Sai Abitha Srinivas1,2, Christopher E Vaughn1,2, Jonathan B Martin1,2, and William A Grissom1,2,3,4
1Biomedical Engineering, Vanderbilt University, Nashville, TN, United States, 2Vanderbilt University Institute of imaging science, Nashville, TN, United States, 3Radiology, Vanderbilt University, Nashville, TN, United States, 4Electrical Engineering, Vanderbilt University, Nashville, TN, United States
Synopsis
Traditional B0 gradients have
several drawbacks including high acoustic noise, PNS, bulkiness, and high cost.
To address this, we demonstrate the use of Bloch-Siegert (BS) RF encoding for phase encoding using an optimized square root
solenoid with a Bucking coil for high efficiency encoding and a nested uniform
saddle coil for the imaging Tx/Rx coil at 47.5mT, a field strength that
is especially attractive due to its low SAR and accessibility. The coil performance was evaluated in simulation and
experimentally, and 2D BS phase encoded imaging and reconstructions were
performed using optimized ‘U’ shaped BS pulses.
Introduction
In conventional MRI, signal is
localized using B0 gradients that link spatial position to
signal frequency, but they are expensive, loud, induce PNS, and are prone to
breakage. B1+ gradients can alleviate these problems [1-8].
However, RF coils producing the appropriate B1+ gradients
are challenging to build. Solenoids with square-root pitch for Bloch-Siegert
(BS) encoding need to be made longer than the imaging volume to achieve the
desired square-root field in the imaging ROI, which limits their efficiency. To
address this, we propose an optimized square root solenoid with a Bucking coil for
encoding for BS spatial encoding, with a geometrically decoupled saddle coil inside
it for excitation and reception. We demonstrate a phase encoding experiment
with optimized frequency swept ‘U’-shaped BS pulses for phase encoding on a
47.5mT system with three reconstruction techniques.Methods
2D phase encoded imaging was implemented on a 47.5mT
scanner (Sigwa 48.7mT, Boston NMR, Boston, MA) shown in Figure 1A. The RF coils
were placed in a shielded box and remaining EMI was removed using EDITER [9] with
two external EMI detector coils.
Figure 1B shows the two-coil setup with 1) The 2.075MHz
litz-wire optimized square root coil for phase encoding, and 2) A uniform
saddle coil (OD of 3D printed former = 3.5cm, length = 12cm) for imaging. Optimization
of the square-root coil was done numerically by minimizing the B1+-field of a
regular square-root solenoid to be close to an ideal square-root B1+-field. The
winding pattern of an optimized coil is shown in Figure 1C. Figure 1D shows the
isolation (-36dB) between the geometrically decoupled solenoid and saddle coils.
Bucking coil windings were added to cancel fringe B1+-field
at the low B1+-field edge of the solenoid coil and make the
coil more efficient. This effect can be seen in Figure 2A wherein an analytic square
root coil is compared to the optimized coil with and without the bucking coil
windings (Total # of windings for each = 34, excluding the bucking coil). Since
the frequency and phase shifts applied to an object by BS encoding are
proportional to B1+2, Figure 2B
shows average B1+2 profiles across the ROI along with linear fits and their
R2. Figure 1C
shows Experimental B1+ maps which were obtained from a 11.5cm(L) x 3cm(D) tube
mineral oil phantom using the optimized ‘U’ shaped BS pulse [10] shown in
Figure 3A (BS frequency-offset =20KHz and Kbs[11] =43.5). Figure 1D shows the
average B1+2 profile across the tube phantom along with a linear fit
and its R2.
The same optimized BS pulse was used for phase encoding in
a 2D GRE sequence (Sequence parameters: TE=17ms, TR=534ms, NPE=67, NFE=128,
Naverages=5) shown in Figure 3B. The BS pulse amplitude was varied for
BS phase encoding (NBS=37). The uniform saddle coil was used for Tx/Rx
for imaging with an active T/R switch [12]. The optimized solenoid was used for
encoding. B0 gradients were used for frequency encoding along the Z
direction(kx). A scanner drift navigator sequence was interleaved with the
acqusition.
The Reconstruction pipeline is shown in Figure 4. Here,
the full phase encoded tube dataset was taken and then scanner drift corrected.
This resulting dataset was used to create an encoding matrix (E). A full
model-based reconstruction(R) done using the following equation:
$$ R = {(E'E + \lambda I)}^{-1} E S$$
Where, S is the hybrid (x-ky) data for a two-ball mineral(2cm) oil phantom and $$$\lambda$$$ is the
regularization parameter. Additionally, a least-squares NUFFT reconstruction was
performed by calculating the slopes of the BS phase encoded lines of the tube
phantom to determine the actual ky locations. A partial Fourier reconstruction was
performed using 70% of the ky lines which comprised all the positive encodes
and 40% of the negative encodes. MATLAB’s lsqnonlin (Mathworks,Natick,MA) was used to reconstruct the image, holding the phase image fixed to
match a zero-filled recon. Results
Linear
fits of B1+2 and their R2 in Figure 2B show that the optimized square root
solenoid had the highest linearity (R2=0.9987) which agreed with the
R2(0.9960) of the linear fit for the experimental results in Figure
2D. Since efficiency depends on the slope of the B1+2-field the y-intercepts are shown for each fit (retaining same slopes) to show
that the optimized coil with the bucking windings was 20% more efficient than
the one without the bucking windings. Figures 5A-D show the results from the
three types of reconstructions in comparison to a fully B0-encoded two-ball phantom image. Figure 5E shows two-ball
phantom 1D profiles at the same points in all the reconstruction methods and
from Figure 5F we can observe that the FWHM’s closely align with the B0-encoded image. The NUFFT reconstruction had the closest
fit to the B0-encoded image. Discussion
We
demonstrated a robust method to perform B1+ phase
encoding using an optimized square root solenoid. Adding the Bucking coil to
null the low-field, we obtained a 20% power efficiency improvement compared to
the optimized design. The interleaved drift navigator scans could be eliminated
by adding an FID measurement during the dead time between the excitation and BS
encoding pulse. In future work, we will linearize the phase encodes to enable
Cartesian FFT reconstruction. Acknowledgements
Funding
source R01 EB030414. We would like to thank Siddhant Gandhi for helpful
discussions about RF coil design. References
1. Hoult D. Rotating
frame selective pulses, J. Magn. Reson. 38 (1980) 369–374.
2. F Bloch and A Siegert. Magnetic resonance for
nonrotating fields. Phys Rev, 57:522–527, 1940.
3. R Kartausch, T Driessle, T Kampf, T C
Basse-Lusebrink, U C Hoelscher, P M Jakob, F Fidler, and X Helluy. Spatial
phase encoding exploiting the Bloch-Siegert shift effect. Magn Reson Mater Phy,
2013.
4. C J Hasselwander and W A Grissom. Bloch-Siegert
phase-encoded MRI with a single RF coil and frequency swept pulses. In
Proceedings 25th Scientific Meeting, International Society for Magnetic
Resonance in Medicine, Honolulu, p5047, 2017.
5. C Zimmerman Cooley. Portable Low-Cost
Magnetic Resonance Imaging. PhD thesis, Massachusetts Institute of Technology,
2014.
6. Torres, E, Froelich, T, Wang,
P, et al. B1-gradient–based
MRI using frequency-modulated Rabi-encoded echoes. Magn Reson
Med. 2021; 00: 1– 12. https://doi.org/10.1002/mrm.29002.
7. Jankiewicz M, Gore JC, Grissom WA. Improved encoding pulses for
Bloch-Siegert B1(+) mapping. J Magn Reson. 2013;226:79-87.
doi:10.1016/j.jmr.2012.11.004.
8. Khaligi M, Rutt B,
Kerr A. Adiabatic RF Pulse Design for Bloch-Siegert B1+ Mapping. Magn. Reson.
in Med. 70:829–835 (2013).
9. Srinivas, SA, Cauley, SF, Stockmann, JP, et al. External
Dynamic InTerference Estimation and Removal (EDITER) for low field MRI. Magn
Reson Med. 2021; 00: 1– 15. https://doi.org/10.1002/mrm.28992 .
10. Martin J, Vaughn
C, Griswold M, Grissom W. Bloch-Siegert |B1+|-Selective Excitation Pulses. In
Proceedings 29th Scientific Meeting, International Society for Magnetic
Resonance in Medicine. p 1911, 2021.
11.Sacolick L,
Wiesinger F, Hancu I, Vogel M. B1 mapping by Bloch-Siegert shift. Magn. Reson.
in Med. 63:1315-1322 (2010).
12. Straney D, Cooley CZ and Rosen MS. An Improved Power Handling Active
Transmit/Receive Switch for Low Field MRI using Reed Relays. In
Proceedings 29th Scientific Meeting, International Society for Magnetic
Resonance in Medicine. p 1395, 2021.