Vasiliki Mallikourti1, James Ross1, Oliver Maier2, Markus Bödenler3, Rudolf Stollberger 4, Lionel Broche1, and Mary-Joan MacLeod5
1Aberdeen Biomedical Imaging Centre, University of Aberdeen, Aberdeen, United Kingdom, 2Institute of Medical Engineering, Graz University of Technology, Graz, Austria, 3Institute of eHealth, University of Applied Sciences FH JOANNEUM, Graz, Austria, 4BioTechMed-Graz, Graz, Austria, 5Institute of Medical Sciences, University of Aberdeen, Aberdeen, United Kingdom
Synopsis
Field-Cycling
imaging (FCI) is a new imaging technique able to image over a range of low magnetic
field levels by rapid switching, to explore the variations in T1 relaxation time as a function of the magnetic field
strength, known as T1 dispersion. Here we measured the T1 dispersion contrast from
patients with ischaemic stroke. The infarct region in T1 maps corresponded to
clinical images. T1 dispersions had different profiles between stroke areas and
healthy brain areas, showing that ischaemic strokes can be imaged below 200 mT
without contrast agents.
Introduction
Computed tomography in acute stroke is
a ‘blunt’ tool that gives limited information in early infarction while conventional
magnetic resonance imaging is of limited practical use due to access, time and
safety issues. New approaches are being explored using low-field devices but
stroke contrast in such regime is poorly known.
Field-Cycling
imaging1,2 (FCI) is a novel modality that can image over a range of
magnetic field strengths through rapid switching between magnetic field levels.
This allows measurement of the field-dependent changes of the longitudinal T1
relaxation time, known as T1 dispersion. T1 dispersion provides
information on molecular dynamics exploiting novel biomarkers that are
invisible to conventional MRI. A whole-body field cycling system2
has been developed and approved for clinical studies in Aberdeen, allowing
multi-field T1 mapping at any field from 0.2 T to 0.2 mT. Here we report the results of the PUFFINS (Potential Use of
Fast Filed Cycling MRI IN Stroke) study that aims to assess whether FCI can
identify sub-acute strokes.Methods
35 patients
gave informed consent to participate in the PUFFINS study from 02/2018 to 06/2021
(ethics approved by NoSREC, number 16/NS/0136). Patients admitted with ischaemic
stroke were scanned on the whole-body FCI system within 24 to 96 h after presentation,
with a second scan 30 days later. Duration of the FCI examination was 45
minutes, including setup, axial and sagittal scout images and FCI images. FCI images
were acquired using a field-cycled inversion recovery spin echo sequence (Figure
1) with four to six evolution fields typically ranging from 0.2 T to 0.2 mT, five
evolution times (see Table 1), in-plane resolution of 2 to 4 mm, FOV of 290 mm,
slice thickness of 10 mm, TE of 16 to 24 ms, one slice and no averaging. As
part of their treatment, patients had a clinical CT scan, some with a 3T MRI
scan, acquired within 24 hours of the FCI images that we used for comparison.
Post-processing
was done in Matlab (Mathworks, 2019a, Natick, USA) using in-house software3.
Multi-field T1 quantification was performed in Python using a joint total
generalised variation (TGV) regularisation in conjunction with an iteratively regularised Gauss-Newton approach4,5.
T1 dispersions were extracted from three ROIs selected manually over grey
matter, white matter and infarct regions, using the clinical CT and MRI images
for validation. The stroke-to-brain T1 contrast across evolution fields was
calculated as6
C = (T1stroke-T1normal)/T1normal
where T1stroke
is the T1 value of the infarct region and T1normal is the mean T1
value of white and grey matter. Results
Ten FCI
scans were included in the study from eight patients. Five patients did not
complete the scan due to claustrophobia and/or unsuitable body habitus, while
in fifteen patients there were technical issues. In five patients the
single-slice scan missed small areas of infarction. Four patients completed
both scans but only two follow up scans were included due to technical issues.
The infarct
region measured by T1 maps corresponded with the CT and 3T MRI images, when available
(Figure 2). The T1 maps exhibited visible T1 contrast below 200 mT (Figure 2). The
in-vivo T1 dispersion profiles obtained from the T1 images showed profiles for
the stroke area that differed from other brain tissues in the low-field regime
(below 200 mT) (Figure 3). The T1 contrast increased markedly as the evolution
magnetic strength field decreased, reaching a maximum at 2 to 6 mT (Figure 4).Discussion
This is the
first-ever measurement of T1 dispersion in patients with ischaemic stroke.
The results showed that significant contrast is available below 200 mT in
patients with stroke, which discriminates between healthy and ischaemic brain
regions but also shows the limitations in T1 contrast at 200 mT. These changes
may reflect modifications of water diffusion across cell membranes due to cellular
adaptation or death7. Conclusions
We have demonstrated
the potential of low fields in ischaemic stroke and created a new tool to
understand the pathophysiology of diseases, in vivo. We hope these results will
help to better understand disease mechanisms in stroke and to guide the design
process of dedicated low-field devices. This first ever FCI in vivo clinical
study greatly helped us to improve the technology and we are currently starting
the second stage of the PUFFINS study to explore potential biomarkers in FCI
for the differentiation of stroke types and mimics.Acknowledgements
This project has received funding from the European Union’s
Horizon 2020 research and innovation programme under grant agreement No 668119
(project “IDentIFY”). We acknowledge the
support of the Scottish Stroke Research Network and Chief Scientist Office
(CSO).References
1.
Lurie, D. J. et al. Fast
field-cycling magnetic resonance imaging. Cr Phys. 2010;11, 136-148.
2.
Broche LM, Ross PJ, Davies GR, MacLeod MJ,
Lurie DJ. A whole-body Fast Field-Cycling scanner for clinical molecular
imaging studies. Sci. Rep. 2019;9(1):10402.
3.
Broche LM, Ross PJ, Davies GR,
Lurie DJ. Simple algorithm for the correction of MRI image artefacts due to
random phase fluctuations. Magn. Reson. Imaging. 2017;44:55-59.
4.
Bödenler M, Maier O, Stollberger R, Broche LM, Ross PJ, MacLeod MJ, Scharfetter H. Joint multi-field T1 quantification for fast field-cycling MRI.
Magn. Reson. Med. 2021; 86:2049–2063.
5. Maier O, Spann SM, Bödenler M,
Stollberger R. PyQMRI: An accelerated Python based
Quantitative MRI toolbox. Open Source
Softw. 2020;5: 2727.
6. Edelman R, Leloudas N, Pang J, Bailes J,
Merrell R, Koktzoglou I. Twofold improved tumor-to-brain contrast using a novel
T1 relaxation-enhanced steady-state (T1RESS) MRI technique. Sci Adv. 2020;
28:6(44):eabd1635.
7. Ruggiero MR, Baroni S, Pezzana S, Ferrante G,
Geninatti Crich S, Aime S. Evidence for the Role of Intracellular Water
Lifetime as a Tumour Biomarker Obtained by In Vivo Field-Cycling
Relaxometry. Angew Chem Int Ed Engl. 2018;57(25):7468-7472.