Jia Guo1,2
1Bioengineering, University of California Riverside, Riverside, CA, United States, 2Center for Advanced Neuroimaging, University of California Riverside, Riverside, CA, United States
Synopsis
Recently velocity-selective saturation (VSS) based dual-module
(dm-) preparation and velocity-selective inversion (VSI) based preparation have
demonstrated improved SNR efficiency in velocity-selective arterial spin
labeling (VSASL) than conventional single-module VSS based preparation. In this
study, a novel strategy was developed to allow dm-VSI labeling to further
improve the SNR efficiency. Implementation of dm-VSASL were evaluated for VSS
and/or VSI combinations. The theoretical performance was compared via ASL
signal modeling, and then validated by in vivo human experiments with an increase
of 6.6%. Dm-VSASL is capable of providing further improved SNR performance, with
potentially better suppression of diffusion attenuation/eddy current artifacts.
Introduction
Velocity-selective ASL (VSASL) 1 is insensitive to inhomogeneous
arterial transit times (ATT) among ASL methods 1, 2,
offering an SNR advantage when ATT is long. VS saturation (VSS) based labeling has
relatively low signal-to-noise ratio (SNR). Recent development on improving the
SNR includes: 1) multi-module VSS preparation 2, and 2) VS inversion (VSI)
preparation 3.
An SNR improvement in the range of 20 ~ 30% compared to single-module (sm-) VSS
based preparation has been reported using both methods 2-4.
A novel labeling strategy is invented to enable dual-module (dm-) VSI labeling to
further improve the SNR efficiency, which also benefits VSS labeling. Theory and Methods
Similar to the dm-VSS preparation 2, two groups of spins are considered
for dm-VSI: 1) one labeled by only the first VS module; 2) one labeled by both VS
modules. The evolution of the magnetization for dual-VSS and dual-VSI is shown
in Figure 1, and SigASL=ΔMz1·TI1.+ΔMz2·TI2.
Dm-VSI
labeling: VSI pulses invert the stationary spins, therefore flipping the label/control
condition in the second VSI is required to avoid signal canceling. Additional
benefits include: 1) the sensitivity to motion, diffusion attenuation and eddy
current (EC) effects are better balanced in the label/control images,
potentially reducing artefacts and improving quantification accuracy; 2) the
inversion effects allow more efficient background suppression (BS).
Dm-VSS labeling: 1) VSS+VSS as described previously 2; 2) VSSinv+VSS or VSSinv+VSSinv
to allow a more balanced label/control design as described above. VSSinv
can be implemented in several ways 2, 4,
such as adding a phase of π to the last segment of the symmetric BIR-8 (sBIR8) pulse.
Dm-VSASL using VSS and VSI: VSS+VSI, VSI+VSS and their inverted
variants. Label/control condition should be flipped in the second VS module when
necessary.
Modeling and simulation
A kinetic ASL model 5 was used to model the ASL
signals, including Pulsed ASL (PASL) and Pseudo-Continuous ASL (PCASL) 6.
The SNR efficiency, SigASL/sqrt(2TR), was evaluated with and
without T2 relaxation for VSASL, and realistic labeling efficiencies
for PASL and PCASL 7.
The maximal bolus durations were: 2s for VSASL 2, 1s for PASL and
unlimited for PCASL.
In vivo experiments
Four healthy human subjects (1F, age 23-38) were studied on
a
3T scanner (Siemens Prisma, Erlangen, Germany) under an IRB approval. Symmetric BIR-8 8 and sinc-VSI 4 pulses were used for VSS and
VSI labeling, respectively. ASL scans were acquired in a randomized order with
BS: 1) PASL: FAIR 9, 10
Q2Tips 11,
TI1=0.8s, TI=2.4s, 2 BS pulses; 2) VSSinv: TI=1.4s, 1 BS
pulses; 3) VSS+VSSinv: TI1/2=1.16/0.83s, 1 BS pulses; 4)
VSSinv+VSS: TI1/2=1.45/0.54s, 1 BS pulses; 5) VSI: TI=1.4s,
1 BS pulses; 6) VSI+VSI: TI1/2=1.45/0.54s, 2 BS pulses. Other parameters
were: 2-interleaved 3D GRASE EPI readout, FOV=220x220mm (64x64), 24 slices, 4mm
thickness, TR=4s (PASL) and 5s (VSASL), 15 and 12 label/control pairs for PASL
and VSASL, cutoff velocity=2cm/s along S/I in VSASL. Reference images were
acquired for quantification.Results
SNR efficiencies of different ASL methods from simulation are
shown in Figure 2. Compared to
sm-VSI, the maximal SNR efficiency of dm-VSI was increased by 11.8% without T2
relaxation; and was the same with T2 relaxation.
Figure 3a
demonstrates the necessity of the label/control flipping in the second VS
module in VSSinv+VSS and VSI+VSI. Figure 3b shows ATT artefacts with PASL in an early testing in sub.
3, while VSASL demonstrated excellent ATT insensitivity.
Normalized ASL signal and cerebral blood flow
(CBF) maps are shown in Figure 4. Averaged
values in gray matter (GM) and white matter (WM) are summarized in Table 1. After correcting for the signal
attenuation from BS (5% reduction each), dm-VSI produced the highest signal in
GM, followed by sm-VSI, and VSS+ VSSinv. Dm-VSI showed a trend of
increasing the signal by 6.6% (p=0.062) compared to sm-VSI labeling; VSS+VSSinv
increased ASL signal by 24.4% (p=0.025) compared to sm-VSS, consistent with
previous findings 2, 4; VSSinv+VSS showed a marginal
increase of 2.4% (p=0.91). Dm-VSI and sm-VSI increased ASL signal by 50% (p≤0.01) compared to sm-VSS. There was no
significant difference in CBF between labeling methods (p=0.19 in GM, and
p=0.070 in WM). In addition, the temporal SNR was doubled with dm-VSASL in GM (details reported in another abstract). A regional ASL signal reduction was
observed in sub. 4 using VSI labeling (Figure
4).Discussion
For dm-VSI vs. sm-VSI, in vivo experiments measured a signal
increase of 6.6%, higher than that predicted by simulation, likely due to more
stable labeling with dual-module labeling. Sinc-VSI is preferred to rect-VSI
because of a shorter effective TE (eTE, 29.4ms vs. 37.6ms using the same
timings in the study). Further reduction of eTE should increase the SNR
advantage of dm-VSASL.
It is important to arrange VS modules properly to
accumulate ASL signals constructively. It is required to flip the label/control
condition in the second VS module if the first one inverts stationary spins.
This can help reduced diffusion attenuation/EC artifacts. Consistent with the
findings in a recent study 4, further reduction of the sensitivity to field
inhomogeneities is needed for VSI pulses.Conclusion
Dual-module
VSS and VSI can be implemented with improved SNR efficiency and/or temporal SNR
with potentially reduced artefacts, may be a preferred VSASL labeling strategy for
clinical applications.Acknowledgements
The authors thank Dr. Jason Langley for data acquisition, and
Dr. Divya Bolar for sharing code for pulse sequence development.
References
1. Wong
EC, Cronin M, Wu W-C, Inglis B, Frank LR, Liu TT. Velocity-selective arterial
spin labeling. Magn Reson Med. 2006;55(6):1334-41.
2. Guo J, Wong EC. Increased SNR
efficiency in velocity selective arterial spin labeling using multiple velocity
selective saturation modules (mm-VSASL). Magn Reson Med. 2015;74(3):694-705.
3. Qin Q, van Zijl PC.
Velocity-selective-inversion prepared arterial spin labeling. Magn Reson Med.
2016;76(4):1136-48.
4. Guo J, Das S, Hernandez-Garcia L.
Comparison of velocity-selective arterial spin labeling schemes. Magn Reson
Med. 2021;85(4):2027-39.
5. Buxton RB, Frank LR, Wong EC, Siewert
B, Warach S, Edelman RR. A general kinetic model for quantitative perfusion
imaging with arterial spin labeling. Magn Reson Med. 1998;40:383-96.
6. Dai WY, Garcia D, de Bazelaire C,
Alsop DC. Continuous Flow-Driven Inversion for Arterial Spin Labeling Using
Pulsed Radio Frequency and Gradient Fields. Magn Reson Med. 2008;60(6):1488-97.
7. Alsop DC, Detre JA, Golay X, Gunther
M, Hendrikse J, Hernandez-Garcia L, et al. Recommended implementation of
arterial spin-labeled perfusion MRI for clinical applications: A consensus of
the ISMRM perfusion study group and the European consortium for ASL in
dementia. Magn Reson Med. 2015;73(1):102-16.
8. Guo J, Meakin JA, Jezzard P, Wong EC.
An optimized design to reduce eddy current sensitivity in velocity-selective
arterial spin labeling using symmetric BIR-8 pulses. Magn Reson Med.
2015;73(3):1085-94.
9. Kim SG. Quantification of relative
cerebral blood flow change by flow-sensitive alternating inversion recovery
(FAIR) technique: application to functional mapping. Magn Reson Med.
1995;34(3):293-301.
10. Kwong KK, Chesler DA, Weisskoff RM,
Donahue KM, Davis TL, Ostergaard L, et al. MR perfusion studies with
T1-weighted echo planar imaging. Magn Reson Med. 1995;34(6):878-87.
11. Luh
WM, Wong EC, Bandettini PA, Hyde JS. QUIPSS II with thin-slice TI1 periodic
saturation: a method for improving accuracy of quantitative perfusion imaging
using pulsed arterial spin labeling. Magn Reson Med. 1999;41(6):1246-54.