Woowon Lee1, Emily Y Miller2, Callan M Luetkemeyer1, and Corey P Neu1,2
1Paul M. Rady Department of Mechanical Engineering, University of Colorado, Boulder, CO, United States, 2Biomedical Engineering Program, University of Colorado, Boulder, CO, United States
Synopsis
Osteoarthritis
(OA) is a degenerative process in cartilage mainly occurring in knee. OA
affects the structure and function of cartilage thus, measuring the mechanical
properties of cartilage is crucial. We use spiral DENSE MRI to quantify the
mechanics of intact and defected bovine articular cartilage. The samples underwent
cyclic compressive load and 27 frames are captured while the joint was
compressed. Displacement and strain gradually increased over time and the
defected cartilage showed higher strain in the tibiofemoral contact areas. We
also found creep response in the defected cartilage which may be a potential
biomarker for OA detection.
INTRODUCTION:
OA
is a degenerative joint disease and a severe medical and socioeconomic burden afflicting
more than 10% of adults over 60 worldwide 1. Currently,
OA can only be detected in the late stage using techniques like radiography since
the pathogenesis of cartilage degeneration is largely unknown. Knowledge of how
the mechanical properties of cartilage alter and how daily activities such as
walking impact the cartilage are lacking. To address this issue, MR-based imaging
methods have been advanced to investigate the structure-function relationship
of soft tissue. We previously used displacement encoding with stimulated echoes (DENSE) MRI to measure in vivo
cartilage deformation under compressive load 2. In an effort to enhance the
temporal resolution, we explored the use of spiral scanning on DENSE MRI 3 and collected multi-frame images
on intact and defected bovine joints during loading. This method allows high temporal resolution
(40ms) imaging, which (unlike previous methods)
facilitates the measurement of viscoelastic parameters. METHODS:
Specimen Preparation: Bovine joints (1-month-old, n=3) were obtained from an abattoir
and stored in a cold room (4°c) for 48 hours. We dissected the external tissue and
mounted both tibia and femur to mechanical testing grips using bone cement. The
sample was placed in the compressive loading device (Fig. 1A) within a clinical
MRI system (3T; Siemens Prismafit). In one intact joint, a defect (~1cm diameter) was generated on the medial condyle.
Spiral DENSE MRI with Exogenous Cyclic
Loading: Bovine joints underwent cyclic
loading (pneumatic;
100-150N; 0.5× body weight)
along the superior-inferior axis on the tibia and an ECG trigger initiated the
imaging. The loading cycle was composed of compression (1s), retraction (1s) and
delay (0.1s) delay before compression. The compressive load reached a plateau
at t=0.52s (frame 13). Total 27 images (frames) were acquired within the
loading time frame. Echo time (TE) and relaxation time (TR) were 2.5 and 20ms,
respectively. Ten interleaves were used to reconstruct each image. Field of
view was 125×125µm2, the number of pixels were 350×350, and the
slice thickness was 1.7mm. Images were collected in the coronal plane and
averaged 8 times for a high image contrast. The displacement encoding gradient
was 0.32cycles/mm. Prior to image acquisition, cyclic loading was applied for
15 minutes to achieve quasi-steady-state response 4 and double echo
steady state (DESS) images were collected for segmentation and to confirm that
joint tissues were responding to load (Fig. 1A).
Displacement and strain calculation:
Cartilage regions of interests (ROIs) were manually
segmented using custom software (MATLAB) and binary masks were created to
represent the articular cartilage. Displacements for each pixel within the ROIs
were determined from phase data as previously described 5.
Displacements then went through locally weighted scatterplot smoothing (LOWESS)
filtering for 100 cycles for a span=10. Subsequently, the displacement values were
used to calculate in-plane Green-Lagrange strains.
Viscoelasticity measurement: The
displacement values were smoothed in time using a Gaussian filter. We fit a natural logarithm curve to the strain-time
response during the constant loading time period (frame 13-27). Ecreep was calculated from the fitted curve which was
defined as the normal strain Eyy at frame 13 (Eelastic)
subtracted from Eyy at frame 27 (Eviscous).
The analysis was conducted on the tibiofemoral contact areas for the intact joints
and defected joint.RESULTS:
Displacement
and strain maps were acquired on 27 time-lapse images during loading (Fig. 1B).
Displacement and strain maps are close to 0 at t=0ms and evolve over time, demonstrating
a heterogeneous pattern across cartilage on both intact and defected joints (Fig.
2). We further analyzed displacement and strain in 12 sections (Fig. 3A). All
sections show a gradual increase in displacement. Displacement in y corresponding
to the loading direction, overall is higher than x. Displacement y
in the tibia is significantly higher than in the femur, shown as separate lines
on Fig. 3B. As seen in previous work, both compression and shear occurred
between the tibia and the femur 2. The tibia contact sections have
the highest Eyy (-0.12) whereas the noncontact sections in the
femur show minimal Eyy (-0.003) (Fig. 3C). The defected joint has higher
normal and shear strain values overall compared to the intact joints. Average
creep strain (Ecreep),
was -0.057 on the defected joint, while the intact joint was -0.019 (Fig. 4).DISCUSSION:
This
study is motivated by the hypothesis that the viscoelastic response of
articular cartilage may be useful biomarkers for early OA diagnosis in addition
to mechanical strain. We collect spiral DENSE MRI images during compressive
loading with high temporal resolution. This approach successfully quantifies
the mechanical response of cartilage under compression while the loading
frequency is comparable with walking cadence. The maximum strain in the defected
joint is significantly higher than the intact joint, indicating compressive
load is more detrimental to defected cartilage. We also note that the
viscoelastic nature of cartilage changes after damage. The viscoelasticity of
cartilage may be useful for distinguishing healthy from diseased tissue 6. However, measuring
viscoelasticity requires special methods regarding data acquisition in time and
is challenging to apply in a clinical setting. This approach not only measures
the viscoelastic response in physiologically relevant motion, but also can be utilized
in in vivo settings.Acknowledgements
The
authors would like to acknowledge funding from NIH R01 AR063712-08.References
1. WHO Department of Chronic
Diseases and Health Promotion. Available at: http://www.who.int/chp/topics/rheumatic/en/.
2. Chan
DD, Cai L, Butz KD, Trippel SB, Nauman EA, Neu CP. In vivo articular cartilage
deformation: Noninvasive quantification of intratissue strain during joint
contact in the human knee. Sci Rep. 2016;6:1-14.
3. Zhong
X, Meyer CH, Schlesinger DJ, et al. Tracking brain motion during the cardiac
cycle using spiral cine-DENSE MRI. Med Phys. 2009;36(8):3413-3419.
4. Neu CP, Hull ML. Toward an MRI-based method to measure
non-uniform cartilage deformation: An MRI-cyclic
loading
apparatus system and steady-state cyclic displacement of articular cartilage
under compressive loading. J Biomech Eng. 2003;125(2):180-188.
5. Chan DD,
Neu CP. Intervertebral disc internal deformation measured by displacements
under applied loading with MRI at 3T. Magn Reson Med. 2014;71(3):1231-1237.
6. Seidenstuecker
M, Watrinet J, Bernstein A, et al. Viscoelasticity and histology of the human
cartilage in healthy and degenerated conditions of the knee. J Orthop Surg
Res. 2019;14(1):256.