Charlotte E Buchanan1, Hao Li2, David M Morris3, Alexander J Daniel1, João Sousa4, Steven Sourbron4, David L Thomas5,6,7, Andrew Nicholas Priest2,8, and Susan T Francis1
1Sir Peter Mansfield Imaging Centre, University of Nottingham, Nottingham, United Kingdom, 2Department of Radiology, University of Cambridge, Cambridge, United Kingdom, 3Centre for Cardiovascular Science, University of Edinburgh, Edinburgh, United Kingdom, 4Department of Infection, Immunity and Cardiovascular Disease, University of Sheffield, Sheffield, United Kingdom, 5Neuroradiological Academic Unit, UCL Queen Square Institute of Neurology, University College London, London, United Kingdom, 6Dementia Research Centre, UCL Queen Square Institute of Neurology, University College London, London, United Kingdom, 7Wellcome Centre for Human Neuroimaging, UCL Queen Square Institute of Neurology, University College London, London, United Kingdom, 8Department of Radiology, Addenbrooke’s Hospital, Cambridge, United Kingdom
Synopsis
Standardisation and multicentre
evaluation of renal MRI measures is crucial for clinical translation. Here we
present results of a Travelling kidney study on GE, Philips and Siemens using a harmonised multiparametric renal MRI protocol, and
show results of B0 and B1
mapping, T1, and T2* mapping. The mode B0
within the kidneys did not vary across vendors, however the FWHM was narrower
for Vendor 1. No significant differences were seen in MOLLI T1 values of the
renal cortex and medulla after this data was corrected for the offset in T1
seen in the NIST phantom.
INTRODUCTION
Harmonisation of MR protocols and standardisation
of measures is crucial for the multicentre evaluation of clinical translation
of renal MRI measures1. Here we perform a Travelling Kidney study to
evaluate the impact of vendor on a multiparametric renal 3T MRI acquisition
protocol that has been harmonised across vendors for field of view (FOV),
spatial resolution, slice orientation and ordering, sequence timings (repetition
time, echo time, inversion time, echo-spacing), bandwidth, flip angle,
fat-suppression scheme and flow compensation, and vendor specific acceleration factors
(ASSET, GRAPPA, SENSE).METHODS
A Travelling Kidney study was
performed using a multiparametric renal MRI protocol harmonised across three 3T
MR vendors (Philips, Siemens and GE) randomly assigned to Vendor 1, 2 and 3. Vendor
1 was a 60 cm bore scanner, and Vendors 2 and 3 were wide bore scanners with a
70 cm bore and one system had dual transmit capability. Data was acquired on
six healthy volunteers (age 32±6 yrs), four participants were scanned on each
of the three vendor scanners, two participants underwent scans on only two
vendors. The protocol comprised of B0 and B1 mapping to
characterise scanners, relaxometry (T1, T2 and T2*
mapping), phase contrast (PC-MRI) and volumetric T1- and T2-weighted
scans, and ASL scans. Results of B0 and B1 mapping, T1,
and T2* mapping will be reported here, all measures were collected
at end-expiration breathhold. B0 mapping used an in-phase
dual-echo mapping scheme with phase images saved to generate B0 maps,
B1+ mapping used the optimal
sequence from each vendor: DREAM2
for Philips, TurboFLASH B1 mapping3 for Siemens and
Bloch-Siegert4 for GE. T2*
and T1 mapping was collected using both a 5-slice sequence acquisition
with 1.5×1.5×5mm3 resolution, a single slice per breath-hold. The
gradient echo (GRE) data was collected in-phase (TE/DTE = 4.61ms) and
the T1 mapping
was acquired using a Modified Look-Locker Imaging (MOLLI) sequence5
with 5(3)3 acquisition scheme. The
ISMRM/NIST system phantom was used to evaluate the standardisation of T1
mapping, these phantom measures provided T1-mapping functions from which to
correct the in-vivo T1 measures. Data were analysed
using UKAT software6. For B0 and B1 analysis,
whole kidney ROIs were manually defined on the B0 magnitude images
and applied to both. For T1 and T2* analysis whole kidney
ROIs were manually defined and histogram analysis was applied to segment the
cortex from medulla7. RESULTS
Figure
1 shows histograms of B0 mapping, no significant difference in the
mode of B0 was seen between vendors, but there was a significant
difference in FWHM of the B0 histograms with Vendor 1 (the narrow
bore system) having a significantly narrower FWHM at p=0.01 and p=0.02 compared
to Vendors 2 and 3 respectively.
The B1 shown on Figure 2 was found
to significantly differ across scanners, with Vendor 1 having a significantly
higher B1 at 103.7±4.1% of the nominal flip angle compared to vendors 2
(93.8±4.3%, p=0.01) and Vendor 3 (85.2±4.6%, p=0.02).
Phantom results for the
MOLLI T1 mapping on the NIST phantom are shown in Figure 3a for each
of the spheres plot against their reference values, the error at high T1
spheres in the range of the medulla is apparent, particularly for Vendor 1. In-vivo
data is shown in Fig 3b, along with plots of the cortex and the medulla which is
shown with and without correction. An intra-subject CoV of 1.6±0.8% was found
for the cortex, and 3.2±1.0% for the medulla, which reduced to
1.2±1.0% after correction. Prior to correction, a difference was seen in the medulla T1
between Vendors 1 and 3 (p=0.002). After correction, no significant differences
were measured in T1 values across vendors.
Figure 4 shows the
results of the T2*mapping. T2* values had a larger CoV at
10.4±6.5% and 12.2±6.3% for the cortex and medulla.DISCUSSION AND CONCLUSION
Here
we characterise the B0 and B1 fields and assess renal
MOLLI T1 and T2* in a Travelling Kidney study across the
three main vendors. B1 maps are shown to vary across vendors/hardware
with Vendor 1 having a significantly higher B1. The mode B0
value within the kidneys did not vary across vendors, however the FWHM was
shown to be narrower for Vendor 1. The effects seen here in B0 and B1
are likely due to the system used for Vendor 1 being a narrow (60cm) bore
system.
No significant differences were seen in MOLLI T1 values of
the renal cortex and medulla after this data was corrected for the offset in T1
seen in the NIST phantom. In future, simulations will be performed of the MOLLI
RF and gradient waveforms used on each of the vendors. T2* data showed a higher CoV
than T1 measures, this could be due to the quality of the breath
hold acquisition for these measures and in future multiple repeats of these
measures will be performed. There was a
significant difference in T2* of the medulla between Vendors 1 and 2
(p=0.02). Acknowledgements
This work is
funded by MRC Partnership grant MR/R02264X/1.References
1. Mendichovszky I, Pullens P,
Dekkers I et al. Technical recommendations for clinical translation of renal
MRI: a consensus project of the Cooperation in Science and Technology Action
PARENCHIMA. Magn. Reson. Mater. Physics, Biol. Med. 2019.
2. Nehrke K, Börnert P. DREAM-a novel approach for robust, ultrafast,
multislice B1 mapping. Magn. Reson. Med. 2012;68:1517–1526.
3.
Klose U. Mapping of the Radio Frequency Magnetic Field With A Mr Snapshot Flash
Technique. Med. Phys. 1992.
4. Sacolick LI, Wiesinger F, Hancu I, Vogel MW. B1 mapping by
Bloch-Siegert shift. Magn. Reson. Med. 2010;63:1315–1322.
4.
Russek SE, Boss M, Jackson EF, et al. Characterization of NIST/ISMRM MRI
system phantom. In: Proc. Intl. Soc. Mag. Reson. Med. 20. Online; 2012 (p. 2456).
5. Messroghli DR, Radjenovic A, Kozerke S, et al.
Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping
of the heart. Magn. Reson. Med. 2004;52:141–6.
6. Daniel AJ, Nery F, et al. UKRIN
Kidney Analysis Toolbox (UKAT): A Framework for Harmonized Quantitative Renal
MRI Analysis in Proc. Intl. Soc. Mag. Reson. Med. 29 proc 3765
7. Cox EF, Buchanan CE, Bradley CR, et al.
Multiparametric renal magnetic resonance imaging: Validation, interventions,
and alterations in chronic kidney disease. Front. Physiol. 2017;8:1–15.