Seraina Anne Dual1,2, Mattia Arduini1, Brigitte Schmittlein 3, Judith Zimmermann1,4, Ellen Roche5, and Daniel Ennis2,6,7
1Department of Radiology, Stanford University, Palo Alto, CA, United States, 2Cardiovascular Institute, Stanford University, Palo Alto, CA, United States, 3Department of Mechanical Engineering, Stanford University, Palo Alto, CA, United States, 4Department of Computer Science, Technical University of Munich, Garching, Germany, 5Institute for Medical Engineering & Science, Massachusetts Institute of Technology, Cambridge, MA, United States, 6Department of Radiology, Stanfor University, Palo Alto, CA, United States, 7Division of Radiology, Veterans Affairs Health Care System, Palo Alto, CA, United States
Synopsis
MRI can be used to evaluate the interaction of soft robotic devices
with biological tissues. The aim of this study was to use MRI to optimize the actuation
pressure of an extra-aortic soft robotic cardiac assist device. We evaluated wall
deformations, displaced fluid volumes and qualitative flows induced by the
actuator on a fluid filled elastic vessel. Our results show that the displaced
volume increases with the actuator pressure and decreases with increasing the
vessel pressure. The maximum actuator pressures do not lead to buckling of the
vessel wall given the current three-bladder design.
Background
Soft robots interact
favorably with humans because the material properties can match those of soft biological
tissue.1 Soft robots have been shown to mechanically assist the failing
heart. While epicardial attachment avoids blood contact, securing the robot to
the heart remains challenging.2 In contrast, soft robots are more easily wrapped around large blood
vessels.3 We built an extra-aortic soft robot for hemodynamic support. Still,
modeling and measuring the interaction of soft actuators with biological
tissues remains a challenge, given their viscoelastic properties, curvilinear
surfaces, and large deformations.
We propose to
use MR as an imaging modality to measure deformation, displaced volume and
flows in mock circulatory flow loops to enable design optimization of a
programmable soft robot. One actuator that suites for extravascular actuation
is the McKibben pneumatic artificial muscle.4 Upon actuation the soft robot contracts, displaces volume, and in
combination with the aortic valve creates forward flow. The McKibben actuator
achieves a high contraction ratio when pressurized and can operate at a similar
frequency to the heart.5
The aim of this
work was to use MRI for assessing the pressure and volume displacement of a
soft robotic McKibben actuator interacting with a compliant synthetic vessel.Methods
We built an MRI
compatible electropneumatic control system and flow loop for testing a soft
robot inside an MRI scanner. The control system2 pressurizes the actuator for 300 ms, then depressurizes for 700 ms
cyclically (60 bpm) at 6, 8, 10, and 12psi and retrospectively gates image
acquisition (Figure 1A). To address both MRI safety and potential
radiofrequency (RF) artefacts, the design of the control system was
Faraday-shielded (Cable Management, Electriduct, USA).
One McKibben
actuator was manufactured using a pre-fabricated mesh sleeve (Flexo PET, Techflex,
USA) and a thermoformed inner bladder (Stretchlon 800, Fibreglast, USA). The three-bubble
shape may minimize vessel wall deformation (Figure 1B). The actuator was
wrapped around compliant tubing (high-temperature silicone rubber, durometer
35A) filled with a blood mimicking fluid (40% glycerol/60% water), surrounded
by static gel, and was placed shortly downstream of a backflow valve (Figure
1C). The intra-vessel pressure was tuned to 50, 75, 100, and 150 mmHg.
Pressure induced by the actuator was measured on the benchtop prior to MRI
acquisition using a pressure transducer (SPR350-S, Millar) sampling at 1 kHz.
We performed MR
imaging using a clinical 3T MRI scanner (Skyra, Siemens). We used (1) fast low
angle shot (FLASH) cine: 9 cross-sectional slices, TE=2.83ms, TR=29.6ms, Flip
angle=12°, FOV=105x260 mm2, two averages, spatial resolution = 1.0x1.0x5.0mm,
and retrospective gating with 40 temporal frames and (2) conventional 4D Flow
sequence with Cartesian k-space sampling: VENC=25cm/s, TE=3.99ms, TR=77.28ms,
Flip angle=15°, FOW=208x319mm2, spatial resolution=2.4x2.4x2.4mm, retrospective
gating with 20 temporal frames. Cross-sectional areas of cine images were
measured by an automated boundary detection method (MATLAB R2021a) and
displaced volumes calculated by multiplying by slice thickness. Pressure data
was down-sampled for pressure-volume loop analysis. Measurements of 4D flow
were Eddy current corrected and visualized using Arterys (Arterys Inc., San
Francisco, USA).Results
The vessel wall was deformed three-ways by each bubble of the
actuator. The actuator is inflated for 300 ms, however the wall deformation
persists until 400ms (Figure 2A). Observable wall deformations extend to
+/- 1 cm from the actuator location (Figure 2B). The maximally induced
wall deformation was dependent on both actuator and loop pressure. At high
actuator pressures the vessel walls did not touch (Figure 2C). The
timing of the incidence of the volume displacement is earlier with higher
actuator pressures (Figure 3A). The maximally displaced volume (4.5–10 mL per beat) was positively related to the actuator pressure and
negatively related to the loop pressure (Figure 3). The trends in timing
were consistent with induced pressures measured on the benchtop. Combining the
pressure and volume time-series data, the actuator was shown to produce
positive stroke work. Qualitative measurements of 4D Flow indicate increased
forward flow with increased operating pressure (Figure 4).Discussion
We designed an MRI-compatible
control system with low ferromagnetic componentry, RF shielding, and one that did
not produce any imaging artifacts.
Traditional optical and benchtop methods
are more limited in characterizing the interaction of soft robots with soft
biological tissue. Particle image velocimetry offers an optical alternative,
but does not capture flow and tissue motion simultaneously and is challenging
upon large deformations.6 Here we show that MRI offers a comprehensive imaging-based
evaluation of an extra-aortic cardiac assist device.
The actuator pressure range tested shows
that the current design does not cause the vessel walls to touch or buckle upon
actuation. Larger actuator pressures displace more volume and may be preferred
for more effective cardiac support. As too large actuator pressures can result
in bladder failure, redesigning bladders with increased volume might help to
further increase displaced volumes. Finally, the effect of actuator pressure on
inflation timing may originate from the long pneumatic tubing used for remote
actuation.
In future work, we intend to use MRI
to enable patient-specific design optimization of extra-vessel soft robots.Conclusion
Using MRI, we
characterized the hemodynamic effects of an extra-aortic soft robot. Maximum
actuator operating pressure does not lead to buckling of the vessel wall and
induces forward flow.Acknowledgements
This study was supported by NIH R01 HL131823 to
DBE. Swiss National Research Foundation P2EZP2_188964 to SAD. References
1. Roche ET, Wohlfarth R,
Overvelde JTB, et al. A Bioinspired Soft Actuated Material. Adv Mater.
2014;26(8):1200-1206. doi:10.1002/adma.201304018
2.
Roche ET, Horvath MA, Wamala I, et
al. Soft robotic sleeve supports heart function. Sci Transl Med.
2017;9(373):eaaf3925. doi:10.1126/scitranslmed.aaf3925
3.
Almanza M, Clavica F, Chavanne J,
et al. Feasibility of a Dielectric Elastomer Augmented Aorta. Adv Sci.
2021;8(6):2001974. doi:10.1002/advs.202001974
4.
Rus D, Tolley MT. Design,
fabrication and control of soft robots. Nature. 2015;521(7553):467-475.
doi:10.1038/nature14543
5.
Chou C-PCC-P, Hannaford B. Static
and dynamic characteristics of McKibben pneumatic artificial muscles. Proc
1994 IEEE Int Conf Robot Autom. 1994;(3):281-286.
doi:10.1109/ROBOT.1994.350977
6.
Medero R, Hoffman C, Roldán-Alzate
A. Comparison of 4D Flow MRI and Particle Image Velocimetry Using an In Vitro
Carotid Bifurcation Model. Ann Biomed Eng. 2018;46(12):2112-2122.
doi:10.1007/s10439-018-02109-9