Guodong Weng1,2, Johannes Slotboom1,2, and Piotr Radojewski1,2
1Institute for Diagnostic and Interventional Neuroradiology, Support Center for Advanced Neuroimaging (SCAN), University of Bern, Bern, Switzerland, 2Translational Imaging Center, sitem-insel AG, Bern, Switzerland
Synopsis
Purpose: comparison
of SLOW-editing and MEGA-editing for detection of 2HG in glioma-suspected
patients at 7T.
Methods:
EPSI-based B0/B0+ robust SLOW-editing and
semiLASER-SVS/CSI-based MEGA-editing were applied in 10 patients. The edited
2HG signal at 4.01 ppm aimed to be detected with TE = 68 – 75 ms.
Results:
8 of 9 patients were identified successfully using SLOW-editing which has a
higher spectral quality and success rate compared to MEGA-semiLASER.
semiLASER-SVS/CSI-based MEGA-editing showed strong ghosting artifacts around
4.01ppm making 2HG-identification impossible.
Conclusions:
in vivo 2HG-editing can be performed using SLOW-EPSI within 10 minutes
measurement time and is the preferrable method at UHF.
INTRODUCTION
2HG
is a biochemical compound which accumulates and can be detected in IDH-mutated gliomas
1. IDH-mutations lead to a distinct
phenotype of gliomas and therefore can be used as a prognostic marker. The detection
of 2HG has direct clinical consequences for further neurosurgical-treatment (e.g.,
maximal resection versus of tumour-debulking). Since 2HG is a J-coupled spin-system
which has multiple resonances hidden by other stronger resonances, multiple
MEGA-based J-difference spectral-editing techniques have been proposed (e.g. 1–3) for
accurate measurement of 2HG. An alternative J-difference editing technique (named
SLOW), which is especially suited for use at UHF, has been proposed recently4. SLOW-editing requires only 8% of the SAR when
comparing to semiLASER-based MEGA-editing5,6,7.
This is because SLOW-editing only requires two different pairs of narrow-band
chemical selective adiabatic phase-compensated 2π-refocussing/editing
pulses (2π-CSAP), instead of four broadband slice selective phase
compensated 2π-refocussing pulses (2π-SSAP) in addition to two
narrowband MEGA-editing pulses. The high SAR results
in a long acquisition time. The low SAR advantage of SLOW allows for a short TR,
and, when combined with an EPSI read-out, a large volume of the brain can be
covered in less 10 minutes.
In this abstract we investigated the 2HG editing-performance in
vivo: MEGA-semiLASER was compared to SLOW-EPSI on our 7T clinical scanner. METHODS
All
MRI, MRS(I) measurements were performed on a Siemens MAGNETOM Terra 7T MR-scanner
in clinical-mode using a Nova 1Tx32Rx head-coil. Up till now, 10 patients with
suspected glioma have been prospectively examed. Currently,
surgery and neuropathology to determine the IDH-status have been performed in 9
out of 10 patients.
The
SLOW-EPSI 4,8,9 (Figure
1A-B) was applied on (i.) the “Braino”-phantom (General-Electric, USA), (ii.)
a spherical 2HG-phantom prepared in-house, (iii.) and 10 patients. The
refocusing/editing chemical-selective adiabatic pulse (2π-CSAP) for
SLOW-editing is 24 ms duration. The bandwidth (BW) of editing-full and
editing-partial ranges from 1.6–4.2 ppm and 2.7–4.2 ppm, respectively (Figure
2A). The editing result was obtained by the subtraction of the editing-full
(acquisition #1) by the editing-partial signal (acquisition #2).
MEGA-semiLASER-SVS
and CSI 5,10 were
additionally performed on all 10 patients during the same examination.RESULTS
Figure 2A shows the pulse profiles
of editing-full (1.6–4.2 ppm) and editing-partial (2.7–4.2 ppm). The
corresponding in vitro measurement for 2HG was shown in Figure 2B. Figure 2C-E
shows the simulation of Glx (glutamate/glutamine=2/1) at 3.76 ppm, GABA at
3.01 ppm and 2HG at 4.01 ppm using SLOW-editing.
Figure 3 shows three typical in
vivo measurements (spectral editing difference) in three IDH-mutated patients.
The 2HG-signal at 4.01 ppm is clearly present in all three patients using SLOW-EPSI
(indicated by red arrows). However, only one case (subject 3) shows two cases
(subject 1 and 3) show unambiguous 2HG signal for MEGA-SVS at 7T (indicated by
red arrows). The MEGA-based-CSI failed to detect 2HG at 7T, respectively.
Subjects 1 and 2 were confirmed to have IDH-mutated glioma by histopathology, while subject 3 has not had surgery yet.
Figure 4 shows the corresponding
MRI and together with VOI of MRSI/MRS.
Figure
5A shows raw metabolic maps of Glx, NAA, Cho, Cr which are non B1+
and B1- inhomogeneity corrected. Figure 5B shows
the same maps but B1+ and B1-
inhomogeneity corrected using the water reference map acquired simultaneously
by the EPSI sequence8. Figure 5C shows the metabolic maps of Glx
(edited), Cho, Cr, and NAA for 4 different semiLASER VOIs sizes/orientations.
The effect of CSDE can clearly be seen when comparing NAA- and Cho-maps. In
addition, all metabolite maps for larger VOI do not show fully refocussed
signal amplitudes in the top right volume, indicated by arrows, which can be
attributed to the lack of adiabatic condition, as rotation by 45-degrees leads
to the same pattern. DISCUSSION
Of
the 9 of 10 prospectively examined patients we currently have histopathology findings.
SLOW-editing was able to identify the correct IDH-status in 8 out of 9 patients
correctly; in the failing case, the tumour was located right above the nasal-cavity,
was poorly B0-shimmed. MEGA-semiLASER-CSI/SVS based 2HG-editing
showed in nine cases often large ghosting artifacts in the 3.8–5.3 ppm range,
often preventing unambiguous 2HG-detection (see three typical cases in Figure 3).
The semiLASER 2π-SSAP
pulse-times were chosen to be 8040 µs, therefore these pulses could reach their correct
nominal 180-degree flip-angles. However, a sufficiently larger B1+-amplitude
(>180 degrees) is required the guarantee robust adiabatic behaviour under
inhomogeneous B1+-conditions. For the narrowband 2π-CSAP of
SLOW, this condition can be easily met. Apparently, the broadband 2π-SSAP of
semiLASER does not meet this criterium in our Nova 1Tx 32Rx head coil. Finally,
the degradation of spectral quality observed in MEGA-semiLASER editing is not
only due to the B1+-inhomogeneities, but also B0-
inhomogeneities and insufficient water suppression, which can be overcome using
SLOW-editing4,9. CONCLUSION
Due
to its low B1+-peak power and SAR-requirements at
UHF, SLOW-editing based MRSI can satisfy the adiabatic condition for 2HG
editing in vivo, while this condition is poorly satisfied for the
MEGA-semiLASER, resulting in large CSDE and strong ghosting-artifacts, which
are also due to B0-inhomoginieties and poor water-suppression. Acknowledgements
The research is supported by the Swiss National Science Foundation (SNSF-182569).References
- Choi C, Ganji SK, DeBerardinis RJ, et al.
2-hydroxyglutarate detection by magnetic resonance spectroscopy in IDH-mutated
patients with gliomas. Nature Medicine 2009 18:4. 2012;18(4):624-629.
- Andronesi OC, Kim GS,
Gerstner E, et al. Detection of 2-hydroxyglutarate in IDH-mutated glioma
patients by in vivo spectral-editing and 2D correlation magnetic resonance
spectroscopy. Science translational medicine. 2012;4(116):116ra4.
-
Moser P, Hingerl L, Strasser B, et al. Whole-slice mapping
of GABA and GABA + at 7T via adiabatic MEGA-editing, real-time instability
correction, and concentric circle readout. NeuroImage. 2019;184:475-489.
- Weng G, Kiefer C, Zubak I, Slotboom J. SLOW: Whole
Brain Spectral Editing EPSI Based Technique using Chemical Selective Adiabatic
2π-Refocusing Pulses applied to 2HG and GABA Editing. In: ISMRM & SMRT
Annual Meeting & Exhibition. ; 2021:1990-1990.
- Mescher M, Merkle H, Kirsch J, Garwood M, Gruetter R.
Simultaneous in vivo spectral editing and water suppression. NMR in Biomedicine.
1998;11(6):266-272.
- Garwood M, Delabarre L. The Return of the Frequency Sweep:
Designing Adiabatic Pulses for Contemporary NMR. Journal of Magnetic
Resonance. 2001;153:155-177.
- Slotboom J, Mehlkopf AF, Bovee WMMJ. The Effects of
Frequency-Selective RF Pulses on J-Coupled Spin-1/2 Systems. Journal of
Magnetic Resonance, Series A. 1994;108(1):38-50.
- Ebel A, Maudsley AA. Improved spectral quality for 3D MR
spectroscopic imaging using a high spatial resolution acquisition strategy. Magnetic
Resonance Imaging. 2003;21(2):113-120.
- Weng G, Sheriff S, Kiefer C, Zubak I, Maudsley A, Slotboom
J. Simultaneous Water and Lipid Suppression Using Chemical Selective Adiabatic
Refocusing Pulses Echo Planar Spectroscopic Imaging (EPSI) at 7T. In: ISMRM
& SMRT Annual Meeting & Exhibition. ; 2021:1985-1985.
- Marjańska M, Auerbach EJ, Valabrègue R, van de Moortele PF,
Adriany G, Garwood M. Localized 1H NMR spectroscopy in different regions of
human brain in vivo at 7T: T 2 relaxation times and concentrations of cerebral
metabolites. NMR in Biomedicine. 2012;25(2):332-339.
- Slotboom J, Vogels BAPM, Dehaan JG, et al. Proton Resonance
Spectroscopy Study of the Effects of L-Ornithine-L-Aspartate on the Development
of Encephalopathy, Using Localization Pulses with Reduced Specific Absorption
Rate. Journal of Magnetic Resonance, Series B. 1994;105(2).
- Slotboom J, Mehlkopf AF, Bovee WMMJ. The Bloch Equations for
an AB System and the Design of Spin-State-Selective RF Pulses for Coupled Spin
Systems. Journal of Magnetic Resonance, Series A. 1994;106(2):159-167.