Ferdinand Seith1, Cecilia Zhang1, Manuel Kolb1, Brigitte Gückel1, Rolf Pohmann2, Petros Martirosian3, and Ahmed E. Othman1,4
1Diagnostic and Interventional Radiology, University Hospital of Tübingen, Tübingen, Germany, 2High-Field MR Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany, 3Section on Experimental Radiology, University Hospital of Tübingen, Tübingen, Germany, 4Neuroradiology, University Hospital of Mainz, Mainz, Germany
Synopsis
We present preliminary results of a prospective study of 96 patients examined with pCASL-MRI of the lung with suspected pulmonary embolism. pCASL-MRI of the lung was acquired in a 1.5T scanner by labeling the pulmonary trunk during systole. Coronal, sagittal and axial images were acquired using a fast bSSFP sequence. Additionally, a mulit-slice coronal bSSFP imaging the whole lung was performed. Two blinded readers evaluated MR-images to detect pulmonary embolism and noted the diagnostic confidence (5=very good). Compared to CT, sensitivity and specificity were 0.971 and 0.951. The median diagnostic confidence was 5 in both readers (kappa 0.739).
Introduction
Acute pulmonary
embolism (PE) requires immediate and precise diagnostic imaging to confirm the
diagnosis and assess the severity of pulmonary perfusion defects.
Contrast-enhanced computed tomography (ceCT) is the reference standard and
detects PE as filling defects in pulmonary arteries. Recently, we presented pseudocontinuous
Arterial Spin Labeling (PCASL) with balanced steady-State Free-Precession
(bSSFP) readout at 1.5T as a non-invasive technique to image pulmonary
perfusion even under free breathing conditions1. However, the
performance of ASL imaging of the lung in clinical routine remains unclear. We
present the first preliminary results of a prospective study aimed at
determining the diagnostic performance of PCASL imaging in direct comparison to
ceCT in 96 patients with suspected PE.Material and Methods
Examination: This study was approved by the local ethics committee
and is registered at the German Clinical Trials Register (DRKS00023599).
Inclusion criteria: adult patients with clinically suspected PE examined in
ceCT within the last 72h, haemodynamically stable, written informed consent.
Exclusion criteria: >110 bpm, mmHg syst. <100, typical contraindications
for MRI. Assuming a prevalence of PE of 0.32, the required sample
size is 97 participants for a two-sided 95% sensitivity confidence interval
with a width of 0.2. PCASL MRI was performed with a 1.5T MR scanner (MAGNETOM
Avantofit, Siemens Healthcare, Erlangen, Germany) under free
breathing conditions. The labeling plane was placed perpendicular to the
pulmonary trunk and the pulse train was triggered by the ECG signal during the
systolic period (labeling duration 300 ms). Following a post-labeling delay of
1000 ms, coronal, sagittal and axial images were acquired using a fast bSSFP
sequence (Fig. 1) with following parameters: TR 2.1 ms, TE 0.9 ms, flip angle
70°, FoV 480×480 mm2, ST 20 mm, matrix 144×192, BW 1240 Hz/Pixel. Additionally,
a multi-slice coronal bSSFP imaging of the whole lungs was performed as follows:
TR 390.6 ms, TE 1.35 ms, flip angle 80°, FoV 480×480 mm2, ST 4 mm,
matrix 210×320, BW 975 Hz/Pixel, iPAT 2. Total examination time ranged from 15-20
min.
Analysis: Registration of
perfusion data was conducted by the open-source toolbox elastix3,4
using an in-house developed MATLAB® (The Mathworks, Natick, MA) script as
decribed in the previous work1. A blinded reading of PCASL and multi-slice
bSSFP images was performed by two senior radiologists with several years of
experience in MRI. Perfusion defects suspected for PE were noted and the
diagnostic confidence was rated using a likert scale (5 is very high).Results
From 11/2020 till
11/2021, 96 of 97 patients were included to the study. In 31 patients, PE was
diagnosed in ceCT, an example is given in Fig. 2. PCASL MRI could be performed
successfully in all patients. Both readers rated 1 patient false negatively and
3 patients false positively in PCASL MRI. Median diagnostic confidence was
rated as 5 in both readers (inter-rater reliability kappa was 0.739).
Sensitivity and specificity were 0.971
and 0.951.Discussion
ceCT is the
standard of care in clinical routine for the detection of PE, but requires
intravenous contrast agent to detect blood clots in the pulmonary arteries
which can cause life-threatening allergic reactions or nephropathy. Radiation
exposure should be avoided in young or pregnant patients which can lead to a
diagnostic dilemma in clinically unclear cases. Moreover, ceCT can not reveal
the resulting functional information, i.e. the perfusion defects in lung
parenchyma. In the present study, we could demonstrate that PCASL-MRI of the
lung can be applied in a clinical setting even under free breathing conditions
in a large patient cohort. Together with multi-slice bSSFPimaging of the whole
lung, PE could reliably be detected with a low rate of false
positives/negatives. These cases need to be evaluated more in detail in the
following data evaluation. The diagnostic confidence was very high in both
readers which a good inter-rater reliability. This underlines the potential of PCASL
MRI of the lung for clinical questions in the future.Acknowledgements
No acknowledgement found.References
1. Seith F, Pohmann R, Schwartz M, et al. Imaging Pulmonary Blood Flow Using Pseudocontinuous
Arterial Spin Labeling (PCASL) With Balanced Steady-State Free-Precession
(bSSFP) Readout at 1.5T. J Magn Reson Imaging. 2020;52(6):1767-1782.
2. Penaloza A, Kline J, .Verschuren F, et al. European and American suspected and
confirmed pulmonary embolism populations: comparison and analysis. JThromb and
Haemost 2012;10:375-381.
3. Klein
S, Staring M, Murphy K, et al. Elastix: a toolbox for intensity based medical
image registration. IEEE Trans Med Imaging. 2010;29(1):196-205.
4. Shamonin DP, Bron EE, Lelieveldt BPF, et al. Fast parallel image
registration on CPU and GPU for diagnostic classification of Alzheimer’s
disease. Front Neuroinform. 2014;7:50.