Felicia Seemann1, Ahsan Javed1, Rachel Chae1, Amanda Potersnak1, Scott Baute1, Kendall O’Brien1, Rajiv Ramasawmy1, Robert J Lederman1, and Adrienne E Campbell-Washburn1
1Cardiovascular Branch, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, MD, United States
Synopsis
Quantification
of changes in lung water using MRI may provide clinical value in early diagnosis
of heart failure. In this study, we demonstrate the ability of our 3D free-breathing spiral ultrashort-TE sequence to quantify and capture changes in lung water using
a high-performance 0.55T MRI system. A phantom
containing mixtures of water and deuterium oxide at varying concentrations was used to validate water
density quantification. Our proposed approach captured the
gravity-induced redistribution of lung water, achieved by imaging subjects in
the supine and prone positions.
Introduction
Quantification of dynamic changes in lung water density (LWD) could
potentially enable an early diagnosis of heart failure, a task that can be challenging
using current conventional clinical assessments.
Magnetic resonance imaging (MRI) has recently been proposed as an ionizing
radiation-free and quantitative method to measure LWD, using half-Fourier single-shot turbo spin-echo (HASTE) or ultrashort
echo time (UTE) sequences at 1.5T and 3T (1–3). However, assessment of lung water dynamics has not been
evaluated for these methods. High-performance low field strength 0.55T
MRI offers reduced image artifacts in tissues in proximity to air, such as lung
parenchyma, and due to reduced susceptibility gradients (4–7). In this study, we present a
free-breathing UTE acquisition and an automated image processing pipeline to
capture changes in lung water density using a high-performance 0.55T MRI system
in patients and healthy volunteers.Methods
In this IRB-approved study, lung MRI was performed on 14 healthy
volunteers (age 29±7 years, 6 women) and two
patients with pulmonary hypertension (44±14 years, 2 women)
using a high-performance 0.55T system (prototype MAGNETOM Aera, Siemens Healthcare, Erlangen, Germany) (4). A 3D free-breathing
stack-of-spirals proton density weighted UTE gradient echo sequence (TE/TR/θ =
0.56ms/9ms/1°, 171 spiral interleaves, 5.0ms spiral readout, 3.5mm isotropic
resolution, 450x450x252mm FOV, 2.5 minute acquisition time) was used to measure
LWD (8). Images were reconstructed
at a stable respiratory phase using a self-navigator signal (8), and corrections for
surface coil shading due to non-uniform receiver coils was applied. In healthy volunteers, a gravity-induced dynamic redistribution of lung
water was achieved by sequentially acquiring UTE images in the supine, prone,
and again supine position. Four images were acquired in each
position. In patients, two images were acquired in the supine position only.
An automated image processing pipeline was
implemented (Figure 1). First, lung segmentation avoiding major
vasculature and airways was performed using a trained convolutional neural
network with a U-net architecture developed in Pytorch. A region of interest in
the liver was automatically placed under the right lung. Occasional manual
corrections were applied when necessary. Pixel-wise LWD maps were calculated as
the ratio of lung tissue signal intensities compared to average liver signal
intensity, with the assumption that the liver has 70% water density (1).
A
test-retest validation of the quantitative capabilities of the UTE sequence was
performed in an array of vials containing mixtures of water and deuterium
oxide. Water densities ranged between 0-100% in
increments of 10%, and were calculated as the
ratio of the average signal intensities in each vial to the 100% water vial.Results
The test-retest phantom experiment validated the capability of the
UTE sequence in quantifying water
density and the reproducibility of the measurement, with a bias of 4.3±4.8% (Figure 2). A redistribution of LWD was
observed in all healthy volunteers when repositioning, with a predominant lung
water accumulation in the posterior lung when supine and in the anterior lung
when prone (Figure 3). Lung water redistribution had occurred before the
first image in each new position (mean time between positions = 6 min), and no
evolution in LWD was observed during four repeated measurements.
The average global LWD was 21±3.5% in healthy volunteers, which
remained unchanged between imaging positions (supine 21±3.3%, prone 21±3.5%,
second supine 21±3.4%) (Figure 4). There was no difference in LWD in
healthy volunteers compared to patients with pulmonary hypertension (21±3.5% vs
23±0.4%). The difference in global LWD between the first and following three
acquired images in each position was -0.54 ± 1.5%,
demonstrating that the proposed LWD metric is repeatable. Discussion
This study presents a free-breathing stack-of-spirals 3D UTE MRI
sequence at 0.55T with an automated image processing pipeline to quantify LWD
in ~2.5 minutes. The method was capable of depicting a redistribution of LWD,
and quantification of global LWD was reproducible in both in-vitro and in-vivo
studies. The
studied cohort was predominantly young and healthy, and future studies and
further method development are warranted to determine if MRI quantification of
dynamic changes in LWD during exercise can unmask latent heart failure.Conclusion
Redistribution of LWD due to gravitational forces can be depicted
and quantified using a validated free-breathing 3D proton density weighted
UTE sequence on a high-performance 0.55T MRI system. Acknowledgements
We would like to acknowledge the assistance of Siemens Healthcare
in the modification of the MRI system for operation at 0.55T, and in the
stack-of-spirals UTE sequence, under an existing cooperative research agreement
(CRADA) between NHLBI and Siemens Healthcare.
This study was supported
by the Intramural Research Program, National Heart Lung and Blood Institute,
National Institutes of Health, USA (Z01-HL006257, Z01-HL006213, and Z01-HL006039).
References
1. Thompson
RB., Chow K., Pagano JJ., et al. Quantification of lung water in heart failure
using cardiovascular magnetic resonance imaging. J Cardiovasc Magn Reson
2019;21(1):58. Doi: 10.1186/s12968-019-0567-y.
2. Meadus WQ., Stobbe RW., Grenier JG., Beaulieu C., Thompson
RB. Quantification of lung water density with UTE Yarnball MRI. Magn Reson Med
2021. Doi: 10.1002/mrm.28800.
3. Burrage MK., Hundertmark M., Valkovič L., et al. Energetic
Basis for Exercise-Induced PulmonaryCongestion in Heart Failure With
PreservedEjection Fraction. Circulation 2021. Doi:
10.1161/CIRCULATIONAHA.121.054858.
4. Campbell-Washburn AE., Ramasawmy R., Restivo MC., et al.
Opportunities in Interventional and Diagnostic Imaging by Using
High-Performance Low-Field-Strength MRI. Radiology 2019;293(2):384–93. Doi:
10.1148/radiol.2019190452.
5. Campbell-Washburn AE., Malayeri AA., Jones EC., et al.
T2-weighted Lung Imaging Using a 0.55-T MRI System. Radiol Cardiothorac Imaging
2021;3(3):e200611. Doi: 10.1148/ryct.2021200611.
6. Simonetti OP., Ahmad R. Low-Field Cardiac Magnetic Resonance
Imaging. Circ Cardiovasc Imaging 2017;10(6). Doi:
10.1161/CIRCIMAGING.117.005446.
7. Atalay MK., Poncelet BP., Kantor HL., Brady TJ., Weisskoff
RM. Cardiac susceptibility artifacts arising from the heart-lung interface.
Magn Reson Med 2001;45(2):341–5. Doi:
10.1002/1522-2594(200102)45:2
8. Javed A., Ramasawmy R., O’Brien K., et al. Self-gated 3D
Stack-of-Spirals Ultra-Short Echo-Time Pulmonary imaging at 0.55T. Magn Reson
Med 2021.