Kartiga Selvaganesan1, Yonghyun Ha2, Gigi Galiana2, and Todd Constable2
1Biomedical Engineering, Yale University, New Haven, CT, United States, 2Radiology and Biomedical Imaging, Yale University, New Haven, CT, United States
Synopsis
Gradient-free
imaging at low-fields can significantly reduce the cost of and increase access
to MRI devices. Here we propose to exploit the Bloch-Siegert shift effect to
perform gradient-free, RF spatial encoding at 24mT. We have developed a
simulation algorithm that designs various nonlinear encoding schemes and
evaluates their performance through image reconstruction. Our results indicate
that this technique is tolerant to noise and B0 inhomogeneities; this is an important
step towards demonstrating the feasibility of performing low-field imaging with
nonlinear RF spatial encoding using the Bloch-Siegert shift.
Introduction
The cost and
complexity of MR systems can be significantly reduced by moving towards lower
field strengths and eliminating gradients coils. Instead, spatial encoding can
be performed using RF coils. One way to perform RF spatial encoding is by using
the Bloch-Siegert shift on coils with spatially varying B1-fields, so the off-resonance
RF pulse produces a spatially varying phase shift1.
Previous work using this
method focused on imposing linear phase shifts which is difficult since RF
coils produce inherently nonlinear field patterns2. Therefore,
we have developed novel nonlinear encoding schemes for gradient free imaging
using the Bloch-Siegert shift at low-field (24mT). Here we present the design
and evaluation of these schemes through simulation studies.Methods
Figure 1
shows a schematic of the simulation workflow used in this study. From a
measured B0-field map, a nonplanar slice containing uniform magnetic field
strength (1MHz), was selected as the imaging slice (Figure 1A). Next, the
B1-field across the slice profile was calculated for each channel of a 3x3
planar RF transmit (Tx) array. Using these B1-fields,
each encoding pattern
was calculated using the following equation: $$B_1 = \sum_1^{c = 9} B_1^c \times A^c$$ where $$$A$$$
is the
amplitude and phase of the Bloch-Siegert off-resonance pulse applied to each channel
(Figure 1B). In this study we
evaluated encoding schemes created from unique combinations of 4, 6 or 8 Tx coils,
with a pulse phase of either 0 or
$$$\pi$$$. The phasor generated by each encoding pattern
was then calculated using Bloch simulations. Figure 1C shows examples of unwrapped
phasors resulting from transmitting the Bloch-Siegert pulse on 4 and 8 coils.
Finally, the MR signal generated by each encoding scheme was used to reconstruct
a grid and 3T liver phantom image using the conjugate gradient method for
nonlinear encoding fields3
(Figure 1D). Signals
were simulated for a spin echo sequence (Figure 1D) implemented with the
following parameters: 256 readout points per echo, 128 echoes,
20cmx20cm FOV, TE=16ms, and maximum Bloch-Siegert pulse length=5.88ms. Images were reconstructed over a matrix of
256x256.Results
Figure 2A
shows the results from reconstructing a digital grid phantom with data generated
from encoding schemes of Tx on 4, 6 and 8 coils. Encoding schemes were
optimized to improve reconstruction on a central ROI delineated by the blue
box. Figure 2B shows plots of the signal profiles along horizontal (purple) and
vertical (green) lines through the center of the FOV for each encoding scheme. The
plot in black is the line profile of the phantom image. These results
demonstrate that increasing the number of Bloch-Siegert transmit coils can help
improve the image resolution within the FOV (blue box).
The effect of adding
noise to the simulated MR signal prior to reconstruction is shown in Figure 3, with
reconstruction results from a liver image extracted from the true position of the
simulated slice. Gaussian white noise was added to the signal at SNR levels of
50dB, 25dB, and 10dB. An SNR of 50dB is comparable to noiseless signal, where
25 and 10dB are closer to what is achievable with the physical hardware. The
normalized root mean square error (RMSE) and structural similarity index (SSIM)
values for these reconstructed images compared to a corresponding 3T liver
image are listed in Figure 4.
Since the B1-field
strength of RF coils drop with depth, the spatial variation in Bloch-Siegert
encoding is to expected decrease with increasing distance from the array coil,
though this can be compensated with longer encoding pulses. Figure 5 shows the
reconstruction results of a liver image at a 1.4cm and 6.4cm depth, encoded
using a 5.88ms and 11.76ms Bloch-Siegert pulse. The results indicate that resolution
is recovered when the length of the Bloch-Siegert pulse is doubled.Discussion
These
simulation results support the feasibility of performing gradient-free,
nonlinear RF encoding using the Bloch-Siegert shift, on our low-field MRI
system. The simulation algorithm developed in this study allows us to robustly design
and assess different encoding schemes and evaluate their performance through MR
image reconstruction.
Overall, the results
showed that the quality of the reconstructed images improves by increasing the
number of Bloch-Siegert transmit coils. While the system is generally quite
tolerant to Gaussian noise, in the case where noise dominates the MR signal
(10dB), encoding schemes with Tx on 4 coils outperformed that of 6 and 8 coils.
This may be because schemes with fewer transmit coils create regions in the FOV
with slower phase windings which are then more tolerant to noise. The results
also illustrate that, though the SNR, resolution and contrast are variable
across the FOV, the variations are small, and the resulting images have significant
diagnostic potential. This characteristic is like that of other imaging methods
that incorporate nonlinear spatial encoding at higher fields4,5.
We showed that one
way to improve image quality is by increasing the duration of the Bloch-Siegert
pulse. To further improve image reconstruction, in the future we will implement
deep learning-based image reconstruction techniques6,7.
Conclusion
In this
study we have designed and numerically analyzed novel nonlinear RF encoding
schemes for gradient free imaging at low-fields. This technique reimagines
conventional MR, making it low-cost, noise-free, and highly accessible. Acknowledgements
No acknowledgement found.References
- Bloch, F. and A. Siegert, Magnetic Resonance for Nonrotating Fields. Physical Review, 1940. 57(6): p. 522-527.
-
Kartausch, R., et al., Spatial phase encoding exploiting the Bloch-Siegert shift effect.
MAGMA, 2014. 27(5): p. 363-71.
-
Li, S., et al., Algebraic
reconstruction technique for parallel imaging reconstruction of undersampled
radial data: application to cardiac cine. Magn Reson Med, 2015. 73(4): p. 1643-53.
-
Wang, H., et al., Experimental
O-space turbo spin echo imaging. Magnetic resonance in medicine, 2016. 75(4): p. 1654-1661.
-
Wang, H., et al., Fast rotary nonlinear spatial acquisition
(FRONSAC) imaging. Magn Reson Med, 2016. 75(3): p. 1154-65.
-
Koonjoo, N., et al., Boosting the signal-to-noise of low-field
MRI with deep learning image reconstruction. Scientific Reports, 2021. 11(1): p. 8248.
-
Zhu, B., et al., Image
reconstruction by domain-transform manifold learning. Nature, 2018. 555(7697): p. 487-492.