Simone Rumac1, Christopher W. Roy1, Jérôme Yerly1, Mariana B. L. Falcao1, Aurélien Bustin1,2,3, Mario Bacher1,4, Peter Speier4, Matthias Stuber1,5, and Ruud B. van Heeswijk1
1Department of Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland, 2IHU LIRYC, Electrophysiology and Heart Modeling Institute, INSERM U1045, Centre de recherche Cardio-Thoracique de Bordeaux, Université de Bordeaux, Bordeaux, France, 3Cardiovascular Imaging, Hôpital Cardiologique du Haut-Lévêque, CHU de Bordeaux, Bordeaux, France, 4Siemens Healthcare GmbH, Erlangen, Germany, 5CIBM Center for BioMedical Imaging, Lausanne, Switzerland, Lausanne, Switzerland
Synopsis
The
most commonly employed T2 mapping techniques are 2D and make use of
ECG-triggering. This may be a limitation in patients with variable heart rate
and complex three-dimensional conditions. To address these limitations, we here
propose an isotropic free-running 3D T2 mapping technique that avoids
ECG triggering by using Pilot Tone navigation.
In three healthy volunteers, our technique produced accurate isotropic T2 maps
when compared to 2D T2 prepared bSSFP (T2=41.1±4.8ms vs. 44.9±3.3ms,
respectively, p=0.1), and cardiac motion was successfully resolved.
Introduction
Cardiac
T2 mapping is routinely used in clinical MR protocols for the
assessment of acute edema.1 Most commonly employed T2
mapping techniques are breath-held single-slice 2D acquisitions. This coverage
can limit the sensitivity, especially in the case of myocardial pathologies
that have heterogeneous 3D patterns throughout the myocardium. To more
completely characterize these spatially complex disease patterns, several 3D T2
mapping techniques have therefore been proposed.2–6
Moreover, all current 2D and 3D T2 mapping techniques make use of electrocardiogram
(ECG) triggering, which may lead to unpredictable scan times, depends on
high-quality ECG signals, may fail in case of arrhythmia, and implies a
time-inefficient sampling of k-space. Most of these limitations might be
avoided by using a free-running acquisition with retrospective self-gating,7 but robust
ECG-free self-gating is very challenging when a wide range of image contrasts is
used, as is the case for mapping.
To
address these limitations, we here propose an isotropic free-running 3D T2
mapping technique that 1) utilizes Pilot Tone8 navigation
to obtain contrast-independent respiratory and cardiac signals, 2) robustly
corrects for respiratory motion through focused navigation (fNAV),9 3)
fully resolves cardiac motion, 4) produces accurate maps through compressed
sensing (CS) combined with patch-based denoising (HD-PROST).10Methods
The
acquisition (Figure 1) consists of a prototype free-running7 3D
radial GRE sequence with a phyllotaxis trajectory,11 field
of view=(220mm)3, isotropic spatial resolution=(1.5mm)3,
TR/TE=4.4/2.2ms, α=5°, four interleaved T2 preparation times T2-prep=0/25/40/55ms,
35200 readouts/T2-prep, and a total acquisition time of 11.3min. All
data were acquired on a 1.5T clinical scanner (MAGNETOM Sola, Siemens
Healthcare, Erlangen, Germany). Data-synchronized respiratory and cardiac
signals were extracted from the Pilot
Tone signal.12
The
respiratory motion was corrected using focused navigation (fNAV),9 which
estimates the amplitude of respiratory motion along all three spatial dimensions
in order to scale the respiratory motion
signals obtained from the Pilot Tone, and translationally corrects each k-space
readout prior to the CS reconstruction, thus reducing the dimensionality of the
dataset.
Source
images were reconstructed with total variation (TV) regularization along the cardiac
dimension, and denoising was subsequently performed by applying HD-PROST to
improve the T2 precision. Extended
phase graph (EPG)13 simulations
were used to generate a sequence-specific dictionary for a range of T2
values. A pixel-wise match of the
source images with this dictionary then resulted in the T2 map.
The
proposed technique was validated in an agar-NiCl2-gel phantom
designed to mimic the in-vivo range of myocardial relaxation times. Gold-standard
values were obtained using a turbo-spin-echo (TSE) sequence with 32 echo times (13-422ms)
and TR=10s. Linear regression and Bland-Altman analyses were performed to compare
our technique and routine breath-held T2-prepared T2
mapping14 to the
TSE relaxation times.
Finally,
both the proposed technique and the clinical routine scans were applied to the
heart of three healthy volunteers (age=33±8y, 1F) with IRB approval and written
informed consent. The myocardial T2 maps
were manually segmented, and a Student’s t-test was
used to compare the average values in the corresponding myocardial area to routine
T2-prepared mapping.14 We
measured the average myocardial T2 in all cardiac phases and
computed its standard deviation as a measure of precision.Results
The
phantom T2 maps demonstrated a high correlation with the gold
standard over the relevant T2 range (y=0.98x-4.6, R2=0.98,
Fig.2). A small (~6ms) underestimation was observed compared to the TSE T2
values, which was consistent in the range of interest. The Bland-Altman
analysis confirmed the bias but did not reveal other trends.
In
all in-vivo cases, and despite the periodic interference produced by the T2-prep
modules, Pilot Tone allowed a robust detection of both cardiac and respiratory
signals. The fNAV correction successfully aligned the respiratory phases as
evidenced by sharp lung-liver interfaces. The CS-based reconstruction produced well-aligned
source images with resolved cardiac motion.
The
resulting motion-resolved T2 maps were accurate when compared to
routine 2D maps (T2=41.1±4.8ms vs. 44.9±3.3ms, respectively, p=0.1).
Moreover, we found that the average T2 value remained coherent
across the cardiac cycle (Figure 4),
with the average standard deviation across all phases being lower than 3ms for
all considered subjects. Discussion
The
proposed technique showed high correlation with the gold-standard TSE
acquisition in the phantom, but a small T2 underestimation, which is
mostly likely due to stimulated echoes in the TSE and is consistent with what was previously
found in dictionary-matching-based techniques.15
The use of Pilot Tone allows for robust identification
of physiological signals regardless of periodic changes in source image
contrast. The fNAV correction proved to be a useful tool in order to reduce
dimensionality and therefore reduced the computational burden while retaining
all information. Finally, the T2 maps demonstrated high consistency across
the different cardiac phases, albeit in a small sample size. Therefore, further
optimization is required to improve both accuracy and precision in a larger cohort.
In conclusion, we successfully demonstrated the feasibility and preliminary
results of an isotropic free-running 3D T2 mapping technique.Acknowledgements
No acknowledgement found.References
- Messroghli, D. R. et al.
Clinical recommendations for cardiovascular magnetic resonance mapping of T1,
T2, T2* and extracellular volume: A consensus statement by the Society for
Cardiovascular Magnetic Resonance (SCMR) endorsed by the European Association for
Cardiovascular Imaging (EACVI). J Cardiovasc Magn Reson 19, 75
(2017).
-
van Heeswijk, R. B.
et al. Self-navigated isotropic three-dimensional cardiac T2 mapping. Magn.
Reson. Med. 73, 1549–54 (2015).
-
Ding, H. et al.
Three-dimensional whole-heart T2 mapping at 3T. Magn. Reson. Med. 74,
803–16 (2015).
-
Milotta, G. et
al. 3D Whole-heart free-breathing qBOOST-T2 mapping. Magn. Reson. Med.
83, 1673–1687 (2020).
-
Bustin, A. et
al. Accelerated free‐breathing whole‐heart 3D T 2 mapping with
high isotropic resolution. Magn. Reson. Med. 83, 988–1002 (2020).
-
Qi, H. et al.
Free-running simultaneous myocardial T1/T2 mapping and cine imaging with 3D
whole-heart coverage and isotropic spatial resolution. Magn. Reson. Imaging
63, 159–169 (2019).
-
Di Sopra, L.,
Piccini, D., Coppo, S., Stuber, M. & Yerly, J. An automated approach to
fully self-gated free-running cardiac and respiratory motion-resolved 5D
whole-heart MRI. Magn. Reson. Med. 82, 2118–2132 (2019).
-
Speier, P., Fenchel,
M. & Rehner, R. PT-Nav: A Novel Respiratory Navigation Method for
Continuous Acquisition Based on Modulation of a Pilot Tone on the MR-Receiver. Magn.
Reson. Mater. Phys. Biol. Med. 28, 1–135 (2015).
-
Roy, C. W. et
al. Motion compensated whole-heart coronary cardiovascular magnetic
resonance angiography using focused navigation (fNAV). J. Cardiovasc. Magn.
Reson. 23, 33 (2021).
-
Bustin, A. et al.
High-dimensionality undersampled patch-based reconstruction (HD-PROST) for
accelerated multi-contrast MRI. Magn. Reson. Med. 81, 3705–3719
(2019).
-
Piccini, D.,
Littmann, A., Nielles-Vallespin, S. & Zenge, M. O. Spiral phyllotaxis: the
natural way to construct a 3D radial trajectory in MRI. Magn. Reson. Med.
66, 1049–1056 (2011).
-
Falcão, M. B. L. et
al. Pilot tone navigation for respiratory and cardiac motion-resolved
free-running 5D flow MRI. Magn. Reson. Med. (2021)
doi:10.1002/mrm.29023.
-
Weigel, M. Extended
phase graphs: Dephasing, RF pulses, and echoes - Pure and simple. J. Magn.
Reson. Imaging 41, 266–295 (2015).
-
Giri, S. et al.
T2 quantification for improved detection of myocardial edema. J. Cardiovasc.
Magn. Reson. 11, 56 (2009).
-
Hamilton, J. I. et
al. Simultaneous Mapping of T1 and T2 Using Cardiac Magnetic Resonance
Fingerprinting in a Cohort of Healthy Subjects at 1.5T. J. Magn. Reson.
Imaging (2020) doi:10.1002/jmri.27155.