Alfredo Ordinola1 and Evren Özarslan1
1Department of Biomedical Engineering, Linköping University, Linköping, Sweden
Synopsis
The distribution of net displacements, commonly
referred to as the ensemble average propagator, has been measured using
Stejskal-Tanner experiments. Here, a recently introduced diffusion
sensitization method is employed in a similar manner to obtain the distribution
of mean positions. By utilizing the two schemes synergistically, low b-value
measurements are employed to decouple dynamic second moments from static
ones. We illustrate our findings on an excised mouse spinal cord imaged using a
benchtop MRI scanner.
Introduction & Theory
Recently, a novel technique for measuring the
diffusion propagator was introduced1. The employed diffusion encoding gradient waveform (Figure 1a) involves two
independent wavenumbers, $$$q$$$ and $$$q’$$$. Sampling different lines on the
$$$qq’$$$ plane as shown in Figure
1b could reveal different characteristics of diffusion. For example, the
blue lines ($$$q’=0$$$ or $$$q=0$$$) depict the sequence by Laun et al.2, used for diffusion
pore imaging. Stejskal-Tanner measurements3 sample the $$$q’=-q$$$ line
(green) and reveal the distribution of net displacements through4,5
$$\bar{P}_\Delta(\Delta\mathbf{x})=\frac{1}{(2\pi)^d}\int\rm{d}\mathbf{q}\,e^{i\mathbf{q}\cdot\Delta\mathbf{x}}\,E(\mathbf{q})\qquad (1).$$
Here, $$$\Delta\mathbf{x}=\mathbf{x}’-\mathbf{x}$$$,
where $$$\mathbf{x}$$$ and $$$\mathbf{x’}$$$ represent the positions encoded by
the two pulses, respectively. The low-$$$\mathbf{q}$$$ regime of the signal is
governed by
$$E(\mathbf{q})\approx\exp\left[-\left(q_x^2\langle{\Delta{x}}^2\rangle+q_y^2\langle{\Delta{y}}^2\rangle+q_z^2\langle{\Delta{z}}^2\rangle+2q_xq_y\langle{\Delta{x}}{\Delta{y}}\rangle+2q_xq_z\langle{\Delta{x}}{\Delta{z}}\rangle+2q_yq_z\langle{\Delta{y}}{\Delta{z}}\rangle\right)\right]\qquad
(2).$$
More generally for the waveform in Figure 1a, a particle’s
coordinates during the application of the second and third pulses ($$$\mathbf{x}$$$
and $$$\mathbf{x’}$$$) are encoded in a frame of reference whose origin is at
the center-of-mass of that particle’s trajectory during the first pulse1,6. For
the case of $$$q’=q$$$, we introduce $$$\mathbf{Q}=2\mathbf{q}$$$ and define
the “mean position” as $$$\mathbf{\bar{x}}=(\mathbf{x}+\mathbf{x’})/2$$$.
The signal $$$E(\mathbf{Q})$$$ acquired through
this scheme can then be used to estimate the mean position distribution through
the following Fourier Transform:
$$\bar{P}_\Delta(\mathbf{\bar{x}})=\frac{1}{(2\pi)^d}\int\rm{d}\mathbf{Q}\,e^{i\mathbf{Q}\cdot\mathbf{\bar{x}}}\,E(\mathbf{Q})\qquad (3).$$
The low-$$$\mathbf{Q}$$$ behavior of the signal
is then,
$$E(\mathbf{Q})\approx\exp\left[-\left(Q_x^2\langle{\bar{x}}^2\rangle+Q_y^2\langle{\bar{y}}^2\rangle+Q_z^2\langle{\bar{z}}^2\rangle+2Q_xQ_y\langle{\bar{x}}{\bar{y}}\rangle+2Q_xQ_z\langle{\bar{x}}{\bar{z}}\rangle+2Q_yQ_z\langle
{\bar{y}}{\bar{z}}\rangle\right)\right]\qquad (4).$$
Using Equation 4, the position-dependent quantities
can be cast as a squared mean-position tensor. Similarly, from Equation 2, a squared
displacement tensor can be obtained. Simple algebraic manipulations of these
tensors’ components reveal 12 unique quantities, six of which are “static” (at
the same point in time) correlations of particle positions: $$$\langle{x^2}\rangle,\langle{y^2}\rangle,\langle{z^2}\rangle,\langle{xy}\rangle,\langle{xz}\rangle,\langle{yz}\rangle$$$.
The remaining quantities for “dynamic” (at different points in time)
correlations are: $$$\langle{xx'}\rangle,\langle{yy'}\rangle,\langle{zz'}\rangle,\langle{xy'}\rangle,\langle{xz'}\rangle,\langle{yz'}\rangle$$$.
Here, we present the very first application of
the above-introduced techniques on biological tissue.Methods
The pulse sequence presented in Figure 2 was implemented on
a benchtop MRI scanner (Pure Devices GmbH, Germany). This pulse sequence
differs from a previous embodiment of the technique6 in that only the first gradient of the
effective waveform of Figure 1a was implemented as a bipolar pulse pair in an effort
to approach the narrow pulse regime.
The investigated sample was an excised mouse
spinal cord (obtained following national and local ethical guidelines). All imaging
was performed with a matrix size of $$$48\times48$$$ and an in-plane voxel size
of approximately $$$0.1\times0.1 mm^2$$$.
For the first experiment, all diffusion
gradients were applied along a single direction, perpendicular to the direction
of the fibers in white matter. Two datasets were acquired, respectively, with
$$$q’=q$$$ and $$$q’=-q$$$. The values of q applied in this study ranged
from $$$-0.8 rad/\mu m$$$ to $$$+0.8 rad/\mu m$$$ with 15 regularly-spaced samples.
Each acquisition was repeated 20 times, averaged, and filtered with a simple Gaussian
filter.
The second experiment involved a two-shell
acquisition with 6 and 21 directions applied at low and high b-values,
respectively. Two datasets were acquired, the first setting $$$q’=q$$$ and the
second with $$$q’=-q$$$. The values of q were set to obtain approximately the
same b-value in both cases. Each acquisition was repeated
25 times, averaged, and filtered with a simple Gaussian filter.
Complex
data acquired from the scanner were used in the reconstruction pipeline for the
first experiment, which involved the estimation of the mean position
distribution $$$\bar{P}_\Delta(\bar x)$$$, and the
displacement distribution $$$\bar{P}_\Delta(\Delta\mathbf{x})$$$4,7.
For the second experiment, magnitude-valued
data from both acquisitions were processed to estimate the squared displacement
and squared mean-position tensor components following Equations 2 and 4,
respectively.
All computations were performed using MATLAB (Mathworks
Inc., Natick, MA, USA).Results
The real part of the estimated displacement
distribution and mean position distribution for voxels corresponding to,
respectively, white matter and gray matter are presented in Figure 3.
Four of the twenty-seven images corresponding
to the spinal cord sample acquired in the second experiment for each
acquisition, i.e., with $$$q’=-q$$$ and $$$q’=q$$$, are presented in Figure 4.
The 12 unique component images corresponding to
both static and dynamic correlations of positions obtained from the squared displacement
and squared mean-position tensors are presented in Figure 5.Discussion & Conclusion
In this study, we introduced a new method for
determining the mean-position distributions. It should be kept in mind that the
mean-position is defined in a frame of reference whose origin coincides with
the average position of the particles during the application of a long pulse.
Because of this, the distributions have finite extent much like the displacement
distributions revealed by the Stejskal-Tanner sequence, which is the standard
method for investigating compartmentalization in biological specimens5.
There are important differences in the two distributions however, as
illustrated by our results from the spinal cord specimen.
The low diffusion sensitivity regime of the
signal is also studied here. The required sampling resembles that for diffusion
tensor imaging8 and complements it by revealing the second moments of the mean
positions. Synergistic application of the two techniques unraveled dynamic
moments from static ones providing various contrasts within the excised mouse
spinal cord.
In conclusion, a sophisticated framework involving
new data acquisition and analysis schemes is introduced. Its feasibility is
illustrated on a benchtop MRI scanner. The technique complements the existing
methods in a meaningful way and could be instrumental in investigating numerous
challenges that demand novel characterization methods.Acknowledgements
We are grateful to Anders Eklund for his support in the acquisition of the MR scanner, and to Walker Jackson for providing the sample investigated in this study.References
[1] Özarslan E. Recovering
almost everything diffusion could reveal. Proc Intl Soc Mag Reson Med, Vol. 29. 2021:p.
3637.
[2] Laun FB, Kuder TA, Semmler W, Stieltjes B. Determination of the Defining Boundary in Nuclear Magnetic Resonance
Diffusion Experiments. Phys Rev Lett. 2011;107(048102).
[3] Stejskal EO and Tanner JE. Spin Diffusion Measurement: Spin Echoes in the Presence of a
Time-Dependent Field Gradient. J Chem Phys. 1965;42:288.
[4] Stejskal EO. Use of spin
echoes in a pulsed magnetic-field gradient to study anisotropic, restricted
diffusion and flow. J Chem Phys. 1965;43 (10):3597–603.
[5] Kärger J and
Heink W. The propagator representation of molecular transport in microporous
crystallites. J Magn Reson 1983;51(1):1–7.
[6] Ordinola Santisteban AM and Özarslan E. Magnetic
resonance measurement and reconstruction of the diffusion propagator,
arXiv:2106.16181. 2021.
[7] Cory DG and Garroway AN. Measurement of
Translational Displacement Probabilities by NRM: An Indicator of
Compartmentation. MRM. 1990; 14:p. 435-444.
[8] Basser PJ, Mattiello J, LeBihan D. Estimation of the effective self-
diffusion tensor from the NMR spin echo. J Magn Reson B 1994;103:247–254.