Evan Cummings1,2, Yuchi Liu2, Yun Jiang1,2, Kathleen Ropella-Panagis2, Jesse Hamilton1,2, and Nicole Seiberlich1,2
1Biomedical Engineering, University of Michigan, Ann Arbor, MI, United States, 2Radiology, University of Michigan, Ann Arbor, MI, United States
Synopsis
In this work, Rosette MRF was modified and deployed to
simultaneously map T1, T2, and T2*. Mapping is
demonstrated on the ISMRM/NIST phantom for verification of quantitative accuracy,
and on four healthy volunteers to demonstrate potential feasibility in vivo.
Introduction
The goal of this work is to rapidly
and simultaneously acquire T1, T2, and T2*
maps on a 0.55T MRI system using MR Fingerprinting with a rosette trajectory. T2*
mapping is increasingly used in the heart and liver to monitor iron concentration
in patients with hereditary hemochromatosis, thalassemia major, or sickle cell
disease1,2. These maps are commonly collected at 1.5T and 3T, but
may be more challenging to collect at lower field strengths due to the increase
in T2* values3,4. While longer T2* values can
be beneficial in some applications, T2* mapping becomes more
difficult, as longer readouts are required to capture the longer decays. This
is particularly challenging in liver and cardiac applications, where there is a
limited amount of data that can be collected within a single breathhold.
MR Fingerprinting5,6 is
an efficient technique for multiparametric mapping using a non-steady state
pulse sequence, and simultaneous T1, T2, and T2*
mapping has been demonstrated at 1.5T using a Dixon MRF technique7. The
rosette trajectory has recently been examined as a readout for MRF, as it
repeatedly crosses the center of k-space8, and can be used to
separate fat and water signals for more accurate T1 and T2
quantification in the heart9. Additionally, rosettes have been used
for T2* mapping outside of the MRF framework10. In this
work, the feasibility of collecting accurate T1, T2, and T2*
maps in the liver in a single breathhold at 0.55T is explored.Methods
All data were collected on a 0.55T FreeMax scanner (Siemens
Healthineers, Erlangen, Germany). Scans were performed on the ISMRM/NIST
phantom as well as four healthy volunteers using a 6-channel body receiver coil
array in this IRB-approved study. Reference T1 values for the
ISMRM/NIST phantom were acquired using an inversion recovery spin echo sequence,
and reference T2 values were acquired using a spin echo sequence. Reference
T2* maps were generated from a gradient echo sequence with the
following parameters: TR=150ms, 12 echoes spaced between 4.06 and 48.28 ms, FA=25°, 128x128
matrix, 300 mm2 FOV, 8 mm slice thickness. Rosette MRF data were
collected using an 18-lobe rosette with the following parameters: TR=41.4 ms,
TE=1 ms, 37.0 ms readout duration, 336 excitations, 14.1 s total scan duration,
FA ranged between 0° and 75°, 192x192 matrix, 300 mm2
FOV, 8 mm slice thickness. Inversions were used every 72 repetitions.
The
reconstruction process for these maps is shown in Figure 1. The rosette trajectory
is split into multiple echoes for each TR10, and the NUFFT11
is used to generate 36 highly accelerated single echo images for each
repetition. PCA is then applied to compress these images over the repetitions
to form a set of five singular images. The single echo images from the first
singular image are used to generate T2* and B0 maps via a
dual exponential fit. The T2* and B0 maps are used to
solve for the initial water and fat contrasts at the beginning of rosette
acquisition for each singular image. The initial water and fat contrasts are
then used in the MRF pattern matching process to generate separate water and
fat T1 and T2 maps. The MRF dictionary was generated with
the following tissue property values: T1=[10:10:2000 2020:20:3000]
and T2=[6:2:100 105:5:300 310:10:500 540:40:1200], and was corrected
for slice profile and inversion efficiency effects12.
Linear
regression was performed on the reconstructed parameter maps of the ISMRM/NIST
phantom to determine the accuracy of the method. T2 values greater than 300 ms and T2* values greater than 150 ms were excluded from the regression. Results from the volunteers
were compared to the reference scans, as well as reported literature values3.Results
Reconstructed maps from the phantom experiment are shown in
Figure 2, along with regression plots comparing experimental T1, T2,
and T2* values to the reference values.
Representative maps from one volunteer are shown in Figure
3, and quantitative data from all volunteers are shown in Figure 4. Mean T1
values for this study were 420 ms, mean T2 values for this study
were 54.6 ms, and mean T2* values for this study were 41.0 ms. Rosette
MRF, GRE, and literature3 were in agreement on T2*
values.Discussion
Overall, the phantom experiments show
good agreement between the experimental and reference values in the ISMRM/NIST
MRI system phantom. The rosette MRF T2* values measured in the liver
agree with both the Cartesian GRE reference scans, as well as published
literature values for healthy volunteers at 0.55T. T1 values
measured with MRF were approximately 81 ms longer than those reported in the
literature for liver at 0.55T, and T2 values were 11 ms lower. The
differences between MRF measurements in this study and previous literature
values are similar to the differences seen at higher field strengths7,9,13,
and the variation between volunteers in this study is similar to the variation
seen in literature3.Conclusion
T2* maps can be collected
simultaneously with T1 and T2 using the rosette MRF
technique, even at lower field strengths.Acknowledgements
NSF/CBET 1553441, NIH/NHLBI R01HL094557, Siemens
HealthineersReferences
1. Sirlin CB, Reeder SB. Magnetic Resonance Imaging
Quantification of Liver Iron. Magn Reson Imaging Clin N Am 2010;
18(3):359-381
2. Triadyaksa P, Oudkerk M, Sijens PE. Cardiac T2*
Mapping: Techniques and Clinical Applications. J Magn Reson Imaging 2020;
52:1340-1351
3. Campbell-Washburn AE, Ramasawmy R, Resivo MC, et
al. Opportunities in Interventional and Diagnostic Imaging by Using
High-performance Low-Field-Strength MRI. Radiology 2019; 293(2):384-393
4. Campbell-Washburn AE, Mancini C, Conrey A, et
al. Evaluation of Hepatic Iron Overload Using a Contemporary 0.55T MRI System. Journal
Magn Reson Imaging. 2021;
5. Ma D, Gulani V, Seiberlich N, et al. Magnetic
Resonance Fingerprinting. Nature 2013; 495:187-192
6. Hamilton JI, Jiang Y, Chen Y, et al. MR
Fingerprinting for Rapid Quantification of Myocardial T1, T2, and Proton Spin
Density. Magn Reson Med 2017; 1446-1458
7. Jaubert O, Arrieta C; Cruz G, et al.
Multi-parametric liver tissue characterization using MR fingerprinting:
Simultaneous T1, T2, T2*, and fat fraction mapping. Magn Reson Med 2020;
84:2625-2635
8. Noll DC. Multishot Rosette Trajectories for
Spectrally Selective MR Imaging. IEEE Trans Med Imaging 1997;
16(4):372-377
9. Liu Y, Hamilton J, Eck B, et al. Myocardial T1
and T2 quantification and water-fat separation using cardiac MR fingerprinting
with rosette trajectories at 3T and 1.5T. Magn Reson Med 2021;
85(1):103-119
10. Bush AM, Sandino CM, Ramachandran S, et al.
Rosette Trajectories Enable Ungated, Motion-Robust, Simultaneous Cardiac and
Liver T2* Iron Assessment. J Magn Reson Imaging 2020; 52(6):1688-1698
11. Fessler JA, Sutton BP. Nonuniform fast Fourier
transforms using min-max interpolation. IEEE Trans Signal Processing. 2003;
51:560-574
12. Hamilton JI, Jiang Y, Ma D, et al. Investigating
and reducing the effects of confounding factors for robust T1 and T2 mapping
with cardiac MR fingerprinting. Magn Reson Imaging 2018; 53:40-51
13. Chen Y, Jiang Y, Pahwa S, et al. MR
Fingerprinting for Rapid Quantitative Abdominal Imaging. Radiology.
2016; 279(1):278-286