Victor Han1, Jianshu Chi1, and Chunlei Liu1,2
1Electrical Engineering and Computer Sciences, University of California, Berkeley, Berkeley, CA, United States, 2Helen Wills Neuroscience Institute, University of California, Berkeley, Berkeley, CA, United States
Synopsis
When multiple RF fields are applied at different frequencies, multiphoton excitation can occur when the sums or differences of integer multiples of these frequencies equal the Larmor frequency. No RF at the Larmor frequency is required. In this work, we describe the general principles of multiphoton pulsed selective excitation, providing a formalized treatment with design examples and implementations on a 3T scanner with an additional homemade z-direction (Bz) coil. With the additional Bz coil, we demonstrate additional flexibility, where the same excitation can be accomplished in several different ways with the same pulse duration.
Introduction
NMR excitation can occur not only at the Larmor frequency, but also at subharmonics of the Larmor frequency, a phenomenon called multiphoton excitation1–3. Although subharmonic excitation at high B0 fields is very inefficient, the efficiency can be greatly increased by using multiple frequencies to generate excitation instead of a single subharmonic. When multiple RF fields (B1 fields) are applied at different frequencies, multiphoton excitation can occur when the sums or differences of integer multiples of these frequencies equal the Larmor frequency4,5. With a proper choice of the RF frequencies, multiphoton excitation becomes practical in high-field MRI where SAR may be a concern.
In this work, we describe the general principles of multiphoton pulsed selective excitation, providing a formalized treatment with design examples and implementations on a 3T scanner with an additional homemade z-direction (Bz) coil. With the additional Bz coil, we demonstrate additional flexibility, where the same excitation can be accomplished in several different ways with the same pulse duration.Theory
With a small-tip angle approximation and ignoring relaxation, the Bloch equations can be written as
[˙Mx˙My]=γ[0Bz(t)−Bz(t)0][MxMy]+γ[−By(t)M0Bx(t)M0].[1]
Denote the transverse magnetization and magnetic field as
mxy=Mx+iMy,[2]
Bxy=Bx+iBy.[3]
For pulsed B fields over the time period from 0 to T, the solution to Eq. [1] is given by
mxy(r,T)=iγM0∫T0Bxy(r,t)e−iγ∫TtBz(r,τ)dτdt.[4]
In the Larmor frequency rotating frame, let us define
Bz=B1,z(t)cos(ωzt+ϕ)+G⋅r,[5]
Bxy=B1,xy(t)e−i((ωxy−ω0)t+θ(t)).[6]
These are the typical fields in MRI, except with the addition of a uniform RF field in the z-direction for multiphoton excitation. When the frequency of the xy-RF is far off from the Larmor frequency, but satisfies the multiphoton resonance condition ωxy−ω0=nωz6, Eq. [4] can be rewritten as
mxy(r,T)=iγM0∫T0B1,xy(t)e−i(nωzt+θ(t))e−iγ∫TtB1,z(τ)cos(ωzτ+ϕ)dτeik(t)⋅rdt,[7]
where −γ∫TtG(τ)dτ=k(t) as in excitation k-space7.
If B1,z(τ) is slowly varying compared to cos(ω1,zτ+ϕ), then the integral of their product is approximately the product of B1,z(τ) and the integral of cos(ω1,zτ+ϕ). With this assumption, and using the Jacobi-Anger expansion shown below, where Jm(−) is the Bessel function of the first kind of order m,
eiγB1,zωzsin(ωzt+ϕ)=Σ∞m=−∞Jm(γB1,zωz)eim(ωzt+ϕ),[8]
Eq. [7] can be rewritten as
mxy(r,T)≈iγM0∫T0B1,xy(t)e−i(nωzt+θ(t))(Σ∞m=−∞Jm(γB1,z(t)ωz)eim(ωzt+ϕ))e−iγB1,z(t)ωzsin(ωzT+ϕ)eik(t)⋅rdt.[9]
Only the term with m=n contributes significantly to the integral, giving
mxy(r,T)≈iγM0∫T0B1,xy(t)e−iθ(t)Jn(γB1,z(t)ωz)e−i(γB1,z(t)ωzsin(ωzT+ϕ)−nϕ)eik(t)⋅rdt.[10]
Eq. [10] shows that Bxy and Bz contribute to the excitation profile in a similar way with B1,xy(t) and Jn(γB1,z(t)ωz) available for amplitude modulation and e−iθ(t) and e−i(γB1,z(t)ωzsin(ωzT+ϕ)−nϕ) available for phase modulation. This contrasts with the standard one-photon case where we would just have mxy(r,T)≈iγM0∫T0B1,xy(t)e−iθ(t)eik(t)⋅rdt.Methods
To demonstrate the principles described in the theory, we simulated and implemented three sets of related pulses. To generate each pulse, the following procedure was followed.
1. Generate a prototype pulse using a conventional method like the SLR algorithm
8.
2. If designing a standard one-photon pulse, directly set
Bxy to the prototype pulse and finish.
3. Else if designing a multiphoton pulse, choose
ωz.
- Based on Eq. [10], choose values such that B1,xy(t)Jn(γB1,z(t)ωz) equals the amplitude modulation of the prototype pulse, and e−iθ(t)e−i(γB1,z(t)ωzsin(ωzT+ϕ)−nϕ) equals the frequency modulation of the prototype pulse. For multiphoton excitation, we have more variables to choose which can together achieve the same effects as in the one-photon case.
- Shift the center frequency of the Bxy pulse by nωz.
Base SLR prototype pulses were generated using SigPy.RF
9. See https://github.com/LiuCLab/multiphoton-selective-excitation for complete details on pulse generation.
ωz/(2π)=25 kHz for all experiments.
Results
Fig. 1 shows the setup using an additional Bz coil. Fig. 2 shows the simulations and experimental results of one-photon, two-photon, and frequency-modulated one-photon pulses producing the same slice selective excitation when designed to be equivalent. Frequency-modulated one-photon pulses are pulses which use e−iθ(t) to imitate the effects of a Bz pulse. Fig. 3 shows the two-photon pulse from Fig. 2, except shifted in position by three different methods. In Fig. 4, two-photon SLR pulses are demonstrated where the first pulse has amplitude modulation fully in the Bxy pulse, the second pulse has amplitude modulation fully in the Bz pulse, and the third pulse has amplitude modulation in both the Bxy and Bz pulses.
Using the same one-photon and two-photon pulses as in Fig. 2, Fig. 5 shows the in-plane results of the one-photon and two-photon pulses in vivo under our institution's IRB approval. No significant differences between the images are observed for this set of parameters.Discussion and Conclusions
When ωz is large enough, the distinction between the xy- and z-direction RF becomes smaller, and the ability to modulate Bz instead of Bxy gives extra flexibility to the multiphoton RF designer. We demonstrated how the same slice profiles can be achieved in many ways. Although not explored here, the principles of using a Bz field for selective excitation could be extended to the use of arrays of z-direction coils for improving excitation homogeneity or tailoring excitation in general. Especially for low-field scanners where SAR is less of a concern, more eclectic applications can also be envisioned. For example, the traditional xy-RF transmit chain could be simplified in favor of a lower frequency z-RF. Alternatively, since the multiphoton pulses do not have any RF at the Larmor frequency, a pulsed version of simultaneous transmit and receive like in10,11 could be implemented.Acknowledgements
The authors thank Ekin Karasan and Miki Lustig for an introduction to HeartVista, Anita Flynn for improving lab spaces, and Karthik Gopalan for help and advice with mechanical engineering. This work was supported in part by NIH grant R21EB030157.References
1. Eles PT, Michal CA. Two-photon excitation in nuclear magnetic and quadrupole resonance. Progress in Nuclear Magnetic Resonance Spectroscopy. 2010;56(3):232-246. doi:10.1016/j.pnmrs.2009.12.002
2. Michal CA. Nuclear magnetic resonance noise spectroscopy using two-photon excitation. The Journal of Chemical Physics. 2003;118(8):3451-3454. doi:10.1063/1.1553758
3. Abragam A. The Principles of Nuclear Magnetism. Clarendon Press; 1961.
4. Eles PT, Michal CA. Two-photon two-color nuclear magnetic resonance. The Journal of Chemical Physics. 2004;121(20):10167-10173. doi:10.1063/1.1808697
5. Zur Y, Levitt MH, Vega S. Multiphoton NMR spectroscopy on a spin system with I =1/2. The Journal of Chemical Physics. 1983;78(9):5293-5310. doi:10.1063/1.445483
6. Han V, Liu C. Multiphoton magnetic resonance in imaging: A classical description and implementation. Magnetic Resonance in Medicine. 2020;84(3):1184-119
7. doi:https://doi.org/10.1002/mrm.281867. Pauly J, Nishimura D, Macovski A. A k-space analysis of small-tip-angle excitation. Journal of Magnetic Resonance (1969). 1989;81(1):43-56. doi:10.1016/0022-2364(89)90265-5
8. Pauly J, Roux PL, Nishimura D, Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging). IEEE Transactions on Medical Imaging. 1991;10(1):53-65. doi:10.1109/42.75611
9. Martin J, Ong F, Ma J, Tamir J, Lustig M, Grissom W. SigPy.RF: Comprehensive Open-Source RF Pulse Design Tools for Reproducible Research. Proceedings of the ISMRM Annual Meeting and Exhibition. Published online 2020:1045.
10. Brunner DO, Pavan M, Dietrich B, Rothmund D, Heller A, Pruessmann K. Sideband Excitation for Concurrent RF Transmission and Reception. Proceedings of the ISMRM Annual Meeting and Exhibition. Published online 2011:625.
11. Brunner DO, Dietrich BE, Pavan M, Prüssmann KP. MRI with Sideband Excitation: Application to Continuous SWIFT. Proceedings of the ISMRM Annual Meeting and Exhibition. Published online 2012:150.