Simona Baroni1, Valeria Bitonto1, Maria Rosaria Ruggiero1, Alessandra Pittaro2, isabella Castellano2, Riccardo Bussone3, Lionel M. Broche4, David J. Lurie4, Silvio Aime1,5, and Simonetta Geninatti Crich1
1Department of Molecular Biotechnology and Health Sciences - Molecular Imaging Center, University of Turin, Torino, Italy, 2Department of Medical Sciences, University of Turin, Torino, Italy, 3Breast Unit, Ospedale Cottolengo, Torino, Italy, 4University of Aberdeen, Aberdeen, United Kingdom, 5IRCCS SDN, Napoli, Italy
Synopsis
A new methodology, based on low field NMR relaxometry,
has been employed to characterize breast tumour tissue and to tackle the
important clinical need for the assessment of tumour margins during
breast-conserving surgery.
The method relies on the acquisition of proton
longitudinal relaxation rate at low magnetic field strengths which are affected
by the changes in the composition of the mammary gland tissue occurring during
the development of neoplasia. A sensitivity, specificity, and accuracy of 92%,
85%, and 89%, respectively, were achieved. The information obtained has the
potential to support the surgeon in real-time margin assessment.
Introduction
As Breast-Conserving
Surgery is routinely applied for treatment of breast cancer (the most common form
of cancer for women), the need for new technology to improve intraoperative
margin assessment has become increasingly important1. The current
gold standard for margin classification is provided by the histopathological
evaluation (H&E staining). The method is robust and accurate, but it is
time consuming2. About 20–40% of BCS procedures result in margins
which are either positive or suspected of having malignant cells at the margins
of the resection region and require reoperation3. Several
intraoperative margin assessment strategies have been proposed to reduce the
need for re-excision, but they all have significant clinical and technical
limitations that have hampered widespread adoption4,5.
In this
study, the potential of fast field-cycling 1H-NMR relaxometry as a
new diagnostic tool was evaluated. It is well known that proton longitudinal
relaxation rate (R1) values of a given tissue decrease as a function
of the applied magnetic field strength, thus providing larger differences among
the tissues the lower the applied magnetic field strength is6. It
has been shown by our group that the Magnetic Nuclear Relaxation Dispersion
(NMRD) profiles can act as a high-sensitivity tool for cancer detection and
staging in ex vivo murine breast tissues7 or in breast cancer
cell lines8 and in vivo with mice transplanted with murine
mammary cancer cells9.Methods
Small freshly excised
tissue samples (n=104, with a range of 2.5–8.9 mm Ø) from 41 patients
undergoing breast cancer surgery were collected and subjected to R1 measurements at very low magnetic field
strengths, then sent to gold standard histopathological evaluation (H&E
staining). The acquisition of R1 was performed on a SpinMaster
FFC-NMR relaxometer (Stelar S.n.c., Mede, Italy), equipped with a
signal-detection microcoil of 10 mm diameter, at 10 °C in the 0.02–1 MHz proton
Larmor frequency range (B0 = 0.48 mT–24 mT) pre-polarising the sample
at 25 MHz10. The water exchange rate constants across the cell
membrane and the extracellular volume fraction were determined by analysing the
data of tumour samples using the 2SX model11. Results
The
histological analysis performed by E&H staining allowed the classification
of the 104 breast tissue samples investigated as follows: 40 healthy (H), 21
tumours (T), and 43 containing a mixture of both (M). Both H and T tissue
relaxation rates increase when the magnetic field strength decreases, but the
relative values and slopes of the two curves are significantly different (Fig.
1). Accordingly, we defined two relaxometric quantifiers that captured this
behaviour and allowed the assessment of the presence of tumour cells in a
breast tissue specimen. The first one was the ratio between the R1
value measured at the lowest (0.02 MHz) and highest (1 MHz) magnetic fields (later
referred to as the Ratio). The second one was the sum of the R1
values measured at 0.39 MHz and 1 MHz (later referred to as the 2R1
value). The Receiver Operating Characteristic curve analysis was used to assess
the performance of the two quantifiers in the discrimination among negative (H)
and positive (M+T) specimens (Fig. 2). The best cutoff values were found to be
2.19 and 24.0 s-1, respectively. The T samples classified as Grade 3
were characterized by high transcytolemmal efflux rate constants.Discussion
The
difference between healthy and malignant tissue appears to be associated with
the different water content and water mobility characteristics of the tissues:
tumour tissue has a protein/fat/water content that is highly altered with
respect to healthy breast tissue, in which adipocytes are dominant and lipids
account for up to 70–80% of tissue content. As previously reported7,
lipid proton relaxation rates show significatively less dispersion with the
magnetic field strength in the range 0.02–10 MHz than water/protein protons. Therefore,
relaxation rates of H tissues show higher values and a less pronounced
dispersion with the magnetic field with respect to T tissues. Preliminary data
indicate that the Ratio and the transcytolemmal efflux rate constants can
report on tumour grading.
Considering
the discriminating ability of the two defined quantifiers, we developed a protocol
(Fig. 3) that privileged higher sensitivity over specificity by applying two
criteria in a specific order (Ratio first, 2R1 second). The
sensitivity (92%) and specificity (85%) attained appear competitive with other
methods and well comparable with the best procedures currently proposed for
intraoperative tumour margin assessment4,5.
The
relaxometric method is carried out without requiring any additional treatment
of the excised specimen and only needs a reasonable time to complete (currently
6–9 min), which is well compatible within the timeframe of BCS. Conclusion
The proposed
relaxometric method has good sensitivity and specificity, is low cost, fast and
intraoperatively applicable with minor modifications.
We believe that the results reported here
clearly indicate that tissue relaxometric data obtained at different magnetic
field strengths may provide a useful tool to assess the presence of tumour
cells in an excised tissue specimen by a simple quantitative analysis that has
the potential to support the surgeon in real-time margin assessment during Breast-Conserving Surgery. Moreover, the water efflux rate constant may be
exploited as a specific biomarker giving a fast and specific determination of
tumour aggressiveness.Acknowledgements
This
work was performed in the frame of the COST Action AC15209 (EURELAX).
This
project has received funding from the European Union Horizon 2020 research and
innovation program under grant agreement No 668119 (project “IDentIFY”) and
from AIRC under IG 2019, ID 23267 project (PI Geninatti Crich Simonetta). Maria
Rosaria Ruggiero was supported by a “FIRC-AIRC fellowship for Italy”. The
authors acknowledge the Italian Ministry of Research for FOE contribution to
the Euro-BioImaging MultiModal Molecular Imaging Italian Node (www.mmmi.unito.it).References
1. Schwarz J and
Schmidt H. Technology for intraoperative margin assessment in breast Cancer. Ann.
Surg. Oncol. 2020;27:2278–2287.
2.
Moran MS, Schnitt SJ, Giuliano AE et al. Society
of surgical oncology-American society for radiation oncology consensus
guideline on margins for breast-conserving surgery with whole-breast
irradiation in stages I and II invasive breast cancer. Int. J. Radiat. Oncol.
Biol. Phys. 2014;88:553–564.
3.
Elmore LC and Margenthaler JA. A tale of
two operations: Re-excision as a quality measure. Gland Surg. 2019; 8:593–595.
4.
Maloney BW, McClatchy DM, Pogue, BW et al.
Review of methods for intraoperative margin detection for breast conserving
surgery. J. Biomed. Opt. 2018;23:1–19.
5.
Pradipta AR, Tanei T, Morimoto K et al. Emerging
Technologies for Real-Time Intraoperative Margin Assessment in Future
Breast-Conserving Surgery. Adv. Sci. 2020;7:1901519.
6.
Korb JP, Bryant RG. Magnetic field
dependence of proton spin-lattice relaxation times. Magn. Reson. Med. 2002;48:21–26.
7.
Di Gregorio E, Ferrauto G, Lanzardo S et al. Use of
FCC-NMRD relaxometry for early detection and characterization of ex-vivo murine
breast cancer. Sci. Rep. 2019;9:1–8.
8.
Ruggiero MR, Baroni S, Aime S, et al.
Relaxometric investigations addressing the determination of intracellular water
lifetime: A novel tumour biomarker of general applicability. Mol. Phys. 2018;117:
968–974.
9.
Ruggiero MR, Baroni S, Pezzana S et al. Evidence
for the Role of Intracellular Water Lifetime as a Tumour Biomarker Obtained by
In Vivo Field-Cycling Relaxometry. Angew. Chemie Int. Ed. 2018;57:7468–7472.
10.
Ferrante G and Sykora S. Technical aspects
of Fast Field Cycling. Adv. Inorg. Chem. 2005;57:405–470.
11.
Springer CS, Li X, Tudorica LA et al. Intratumor
mapping of intracellular water lifetime: metabolic images of breast cancer? NMR
Biomed. 2014;27:760–773.