3 have shown the feasibility of detecting
PE at 3T using the single-voxel MEGA-PRESS-sequence. 1H-MRS has the
advantage of having a higher spatial-resolution and improved SNR than 31P-MRS.
However, due to the high B1+-peak power
requirements of the PRESS-sequence, needed to minimize the chemical-shift
displacement-error (CSDE), its usage at 7T is not feasible. Here we present an
alternative EPSI-based SLOW-editing method 4,5, which enables whole-brain proton
PE-editing at UHF.METHODS
MRI/MRSI was performed on a Siemens
7T-scanner in clinical mode (MAGNETOM Terra, Germany) using the Nova 1Tx 32Rx
head coil.
The SLOW-EPSI 4–6 (Figure 1A-B) was applied on a
PE-phantom (~14 mmol/L) and in three healthy volunteers (TE=90 ms, TR=1500 ms,
matrix = 65x23x9, FOV=280x180x70 mm, and TA=9:04 min). The refocusing/editing
chemical-selective adiabatic pulse (2π-CSAP) for SLOW-editing has 35 ms
duration. The bandwidths (BW) of editing-full and editing-partial ranges are
1.7–4.3 ppm and 1.7–3.6 ppm, respectively (Figure 2A). Editing results are
obtained by the subtraction of editing-full (acquisition #1) and
editing-partial (acquisition #2) spectra. In SLOW-editing, the 2π-CSAP acts as
both a refocusing and as an editing pulse.
Simulations of all spin-systems were
performed using in-house MATLAB (R2019b) code, by solving relaxation-free
Liouville-von-Neumann equation7.RESULTS
Figure 2A shows the pulse-profiles
of editing-full (1.7–4.3 ppm) and editing-partial (1.7–3.6 ppm). The
corresponding in vitro measurement was shown in Figure 2B. Figure 2C,D shows
the simulation of PE and Glx (glutamate/glutamine = 2/1) using SLOW-editing.
Figure 3
illustrates the in vitro integration maps of editing-full and -partial
and of the difference at 3.22 ppm, which are not corrected for B1+/B1--inhomogeneity.
Figure 4A
shows the in vivo measurement (spectral editing difference) of three healthy
volunteers. The PE signal at 3.22 ppm is clearly present in both gray- and
white-matter, as well as the co-edited Glx at 2.11 ppm. Figure 4B shows the results
of editing-full and -partial of subject 1. The displaced volumes (~1.3 x 2.2 x
0.8 cm, i.e., 2.3 cm3) from which the spectra were computed are
indicated by blue (gray matter) and orange (white matter) rectangles
superimposed to the MR-images.DISCUSSION and CONCLUSION
A novel whole brain
SLOW-editing-based pulse-sequence for PE at 7T has been developed and successfully
tested in vitro and in vivo. The simulation results (Figure 2) of
PE and Glx spectral profiles agree very well with the results obtained in a
phantom and volunteers (Figure 4), indicating the close-to-ideal-behaviour of
the phase-compensated complex-sech-modulated 2-CSAP. Figure 3 shows the metabolite and derived maps
of PE, which show that there is no in-plane CSDE due to the use of 2-CSAP, in contrast to RF refocussing pulses in
MEGA-PRESS and to a lesser extend in MEGA-semiLASER. The variation observed in
the maps reflect therefore the inherent B1+-inhomogeneity
of the non-adiabatic excitation pulse. This inhomogeneity of the excitation
pulse is inevitable at 7T when using non adiabatic excitation in combination
with a non-pTx RF head-coil. Whole brain in vivo PE-editing can be
performed in less than 10 minutes measurement-time and is therefore clinically
applicable. In contrast to the referenced MEGA-PRESS sequence, our method has
the advantage that an unlimited number of ROIs can be selected and analysed in
the resulting 3D MRSI-dataset.Acknowledgements
The
research is supported by the Swiss National Science Foundation (SNSF-182569).References
- Shah T, Krishnamachary B, Wildes F, Wijnen JP,
Glunde K, Bhujwalla ZM. Molecular causes of elevated phosphoethanolamine in
breast and pancreatic cancer cells. NMR in Biomedicine. 2018;31(8).
- Osawa T, Shimamura T, Saito K, et al.
Phosphoethanolamine Accumulation Protects Cancer Cells under Glutamine
Starvation through Downregulation of PCYT2. Cell Reports. 2019;29(1).
- Hui SCN, Zöllner HJ,
Oeltzschner G, Edden RAE, Saleh MG. In vivo spectral editing of
phosphorylethanolamine. Magnetic Resonance in Medicine. Published online
August 19, 2021.
- Weng G, Kiefer C, Zubak I, Slotboom J.
SLOW: Whole Brain Spectral Editing EPSI Based Technique using Chemical
Selective Adiabatic 2π-Refocusing Pulses applied to 2HG and GABA Editing. In: ISMRM
& SMRT Annual Meeting & Exhibition. ; 2021:1990-1990.
- Weng G, Sheriff S, Kiefer C, Zubak I, Maudsley
A, Slotboom J. Simultaneous Water and Lipid Suppression Using Chemical
Selective Adiabatic Refocusing Pulses Echo Planar Spectroscopic Imaging (EPSI)
at 7T. In: ISMRM & SMRT Annual Meeting & Exhibition. ;
2021:1985-1985.
- Ebel A, Maudsley AA. Improved spectral quality
for 3D MR spectroscopic imaging using a high spatial resolution acquisition
strategy. Magnetic Resonance Imaging. 2003;21(2):113-120.
- Slotboom J, Mehlkopf AF, Bovee WMMJ. The Effects
of Frequency-Selective RF Pulses on J-Coupled Spin-1/2 Systems. Journal of
Magnetic Resonance, Series A. 1994;108(1):38-50.
- Slotboom J, Vogels BAPM,
Dehaan JG, et al. Proton Resonance Spectroscopy Study of the Effects of
L-Ornithine-L-Aspartate on the Development of Encephalopathy, Using
Localization Pulses with Reduced Specific Absorption Rate. Journal of
Magnetic Resonance, Series B. 1994;105(2).