Anne-Sophie van Schelt1, Nienke P.M. Wassenaar1, Jurgen H. Runge1, Jules L. Nelissen1, Marian Troelstra1,2, Christian Guenthner3, Ralph Sinkus2,4, Jaap Stoker1, Aart J. Nederveen1, and Eric M. Schrauben1
1Radiology and Nuclear Medicine, Amsterdam UMC location AMC, Amsterdam, Netherlands, 2Department of Imaging Sciences and Biomedical Engineering, King’s College London, London, United Kingdom, 3Department of Biomed, ETH Zurich, Zurich, Switzerland, 4Inserm U1148, LVTS, Univeristy Paris Diderot, Paris, France
Synopsis
Magnetic resonance
elastography (MRE) allows for non-invasive determination of pancreatic
visco-elastic properties in four consecutive breath-holds. The aim of this
study was to develop and test a single breath-hold MRE acquisition accelerated using
prospective undersampling and compressed sensing (CS) reconstruction. Testing
was done on a retrospective undersampled phantom dataset and in-vivo. These showed that CS
accelerated (R< 8.7), single breath-hold MRE is feasible without hampering
visco-elastic reconstruction in tissues with low shear stiffness
(|G*|<1.6kPa). Further research is necessary to guarantee accuracy of the
measured shear stiffness, notably for high stiffness tissues such as tumors,
and at higher acceleration factors.
Introduction
Magnetic resonance
elastography (MRE) allows for non-invasive determination of tissue
visco-elastic properties, which differ for healthy and diseased tissues. Longitudinal
changes in tissue stiffness in the course of pancreatitis, endocrine dysfunction or pancreatic cancer could
reveal new insights in the disease pathophysiology[1]. However, assessing the
pancreas with MRE remains challenging due to its small size, elongated shape,
and central abdominal location. The quantitative accuracy of MRE in small
structures depends on the local stability of the tissue being imaged over all wave-offsets
and motion-encoding directions. However, current pancreatic MRE requires four
consecutive (16s) breath-holds[2]. These may introduce errors in pancreas
location, limit spatial and temporal resolution, and can be particularly
uncomfortable for patients.
The aim of
this study was to develop and test a single breath-hold MRE acquisition accelerated
using prospective undersampling and compressed sensing (CS) reconstruction in
the pancreas.Methods
We modified
the 3D Ristretto GRE-MRE acquisition to accommodate for compressed sensing
acceleration[3,4]. Feasibility of CS-accelerated MRE was first established by
retrospective undersampling (factors:R=0-15) and reconstruction of a 3D MRE
dataset of a CIRS phantom with inclusions (CIRS,Inc.,Norfolk,USA). Data was
acquired by Guenthner et al. using 36Hz vibration frequency and eight wave-phase
offsets in 14 slices (FOV:192x156x42mm)[3]. Scan parameters are reported in Figure
1a.
In-vivo testing was performed on six healthy
volunteers (female=6, mean age=27±3 years). A gravitational transducer [5] with
a 3D-printed curved polyactic-plate was strapped to the right flank at the
level of the pancreas head (Figure.1c). High-resolution T2-weighted images were
made for anatomical reference. MRE acquisitions were performed using 40Hz vibration
frequency, nine slices (FOV:336x192x27mm) and five wave-phase offsets. Six
consecutive pancreatic MRE scans were performed: (I)standard Multi-Slice (MS), (II)3D
SENSE-accelerated four breath-hold acquisition, (III)3D CS-accelerated four
breath-hold acquisition and (IV-VI)three 3D CS-accelerated single breath-hold
acquisitions. Scan parameters are reported in Figure 1a.
Incoherent
undersampling was achieved using variable-density Poisson disks with half-scan
in both ky (67%) and kz (89%) directions(Figure.1b). Five
different undersampling patterns with different undersampling factors (R=3.6,8.7,9.8,13.4)
were used for each wave-phase offset and were repeated for each encoding
direction. For CS acquisitions, reconstruction was performed offline using
ReconFrame (Gyrotools,Zurich,Switzerland) in MATLAB (MathWorks,Natick,MA,USA)
together with the Berkeley Acquisition Advanced Reconstruction Toolbox (BART) using
total variation sparsifying transform in the wave-offset dimension (l=0.001,
20 iterations)[6,7,8].
Phantom
post-processing was performed using MATLAB. The shear stiffness of each
inclusion and background was calculated and compared to the non-accelerated stiffness.
In-vivo MRE post-processing was
performed using dedicated software(ROOT,KIR). Phase data are unwrapped and
smoothed before 3D transformation using a 3D Blackman-Harris window. The shear
wave displacement was calculated per voxel and visco-elastic parameters were
extracted with a finite element method. Mean shear wave speed (SWS) and shear
stiffness (|G*|) were calculated by manually drawn region-of-interest over the
pancreatic head and liver on MRE magnitude images guided by anatomical
high-resolution T2-weighted images.
Mean SWS, |G*|
and the nonlinearity (ratio of the second harmonic over the first harmonic) in
both the pancreas head and liver were analyzed, with 3D-SENSE-MRE allocated as
index test for CS. Statistical analysis was done using repeated measures ANOVA
and pairwise comparison with a Bonferroni correction(p<.05) for SWS and
nonlinearity parameters of the pancreas and liver. Results
Mean |G*|and
example elastograms of the phantom data are shown in Figure 2. In-vivo examples of SWS-maps in the
pancreas are shown in Figure 3. The mean SWS, |G*| and nonlinearity for the
pancreas and liver are given in Figure 4. Mean SWS and nonlinearity of each
method for the pancreas and liver are shown in Figure 5. Repeated measures
ANOVA showed no significant differences in SWS overall(f=5/1,p=0.07). Exploratory
pairwise comparison showed significant difference between mean SWS of MS and
CS13.4(p=.03) in the pancreas and nonlinearity in the liver between CS3.6 and
CS8.7(p=.02).Discussion
Phantom retrospective
undersampling showed accurate results for low stiffness inclusions and
background (|G*|<1.6kPa) up to an acceleration factor of 15 (Figure.2).
However, for the two stiff inclusion the apparent stiffness deviated from non-accelerated
R0 values, with underestimation increasing for higher acceleration factors. The
relative low mechanical frequency (and hence long wavelength) could explain
this disparity[9].
CS-accelerated
MRE gave comparable stiffness values in-vivo up to an acceleration factor of 8.7.
The nonlinearity showed no significant difference between all scans in the
pancreas, indicating that higher acceleration does not significantly hamper visco-elastic
reconstruction. However, accuracy of MRE inversion is only guaranteed for a
nonlinearity below 50%, therefore CS acceleration of 13.4 (nonlinearity=50.5%)
could be considered borderline acceptable.
Liver nonlinearity
shows a significant difference between CS3.6 and CS8.7, however, this is not
seen for higher acceleration. This could be caused by the large spread observed
at CS13.4. A large spread in quality parameters, such as nonlinearity, could
indicate loss in accuracy and thereby reproducibility. Additionally, while acceleration
of CS=8.7 displayed no difference in SWS and nonlinearity compared to MS, the
breath-hold duration (20s) may be too long for some patients. Conclusion
Phantom and
in-vivo testing shows that compressed
sensing accelerated, single breath-hold MRE is feasible without hampering
visco-elastic reconstruction in tissues with low shear stiffness
(|G*|<1.6kPa). Further research is necessary to guarantee accuracy of the
measured shear stiffness, notably for high stiffness tissues such as tumors,
and at higher acceleration factors. Acknowledgements
No acknowledgement found.References
[1] MacCurtain BM, Quirke NP, Thorpe SD, Gallagher TK.
Pancreatic Ductal Adenocarcinoma: Relating Biomechanics and Prognosis. Journal of Clinical Medicine. 2021; 10(12):2711. https://doi.org/10.3390/jcm10122711
[2] van Schelt A, Wassenaar N, Schrauben
E, Troelstra M, Nelissen J, Sinkus R, Stoker J, Nederveen A, Runge J. (2021) Optimization of GRE-MRE
in the pancreas. ISMRM Abstract number 1021
[3] Guenthner
C, Sethi S, Troelstra M, Dokumaci AS, Sinkus R, Kozerke S. Ristretto MRE: A
generalized multi-shot GRE-MRE sequence. NMR Biomed. 2019;32(5).
doi:10.1002/nbm.4049
[4] Christian
Guenthner, Sweta Sethi, Ayse Sila Dokumaci, Ralph Sinkus, and Sebastian Kozerke
(2018) ISMRM Abstract number 5584
[5] Runge,
J. H., Hoelzl, S. H., Sudakova, J., Dokumaci, A. S., Nelissen, J. L.,
Guenthner, C., ... Sinkus, R. (2019). A novel magnetic resonance elastography
transducer concept based on a rotational eccentric mass: Preliminary
experiences with the gravitational transducer. Physics in Medicine and Biology,
64(4):045007. doi:
10.1088/1361-6560/aaf9f8.
[6] Mazzoli, V, Gottwald, LM, Peper, ES, et al. Accelerated 4D phase
contrast MRI in skeletal muscle contraction. Magn Reson Med. 2018; 80: 1799–
1811.
[7] Uecker M, Ong
F, Tamir JI, et al. Berkeley Advanced Reconstruction Toolbox. Proc Intl Soc Mag Reson Med.
2015;23(1):2486.
[8] Beck, Amir, and Marc Teboulle. A fast iterative
shrinkage-thresholding algorithm for linear inverse problems. SIAM journal on imaging sciences 2.1
(2009): 183-202.
[9] Guenthner, C, Sethi, S, Troelstra, M,
et al. Unipolar
MR elastography: Theory, numerical analysis and implementation. NMR in
Biomedicine. 2020; 33:e4138. https://doi.org/10.1002/nbm.4138