Dinesh K Deelchand1, Pierre-Gilles Henry1, James M Joers1, Edward Auerbach 1, Young Woo Park1, Firat Kara2, Eva Ratai3, Kejal Kantarci2, and Gülin Öz1
1University of Minnesota, Minneapolis, MN, United States, 2Mayo Clinic, Rochester, MN, United States, 3Massachusetts General Hospital, Charlestown, MA, United States
Synopsis
We
developed an automated advanced-single-voxel MRS acquisition protocol at 3T to
facilitate acquisition of high-quality spectroscopic data without local MRS
expertise. Voxel-based B0 and B1 calibrations were
incorporated into the consensus-recommended semi-LASER sequence and combined
with automated voxel prescription. Automated B0 and B1
calibrations saved ~4.5min per voxel vs. manual calibrations. All spectra
acquired with the automated protocol by rotating MR technologists were usable,
while only 83% of those collected with the manual protocol were usable and
spectral quality was more variable. The protocol allows automated acquisition
of high-quality MRS data with high success rate on a clinical 3T platform.
Introduction
Proton
magnetic resonance spectroscopy (1H MRS) allows non-invasive
measurements of brain biochemistry and up to 18 metabolites can be quantified
in vivo1. MRS complements structural MRI, which provides spatial
distribution of water in various tissues. Despite providing noninvasive access
to neurochemical changes in a multitude of central nervous system disorders2,
MRS has not found wide utility in the clinical setting due to challenges
associated with acquiring high-quality and reproducible spectroscopic data.
Reliable metabolite quantification from MRS data requires accurate localization
of the volume-of-interest (VOI), high signal-to-noise ratio, optimal
voxel-based B0 inhomogeneity correction, efficient water suppression
(WS) and minimal lipids/ artifacts in the spectrum3.
Advanced
MRS protocols improve data quality and reproducibility relative to
vendor-provided protocols4,5, however, are challenging to
incorporate into the clinical workflow and require local MRS expertise for
successful implementation6. Here, we developed an automated advanced
single-voxel MRS acquisition protocol at 3T to facilitate acquisition of
high-quality spectroscopic data without local MRS expertise.Methods
The
semi-LASER (sLASER) sequence was recently recommended by MRS experts’ consensus
at high fields to minimize CSD artifacts that are prominent with conventional
protocol. The compact sLASER sequence7 with interleaved adiabatic
refocusing, optimized gradient scheme and 3D outer-volume-suppression (OVS)
that was recently harmonized across vendors8 was selected for the
automated protocol.
First,
a B0 shimming protocol was selected for automation by comparing
three widely used B0 algorithms; two Siemens-provided protocols: Advanced
shim and “Brain” mode shim and FAST(EST)MAP (FM)9. Five healthy
participants were scanned at a 3T Prisma scanner and water signal was measured
from different six volume-of-interest (VOI) across the brain.
The
B1 transmit field for all radio-frequency pulses used for WS, OVS
and localization in the MRS sequence needs to be correctly set for efficient WS
and accurate localization. Voxel-based B1 adjustment consists of
arraying the flip angle (FA) (or transmit voltage) of the excitation pulse in
the localization sequence and measuring the water signal at each step.
These
voxel-based B0 and B1 calibrations were incorporated into
the sLASER sequence and combined with AutoVOI, a recently developed method for
automated voxel prescription. The complete automated advanced MRS protocol is shown
in Figure 1.
The
efficiency of collecting data from a clinical cohort (N=40) with the automated
protocol (calibration time and fraction of usable datasets) was compared with
the non-automated semi-LASER protocol (N=46) whereby all prescan calibrations
were executed manually in the academic hospital setting with rotating MR technologists
in the neuroradiology unit6. MRS data were collected from the
bilateral posterior cingulate cortex (8 mL) using sLASER and FM shimming on a
Siemens 3T.Results
A
multi-iteration FAST(EST)MAP protocol resulted in narrower water linewidths
than vendor’s B0 shim protocols for data acquired from six brain
locations (P<1e-5) and was
selected for automation (Figure 2). Since the water linewidths were narrowest
and within 10Hz for all VOIs with FM (P<1e-5
vs. vendor routines), FM was therefore incorporated into the automated MRS
protocol.
The
time taken for manual B0 and B1 calibrations ranged from
5 to 15.5min with a mean time of ~8min. In contrast, the time taken was
identical (~4min) for all subjects in the automated MRS protocol, demonstrating
an average gain of ~4.5min per VOI (P<1e-5)
compared to the manual protocol.
All
spectra acquired with the automated protocol were usable, while only 83% of
those collected with the manual protocol were usable (N=38). After exclusion of
unusable spectra, comparable spectral quality was observed between protocols
(Figure 3), with higher variation in water residual signal with the manual
protocol. Although similar water linewidths (mean±standard deviation, 6.1±2.0Hz
with manual vs. 6.7±0.5Hz with automated) were observed in PCC with both protocols,
the between-subject coefficients-of-variation was larger with the manual
protocol (33% with manual vs. 8% with automated protocol) due to higher variation
of linewidth in manual protocol.Discussion
This
study showed that an automated advanced MRS protocol can be implemented on
clinical scanners whereby both B0 and B1 adjustments are
executed without operator intervention. The protocol yields savings in data
acquisition time, increases the rate of successfully acquired datasets and
improves between-subject consistency in spectral quality compared to the same
protocol executed manually where the MR operator needs to set the 1st and/or
2nd order shim values after each FM iteration and also determine and input the optimal transmit voltage and WS-FA. This
is a very important step to facilitate streamlined MRS acquisition in research
and clinical settings.
A
limitation of the current study was that the automated protocol was only
evaluated for single-voxel MRS. However, the same automated principles can be
applied to multi-voxel MRS imaging10 where B0
and B1 adjustments would be carried out on a larger voxel or a whole
slice depending on the sequence utilized. An advantage of the proposed
automated protocol is that its implemented is not restricted to sLASER, but it
has potential to be implemented for other advanced MRS sequences at various
fields. Although, this automated approach was demonstrated on a Siemens scanner
here, this technique could be implemented on other vendors’ platform.Conclusion
The
plug-and-play advanced MRS protocol allows automated acquisition of
high-quality MRS data with high success rate and consistency on a clinical 3T
platform.Acknowledgements
This
work was supported by funding from the National Institutes of Health (NIH) (R01
NS080816, P41 EB027061, P30 NS076408). We would like to thank the MR
technologists at the Mayo Clinic for the MRS data acquisitions and all study
participants.References
-
Pfeuffer
J et al. Toward an in vivo neurochemical profile: quantification of 18
metabolites in short-echo-time 1H NMR spectra of the rat brain. J Magn Reson
1999;141(1):104-120.
- Öz
G et al. The MRS Consensus Group. Clinical Proton MR Spectroscopy in Central
Nervous System Disorders. Radiology 2014;270(3):658-679.
- Tkáč
I & Gruetter R. Methodology of 1H NMR spectroscopy of the human brain at
very high magnetic fields. Appl Magn Reson 2005;29(1):139-157.
- Terpstra
M, Cheong I, Lyu T, Deelchand DK, Emir UE, Bednařík P, Eberly LE, Öz G.
Test-retest reproducibility of neurochemical profiles with short-echo,
single-voxel MR spectroscopy at 3T and 7T. Magn Reson Med 2016;76(4):1083-1091.
- Deelchand
DK et al. Two-site reproducibility of cerebellar and brainstem neurochemical
profiles with short-echo, single-voxel MRS at 3T. Magn Reson Med
2015;73(5):1718-1725.
-
Deelchand
DK et al. Improved localization, spectral quality, and repeatability with
advanced MRS methodology in the clinical setting. Magn Reson Med
2018;79(3):1241-1250.
- Öz
G & Tkac I. Short-echo, single-shot, full-intensity proton magnetic
resonance spectroscopy for neurochemical profiling at 4 T: Validation in the
cerebellum and brainstem. Magn Reson Med 2011;65(4):901-910.
- Deelchand
DK et al. Across-vendor standardization of semi-LASER for single-voxel MRS at
3T. NMR in Biomedicine 2021;34(5):e4218.
- Gruetter
R & Tkac I. Field mapping without reference scan using asymmetric
echo-planar techniques. Magn Reson Med 2000;43(2):319-323.
- Maudsley
AA et al. Advanced magnetic resonance spectroscopic neuroimaging: Experts'
consensus recommendations. NMR in Biomedicine 2021;34(5):e4309.