Jie Zheng1,2, Ran Li1,2, Cihat Eldeniz1, Thomas H Schindler1, Linda R Peterson3, and Pamela K Woodard1,2
1Mallinckrodt Institute of Radiology, Washington University in St. Louis, Saint Louis, MO, United States, 2Department of Biomedical Engineering, Washington University in St. Louis, Saint Louis, MO, United States, 3Department of Medicine, Cardiovascular Division, Washington University in St. Louis, Saint Louis, MO, United States
Synopsis
A previously developed MRI method for quantitative myocardial oxygen
extraction mapping showed promising results, but image quality suffered from
distortion and low in-plane resolution. Therefore we developed a new image
acquisition method which both doubled the in-plane spatial resolution and
corrected image distortion and tested it in healthy subjects. Reproducibility
studies showed comparable results between the two methods. Rigorous animal
and/or human validation studies are warranted to study its translational
potential in the assessment of patients with myocardial metabolic dysfunction.
Introduction
In
several pathophysiologic conditions (myocardial infarction, heart failure,
valvular heart disease, etc), myocardial oxygen supply and demand (oxygen
consumption) become uncoupled. To compensate, the heart increases its
myocardial oxygen extraction fraction (mOEF), which usually precedes ventricular
mechanical dysfunction. A noninvasive, robust, and early means of assessment of
mOEF would be highly desirable. We recently developed a noncontrast
cardiovascular magnetic resonance (CMR) imaging technique to measure absolute mOEF.1
However, the current mOEF imaging technique suffers from image distortion
artifacts and relatively low spatial resolution. The objective of this study was
to develop a new imaging acquisition technique for the quantification of mOEF
which would minimize artifacts and improve image quality.Methods
The
previous CMR mOEF sequence is a 2D Asymmetric-spin-echo (ASE) echo-planar
readout sequence (ASE-EPI, dark blood), with an inplane spatial resolution of
3.4 x 3.4 mm2.1 The new CMR mOEF sequence is a 2D Asymmetric-spin-echo
(ASE) prepared balanced steady state free precession (SSFP) readout sequence
with TE of 50-60 msec (ASEprep, bright blood). Figure 1 shows the new sequence diagram with one TE. The ASE
preparation module consists of 90o-180o-180o-90o
RF pulses with alternate 0 and 180o phase to minimize B0
inhomogeneity. For a time shift of τ towards opposite directions in the
timing of the two 180o pulses, the signal intensity acquired
represents echo shift of dTE = 4τ. To accommodate short breath-hold
time, the sequence collects 14-16 images when the time shift τ changes from 0 to 3.9 – 4.5 msec. Other imaging parameters are: TE = 50 msec, inplane
resolution = 1.7 x 1.7 mm2. A trigger delay (TD) ensures the data
acquisition is within the quiescent period of the heart.
Six
healthy volunteers (24 – 30 years old, 3F) were recruited for the initial
evaluation of the sequence and reproducibility study at rest, i.e., same
imaging scans at two different days (D1 and D2). The CMR mOEF measurements was
performed along short-axis using both ASE-EPI and ASEprep sequences. For
EPI-ASE, a total of 17 images per slice with different dTE were acquired for a
total acquisition time of 17 RR-intervals within one breath-hold. For ASEprep,
a total of 16 images per slice were acquired for a total acquisition time of 16
RR-intervals. One mid slice each for the first two subjects and three slices
each for the last 4 subjects were prescribed: basal, mid, and apex. The mOEF
maps were derived with a custom-made software program and a region-of-interest
(ROI) was drawn in the septum area for all mOEF measurements to compare both
sequences. The study was performed at a 3T Prisma Siemens MR system (Siemens
Healthcare, Malvern, PA). The reproducibility was represented by the
coefficient of variation (CV) between D1 and D2.Results
Two apex slices were excluded for analysis due to severe cardiac motion
in two subjects, otherwise a total of 12 slices were included for each method
in each patient. Table 1 shows the
comparison of the resting mOEF measured in septum by both mOEF sequences. ASEprep
and EPI-ASE measurements showed no significant difference and comparable
reproducibility between the two sequences. Correlation between two sequence
measurements for septal mOEF were r =
0.8 and 0.79, slope = 1.02 and 0.98, in D1 and D2, respectively. Figure 2 shows examples of mOEF images
and maps obtained by the two sequences, demonstrating improved image quality by
using the new ASEprep sequence. In these limited number of subjects, no
significant difference in mOEF was shown between male and female subjects.
Discussion & Conclusion
We have successfully implemented a new CMR sequence to quantify mOEF,
without image distortion and with doubled in-plane resolution. Comparable
resting mOEF and reproducibility are demonstrated in this initial study.
Rigorous animal and/or human validation studies are warranted to study its
translational potential in the assessment of patients with myocardial metabolic
dysfunction.Acknowledgements
No acknowledgement found.References
1. Lu L, Eldeniz C, An H, et al. Quantification of
myocardial oxygen extraction fraction: A proof-of-concept study. Magn Reson
Med, 2021;85:3318-3325.