Edwin Versteeg1, Mark van Uden2, Luke van der Hofstad2, Kevin Bax2, Martino Borgo2,3, Wout Schuth3, Jeroen Siero1,4, Mark Gosselink1, and Dennis Klomp1
1Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Tesla Dynamic Coils, Zaltbommel, Netherlands, 3Futura Composites, Heerhugowaard, Netherlands, 4Spinoza Centre for Neuroimaging, Amsterdam, Netherlands
Synopsis
The sound level
in MRI can be lowered by using a silent gradient axis that adds extra soundless
encoding. In this work, we present the field characteristics and first images
for a silent gradient designed for a clinical 3 Tesla system. This silent
gradient featured a built-in 32-channel receive coil and weighs 21 kg. The
gradient field could be oscillated at 23.5 kHz and featured a linear region of
20 cm. The coil had limited interactions with the transmit B1 field and did not
affect B1 homogeneity, which was showcased in phantom and in-vivo images.
Introduction
Most MR-exams
intrinsically feature loud acoustic noise as the imaging gradients need to be
switched fast and at high amplitude to ensure short acquisition times. Together
with claustrophobia these loud sounds are the main sources of patient anxiety
during MR-exams, which might lead to a degradation of the image quality (1,2). Previously, we have presented a silent
gradient axis that aims to lower sound levels by providing extra soundless
spatial encoding and allow the other imaging gradients to be switched slower,
which reduces the sound level without loss of imaging time(3). In this work, we present the field
characteristics and first images for a silent gradient designed for a clinical 3 tesla MR-system.Methods
Gradient
design
An open and
lightweight design was targeted for the silent gradient (Futura composites,
Heerhugowaard & Tesla DC, Zaltbommel).
The silent gradient coil consisted of 32 windings distributed in two
groups (16 windings per group), which were spaced 25 cm apart and produced a
z-gradient field. By grouping the windings like this, the amount of epoxy
needed to pot the gradient coil was minimized resulting in a weight of 21 kg of
the whole coil, which facilitates easy installation comparable to a standard
head coil. The gradient coil was made
resonant at 23.5 kHz using two capacitors and combined with a modified gradient
amplifier (NG500, Prodrive Technologies, Son) to maximize the silent encoding (4). Furthermore, a 32-channel receive coil array
was integrated into the coil with the receive elements distributed over 4 rows
extending in the feet-head direction to facilitate parallel imaging along the
direction of the silent gradient. This receive array consisted of two attached PCB’s
containing all coils which were decoupled based on a predetermined overlap. Figure 1 shows the final coil design.
Characterization
The gradient
field distribution was calculated using the Biot-Savart law and was used to determine
gradient efficiency and field linearity. The exact resonance frequency and
spatiotemporal gradient field behavior were measured using a dynamic field
camera (Skope Magnetic Resonance Technologies
AG, Switzerland). During these measurements the gradient coil was
positioned in a 7T MR-system (Philips, The Netherlands).
To investigate
transparency of the setup to the transmit field of the 3T RF body-coil, B1-mapping
was performed with the same subject, once with the setup and once with the
standard Philips head coil.
Images
For imaging
experiments, the RF coil array with the embedded silent gradient was positioned
in a 3T MR-system (Ingenia CX, Philips, Best).
A phantom consisting of a water-filled bottle was imaged to investigate
the imaging performance along the linear region of the gradient. Here, a
modified 2D gradient-echo sequence that plays out the silent gradient during
readout was used with the following imaging parameters: coronal slice
orientation, in-plane resolution = 2 x 2 mm2, FOV = 256 x 512 (RL x
FH) mm2, slice-thickness = 4 mm, TE = 5.6 ms, TR = 60 ms, and flip
angle = 16°. Additionally, a similar
sequence was used to image a volunteer with the following imaging parameters:
coronal slice orientation, in-plane resolution = 1 x 1 mm2, FOV =
256 x 256 (RL x FH) mm2, slice-thickness = 6 mm, TE = 21 ms, TR = 60
ms, and flip angle = 21°. All images were
acquired with the silent gradient operating at 22.5 mT/m. Reconstruction was
performed off-line in MATLAB using the field camera data and a non-uniform
Fourier transform.Results and Discussion
Figure 2a shows
the field distribution of the silent gradient. Here, the gradient efficiency
was determined to be 0.56 mT/m/A. Figure 2b shows the deviation from a
perfectly linear gradient field for the silent gradient, which shows that the
gradient field has less than 10% deviation from linearity in a 20 cm DSV. The
field measurements in figure 2C shows the gradient field oscillating at 23.5
kHz and at an amplitude of 22.5 mT/m. Additionally, the silent gradient needs
about 1.5 ms to reach a steady state and ~1 ms to return back to zero.
The B1-maps (Fig 3ab)
shows similar performance with both setups in the MR-system suggesting that the
interaction with the gradient coil is limited. On average the B1 measured with
the silent gradient system was ~10% lower. However, the B1-field homogeneity
was not affected as the distribution of B1-values (Figure 3c) was similar for both
setups (σB1-silent gradient = 2.3 µT vs σB1-standard
coil = 2.4 µT).
Figure 4a shows
that no significant distortion is observed over the 20 cm linear region.
However, some signal loss was seen outside this region, which might be recovered
by incorporating the field non-linearity in the reconstruction. In Figure 4b,
the in-vivo image shows that the whole-brain fits in the 20 cm linear region.
Earlier work on
silent gradient axes was performed at 7 tesla, where a 26 dB reduction in sound
level was achieved for a T1-weighted anatomical scan (5). Translating the concept of the silent
gradient axis to 3 tesla is expected to yield similar or lower sound levels as
the Lorentz forces at 3 tesla are intrinsically lower.Conclusion
We have presented
a silent gradient designed for 3 Tesla MR-system, which can enable quiet and
fast brain imaging.Acknowledgements
No acknowledgement found.References
1.
McNulty JP, McNulty S. Acoustic noise in magnetic resonance imaging: An ongoing
issue. Radiography 2009;15:320–326 doi: 10.1016/j.radi.2009.01.001.
2. Dewey M, Schink T, Dewey CF. Claustrophobia during
magnetic resonance imaging: Cohort study in over 55,000 patients. J. Magn.
Reson. Imaging 2007;26:1322–1327 doi: 10.1002/jmri.21147.
3. Versteeg E, Klomp DWJ, Siero JCW. A silent
gradient axis for soundless spatial encoding to enable fast and quiet brain
imaging. Magn. Reson. Med. 2021;00:1–12 doi: 10.1002/MRM.29010.
4. Welting D, Versteeg E, Voogt I, et al. Towards 8ch
multi transmit with high power ultrasonic spirals and 72ch receive setup for
ultimate spatial encoding at 7T. In: Proceedings of the 29th Annual Meeting of
ISMRM. ; 2021. p. 0563.
5. Versteeg E, Jacobs SM, Oliveira ÍAF, Klomp DWJ,
Siero JCW. Fast and quiet 3D MPRAGE using a silent gradient axis- sequence
development. In: Proceedings of the 29th Annual Meeting of ISMRM. ; 2021. p.
0842.