Karl Philipp Kunze1, Nina Mellor2, Tracy Moon2, Kuberan Pushparajah3, Radhouene Neji1, and Amedeo Chiribiri3
1MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom, 2Evelina London Children’s Hospital, Guy’s and St Thomas’ NHS Foundation Trust, London, United Kingdom, 3Biomedical Engineering and Imaging Science, King's College London, London, United Kingdom
Synopsis
This abstract describes a feasibility study for
a high-resolution perfusion imaging framework, including a dedicated sampling
pattern and a motion compensation framework integrated into temporal regularization and implemented
in-line on the scanner with GPU support. The approach was employed successfully in patients with suspected myocarditis at rest, showing superior
ability to handle both respiratory and cardiac motion at an image resolution of 1.3 mm2, as compared to a clinical reference perfusion sequence and supported by Late Gadolinium Enhancement imaging findings.
Introduction
Standard clinical myocardial perfusion MR techniques
are hampered by a lack of spatial resolution and coverage. Both cardiac and
respiratory motion are a challenge to image quality, and necessitate breath held
exams and very fast single-shot acquisitions, with acquisition duration of
single slices being the limiting factor for anatomical coverage. To increase
spatial resolution and/or shorten single-shot acquisition times, high acceleration
factors are usually unavoidable. These are, however, only achievable with some
degree of regularization along the temporal dimension of the dynamic perfusion
series, which leads to strong blurring artifacts in the case of motion between
frames. This abstract describes a feasibility study for a high-resolution
perfusion imaging framework, including a dedicated sampling pattern and a reconstruction which incorporates motion compensation in the temporal regularization implemented
in-line on the scanner with GPU support.Methods
An MR perfusion prototype sequence for a 2D
single-shot, saturation-recovery acquisition was implemented featuring a
dedicated sampling pattern combining incoherent undersampling for high-, and linear
undersampling low spatial frequencies. A prototype reconstruction with motion
compensation was implemented inline on the scanner with GPU support, featuring a
preliminary reconstruction of all frames at reduced resolution without temporal
regularization, an estimation of non-rigid motion fields between all frames1,
and a final reconstruction with incorporation of motion fields into a temporal
regularization term2,3. The described sequence and reconstruction were
applied in 3 patients with suspected Myocarditis (ages 12, 16 and 17 years) on
a 3T Scanner (MAGNETOM Vida, Siemens Healthcare, Erlangen, Germany), with the
following parameters: 1.3mm2 acquired resolution, 8 mm slice
thickness, 3-4 slices per heartbeat, acceleration factor 5.5-6.0, FLASH
readout, TE/TR 1.19/2.71 ms, BW 707 Hz/px, TI 110 ms, FA 15 deg, chemical-shift
selective fat saturation, single slice readout time 105 ms (excluding sat
pulses, delay times and FatSat). The sequence was applied for 90 dynamics in
free breathing during a bolus injection (0.15 mmol/kg Gadobutrol), and
reconstructed with (MC) and without (NMC) the integrated motion compensation. In
addition, a standard clinical sequence was applied as a reference directly
afterwards for 40 dynamics (acquired/reconstructed resolution 2.3 x 1.9 / 1.9 x 1.9 mm2,
TPAT acceleration factor 2, TE/TR 1.03/2.16 ms, BW 1003 Hz/px, TI 95 ms, single
slice readout time 125 ms). Results
In-line image reconstruction times for the described
high-resolution sequence were 20-30 s per slice. Image quality for MC was excellent
in all cases despite heavy respiratory and cardiac motion as revealed by the
comparison of MC and NMC reconstructions. Figure 1 shows a comparison of MC and
NMC images in a patient at different points during the perfusion series. For
the same patient, Figure 2 shows MC, TPAT and LGE images, the latter confirming
acute myocarditis and validating the contrast pattern seen in both MC and TPAT images
during the last frames. Figure 3 shows a second patient with both cardiac and
respiratory motion, leading to significant blurring in NMC, but no significant
degradation in MC images.Discussion
Despite the limited amount of clinical data, the
study at hand demonstrates feasibility for a fast in-line implementation of free-breathing,
high-resolution myocardial perfusion data in a clinical setting. While further
validation studies are warranted to demonstrate the added clinical value of the
presented approach, cases shown here, including confirmation by LGE findings,
hint at the potential for superior depiction of small perfusion abnormalities. Conclusion
A high-resolution, free-breathing myocardial
perfusion approach with a fast, inline motion-compensated reconstruction was
employed successfully in patients with suspected myocarditis, showing very good ability to handle both respiratory and cardiac motion at a resolution of 1.3mm2.Acknowledgements
No acknowledgement found.References
1) Xue H,
Zuehlsdorff S, Kellman P, et al. Unsupervised inline analysis of cardiac
perfusion MRI. In Proceedings of the International Conference on Medical Image
Computing and Computer-Assisted Intervention, 2009, 741-749.
2) McElroy S, Kunze
KP, Nazir S, et al. Stress first pass myocardial perfusion using simultaneous
multi-slice (SMS) SSFP with iterative reconstruction and integrated motion
compensation. Proceedings of the Annual SCMR Meeting, 2021.
3) Muehlberg
F, Stoetzner A, Forman C, et al. Comparability of compressed sensing-based gradient
echo perfusion sequence SPARSE and conventional gradient echo sequence in
assessment of myocardial ischemia. European Journal of Radiology, 2020, 131:109213.