Michael Eijbersen1,2, Bart Steensma1,2, Cornelis A.T. van den Berg1,2, and Alexander J.E. Raaijmakers1,2,3
1Radiotherapy, UMC Utrecht, Utrecht, Netherlands, 2Computational Imaging Group, UMC Utrecht, Utrecht, Netherlands, 3Biomedical Engineering, Eindhoven University of Technology, Eindhoven, Netherlands
Synopsis
A new computational method is presented for determining a transfer function of implants using a MRI only paradigm. By relaxing the transciever phase approximation better results are obtained for longer implants, however an optimization for both B1+ and B1- based on magnitude only maps is required.
Introduction
The transfer function (TF) is an important
characteristic of elongated implant structures. It relates the incident
electric field along the implant to the electric field enhancement at the tip
of the implant. The assessment of the transfer function is part of the RF
safety assessment for MRI-safe labelling as prescribed in the ISO standard [1]. The transfer function is conventionally measured using a bench
setup. However, Tokaya et al introduced an alternative method where the MRI
system is being used to measure the transfer function of elongated implant
structures in an ASTM phantom from the artefacts that the induced RF currents
generate [2]. The method, however, assumes negligible z-components of
the B1-field (transceiver phase approximation), which is only valid for
relatively short implant structures. In this work, we introduce an improved computational
method to asses the transfer function of an implant, without this assumption making the method
applicable for implant wires/leads as long as the field of view of the scanner.
Using simulated data generated by a
model in Sim4Life we show that this
computational method is more accurate than previous work.Materials and Methods
The MRI-based TF measurement techniques
makes use of the concept of the transfer matrix (TM) which relates incident
electric fields along the implant to induced current on the implant:
$$I=ME_{inc}$$.
A parameterized version of the TM can be
determined if incident Ez-field and induced current are known.
From Maxwell’s equations we may derive the relation
between measured B1- fields and the Incident Ez-fields.
$$ E_z=\frac{1}{\mu_0\sigma+i\omega\mu_r\mu_0\epsilon_r\epsilon_0}\left(-i\frac{\partial B^{1+}}{\partial x}-\frac{\partial B^{1+}}{\partial y}+i\frac{\partial B^{1-}}{\partial x}-\frac{\partial B^{1-}}{\partial y}\right)$$
The transceive phase approximation assumes $$$\frac{\partial Bz}{dz}=0$$$ in which case the following relation
holds:
$$\frac{\partial B_1^{+}}{\partial x}+i\frac{\partial B_1^{+}}{\partial y}=-\frac{\partial B_1^{-}}{\partial x}+i\frac{\partial B_1^{-}}{\partial y}$$
This allows the incident electric field to be determined by only B1+. If the assumption is not valid, both field
components need to be determined which, for a homogeneous phantom, can be
obtained from the proton density scaling factor in the multi-flip angle
B1-measurement method [2,3].
Induced current is determined by fitting
the B1+ and B1- fields to a EM
field model where the background field is modeled as a superposition of spherical harmonics while the scattered
field is modeled through
Jefimenko’s generalization of the
Biot-Savart law:
$$B_{1,bg}^{\pm}=\sum_{n,l,m} j_l(kr)Y^l_m(\theta,\phi)$$ $$B_{1,sc}^\pm(\vec{r})\exp(i\omega t)=\frac{4\pi}{\mu_0}\int_{\text{wire}}I^\pm(\vec{r}')\exp\left(t-\frac{|\vec{r}'-\vec{r}|}{c_m}\right)\left(\frac{1}{|\vec{r}'-\vec{r}|^3}+\frac{i\omega}{c_m|\vec{r}'-\vec{r}|^2}\right)dr'\times (\vec{r}'-\vec{r})$$
The combined scattered and background
field (d) in the data may be written as a matrix
vector product:
$$\vec{d}=A\vec{x}+J\vec{y}$$
Here the matrix A and J contain
discretized versions of the background and scattered field terms respectively. Since only magnitudes of the data
points are available in experiments, the problem is
reformulated in a Magnitude Squared Least Squares (MSLS)
setting. This allows for working with
inner products which are linear and allows us to rewrite the problem in
terms of z, which is the concatenation of
x and y with their real and imaginary part splitted. In this formulation
we may derive equations for the cost
function, gradient and hessian in order to use a Trust Region Newton
algorithm to find an acceptable local minimum. A schematic representation of the derivation is provided in figure 2.
The proposed method was tested using
simulated |B1+| and |B1-| field distributions of a realistic birdcage body coil
in the conventional ASTM phantom (ϵr = 78.0 and
σ = 0.34 S/m) which
contains a 42 cm insulated copper wire (figure 1). Resulting Ez and Iind
vectors were used to determine the TM of which the first column is the TF.
Results and Discussion
Figure 3 shows the
simulated B1+ and B1- distributions
along with the corresponding reconstructed EM field models. Clearly, the model
captures the behavior of background and scattered field very well. The reconstructed
Ez and Iind values along the wire are indicated in figure 4 along with their simulated
ground-truth counterparts. Computation of Ez using only B1+ based on the transceiver phase approximation clearly shows unacceptable deviations from the ground truth in this case. Data shows that both magnitude and
phase reveal some minor deviations, which do not severely affect the TF
accuracy. The resulting normalized TM and TF are presented in figure 5 along
with the ground truth simulated counterparts. These results show that the
proposed method is capable of calculating the transfer function from MRI-accessible
data, even for longer implants where the transceive phase approximation is no
longer valid.Conclusion
A MRI-based measurement method for TF
determination without restrictive assumptions was presented. For this purpose,
a calculation pipeline was developed to acquire incident electric fields from
B1+ and B1- distributions. A EM-model was setup to parameterize the
background-scattered field combination. Application on simulated field
distributions shows that the method is capable of determining the transfer function
from MRI-accessible data.Acknowledgements
No acknowledgement found.References
- Implants ISA. Iso/ts 10974:2012(en). Assessment of the safety of magnetic resonance imaging for patients with an active implantable medical device. 2012
- Tokaya, Janot P., et al. MRI‐based transfer function determination through the transfer matrix by jointly fitting the incident and scattered field. Magnetic resonance in medicine, 2020, 83.3: 1081-1095.
- van den Bosch, Michiel R., et al. New method to monitor RF safety in MRI‐guided interventions based on RF induced image artefacts. Medical physics, 2010, 37.2: 814-821.