Can Wu1, Neelam Tyagi1, Marsha Reyngold2, Christopher Crane2, and Ricardo Otazo1,3
1Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 2Department of Radiation Oncology, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 3Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, United States
Synopsis
Real-time 3D MRI for low-latency volumetric motion
tracking was successfully implemented on a 1.5T MR-Linac system using the MRSIGMA
framework. A first scan was performed for offline learning to obtain a training
dictionary of ten 3D motion states. A second scan with the same sequence parameters
was performed to generate the motion signature in real-time for online matching
and was also used as a reference for retrospective self-validation. The
feasibility of the technique was demonstrated on a healthy volunteer and a
patient with pancreatic cancer which presented high quantitative concordance between
contours of real-time MRSIGMA matching and the reference.
INTRODUCTION
The MR-Linac system
offers opportunities for real-time adaptive treatment for tumors affected by
continuous motion, such as those located in the upper abdomen which are
affected by respiratory and digestive motion.1,2 Real-time adaptive
treatment holds the promise to focus the radiation beam to the tumor and spare
nearby organs at risk which can enable definitive doses to be given. To achieve
this goal, real-time 3D MRI with sufficient volumetric coverage to include the
tumor and organs at risk is required to track motion and shape the radiation
beam accordingly. However, current real-time MRI technology on the MR-Linac is
limited to 2D imaging, which limits the performance of motion tracking. This
work presents the implementation of the MR signature matching (MRSIGMA)3,4
technique on a 1.5T MR-Linac system to enable low-latency 3D MRI to track
motion of pancreatic tumors and organs at risk.METHODS
Concept of MRSIGMA: MRSIGMA
shifts all the acquisition and reconstruction burden to an offline learning
step (Figure 1), where continuously acquired stack-of-stars data (~5min) are
reconstructed into 3D motion states using XD-GRASP.5 The motion
range along the z dimension in each motion state is the motion signature. After
offline learning, MRSIGMA moves to a real-time online signature matching step,
where the same acquisition is employed, but for each angle (~250ms), a motion
signature is computed and matched to one of the pre-computed 3D motion states (~50ms),
enabling low-latency 3D MRI.
MRSIGMA Data Acquisition: MRSIGMA data was acquired from a healthy volunteer (female,
31 years old) and a patient with pancreatic cancer (male, 59 years old) on a 1.5T
Unity MR-Linac System (Elekta AB, Stockholm, Sweden). A free-breathing
T1-weighted 3DVANE sequence was modified for 3D radial golden-angle stack-of-stars
acquisitions. A training dataset was first acquired for offline learning and a
second scan with the same sequence parameters was performed to generate motion
signature for online matching and was also used as a reference for
retrospective self-validation. Data acquisition parameters are as follows: TR/TE = 5.0/2.1 ms, flip angle = 12°, FOV = 370×370×208
mm3, voxel size = 1.5×1.5×4.0 mm3, number of radial
spokes = 905, bandwidth = 720 Hz/pixel, total scan time = 5:50 min. The latency
of acquiring a motion signature (one angle) is about 275 ms.
MRSIGMA Image Reconstruction: K-space data in raw data format
(.data/.list files) was transferred to a high-performance computer and MRSIGMA
reconstruction of 4D images (10 motion states of 3D volume) was performed in
Matlab (MathWorks, Natick, MA) using in-house algorithms, as described
previously.3,4
Image Contouring: Clinical contours of the patient were transferred to the first
phase (expiration) of the two 4D image sets after a fusion of a clinical
T1-weighted scan and the two MRSIGMA scans. The contours were then
automatically propagated from the first phase to all remaining phases using a
deformable propagation algorithm in the MIM software (MIM Software Inc,
Cleveland, OH) and were visually inspected thereafter for minor manual
adjustment.
Motion Tracking Performance
Evaluation: Gross
tumor volume (GTV) of the pancreas, the duodenum and stomach (Duo-stomach) as
the high-risk organs, and three other nearby organs at risk (small and large
bowel, and liver) were selected for evaluation. Dice coefficient was calculated
for all contours of the same organ between the online matching and the reference.RESULTS
Selected motion states
(expiration, middle and expiration phases) of the offline training dictionary
and non-real-time reference are shown in Figure 2 and Figure 4 for the healthy
volunteer and patient, where the pancreas, duodenum, and stomach are indicated
by arrows in green, yellow, and white, respectively. Motion of these organs
from expiration to inspiration can be readily seen for both scans. Figure 3
illustrates a good agreement of organ motion between online matching and the reference
as synchronized by the motion signature. Figure 5 shows the change of the
contours (only GTV and Duo-stomach are shown) for online signature matching in red,
which is in good agreement with the contours of the reference in green. The Dice
coefficients of the contours are: 0.912±0.016 (GTV), 0.863±0.032 (Duo-stomach),
0.876±0.061 (small bowel), 0.777±0.036 (large bowel), and 0.963±0.007 (liver),
further confirming that online signature matching agrees well with the
reference.DISCUSSION
The preliminary work demonstrated
the feasibility of real-time 3D MRI for low-latency volumetric motion tracking on
a 1.5T MR-Linac system using the MRSIGMA framework. This opens the door for
adaptive planning in radiation therapy with smaller treatment margins and more focused
dose delivery. There are a few ways to further reduce the latency of acquiring
a motion signature, for example, using a lower acquisition resolution, reducing
the number of slices to only cover the lung-liver interface, or using a pencil
beam liver-lung-navigator. Further work is needed to implement a streamlined workflow
for fast raw-data transfer and deep-learning-based image reconstruction to
achieve total latency below 300 ms for real-time motion tracking in radiation therapy.CONCLUSION
Real-time 3D MRI for low-latency volumetric motion
tracking was successfully implemented on a 1.5T MR-Linac system using an online
matching of the motion signature to the offline training dictionary. The
feasibility was demonstrated on a healthy volunteer and a patient with
pancreatic cancer with good motion tracking performance.Acknowledgements
The work was supported by NIH Grant R01CA255661.References
1. Mutic
S, Dempsey JF. The ViewRay system: Magnetic Resonanceāguided and controlled
radiotherapy. Semin Radiat Oncol. 2014; 24(3):196-199.
2. Raaymakers
BW, Lagendijk JJW, Overweg J, et al. Integrating a 1.5 T MRI scanner with a 6
MV accelerator: proof of concept. Phys Med Biol. 2009. 54(12):N229-N237.
3. Feng
L, Tyagi N, Otazo R. MRSIGMA: Magnetic Resonance SIGnature MAtching for
real-time volumetric imaging. Magn Reson Med. 2020; 84(3):1280-1292.
4. Kim
N, Tringale KR, Crane C, et al. MR SIGnature MAtching (MRSIGMA) with
retrospective self-evaluation for real-time volumetric motion imaging. Phys Med
Biol. 2021 Oct 26;66(21).
5. Feng
L, Axel L, Chandarana H, et al. XD-GRASP: Golden-angle radial MRI with
reconstruction of extra motion-state dimensions using compressed sensing. Magn
Reson Med 2016;75(2):775-788.