Brian Johnson1, Jonathan Chia1, Dave Hitt1, Rob Lay1, Tom Lowe1, Michael Pawlak1, John Penatzer1, James Snicer1, Marcie Stopchinski1, Gregory Thomas1, Kristen Williams1, and Paul Worthington1
1Philips Healthcare, Gainesville, FL, United States
Synopsis
Consistent high-quality magnetic resonance imaging (MRI) of the cervical
spine still remains challenging because of the inherent small anatomical
structures and degradation of image quality due to motion artifacts1. Motion
artifacts can arise from several sources including swallowing, respiration,
cerebrospinal fluid (CSF) pulsation, blood flow, and bulk patient movement. MRI of the cervical spine is one the highest performed
exams globally. Here we present a
complete cervical spine MRI protocol utilizing 2D and 3D radial k-space
sampling techniques which are inherently motion insensitive to increase image
quality.
Background
Consistent high-quality magnetic
resonance imaging (MRI) of the cervical spine still remains challenging because
of the inherent small anatomical structures and degradation of image quality due
to motion artifacts1.
Motion artifacts can arise from several sources including: swallowing,
respiration, cerebrospinal fluid (CSF) pulsation, blood flow, and bulk patient
movement2.
MRI of the cervical spine is one the highest performed exams globally3.
Therefore, optimization of cervical spine MRI protocols to improve IQ by
reducing motion artifacts can have a significant clinical impact and increase
diagnostic confidence. Radial k-space filling techniques such as PROPLLER (GE),
BLADE (Siemens), and MultiVane (Philips) help to reduce motion artifacts (figure
1) by oversampling the center of k-space and can help improve image quality in
the cervical spine1 (figure 2). In addition, a radial k-space trajectory
helps to improve signal-to-noise (SNR) by the repeated oversampling of the
center of k-space4 which can help to better visualize
the smaller structures of the cervical spine.
While use of sagittal T2-weighted radial acquisitions of the cervical
spine have been shown to be advantageous in reducing motion artifacts1 this only represents a single
sequence and contrast in a comprehensive MRI cervical spine exam. The American College of Radiology (ACR) guidance
for MRI of the cervical spine includes acquiring multiple image contrasts in
multiple imaging planes as well as utilization of 3D acquisitions to visualize
small structures5.
The ACR also recommends the use of motion suppression techniques where
possible5.
Here we present a complete MRI cervical spine protocol using both 2D and
3D radial k-space filling techniques to provide all necessary contrasts while
reducing motion artifacts. Teaching Point
Current commercial 2D radial imaging
offerings allow for the acquisition of turbo spin echo (TSE/FSE) and inversion
recovery (IR) sequences to provide T1-weighted, T2-weighted, proton density,
fluid attenuated inversion recovery (FLAIR), and diffusion weighted imaging
(DWI) contrasts. Fat saturation
techniques can also be applied to T2-weighted radial acquisitions to provide
water sensitive images that meet ACR requirements. However, fat saturated T2-weighted images may
fail compared to short tau inversion recovery (STIR) sequences over large a large
field of view, in the presence of metal implants, and variations in B0
homogeneity. B0 homogeneity is drastically affected
by all the motion in the cervical spine (i.e. swallowing, respiration, CSF and
blood flow, etc.) which will lead to incomplete fat saturation based on
spectral techniques. Simple modification
of the inversion time (TI), echo time (TE), repetition time (TR), and echo
train length (ETL) of the 2D radial FLAIR can be done to produce STIR images (Figure
3). Therefore, current commercially
available 2D radial techniques (PROPELLER, BLADE, and MultiVane) can provide
all necessary contrasts in the sagittal and axial plane. Advancements in radial techniques now allow
for commercially available 3D stack of stars sequences (figure 1) to be
acquired with and without fat saturation6.
These techniques include StarVIBE (Siemens) and 3D VANE (Philips) and are
most often used for T1-weighted free-breathing abdominal imaging. These sequences can be modified for cervical
spine imaging (figure 3) especially in cases when the use of gadolinium-based
contrast agents (GBCAs) are indicated.
While axial 2D T2 and T2*-weighted images can be currently acquired,
acquisition of a 3D radial T2*-weighted sequence has remained a missing element
to provide a full compliment of radial image contrasts. Modification of the current 3D VANE
implementation can be done to create a 3D T2*-weighted balanced fast field echo
(bFFE) image of the cervical spine (figure 4).Summary
By simple modification and optimization of imaging parameters, it is
possible to create a comprehensive motion-compensated cervical spine MRI
protocol with all image contrasts. With
the current ability to apply parallel imaging techniques (i.e. SENSE) to 2D and
3D radial acquisitions it makes the possibility of using a completely radial
approach to reduce motion artifacts in the cervical spine clinically feasible. Overall protocol and scan times may be longer
with radial acquisitions compared to Cartesian k-space trajectories with the
same resolution, but even if one sequence needs to be repeated due to motion
the comprehensive radial approach is faster (Table 1). Furthermore, the ability to acquire 3D
T2*-weighted images is novel and provides an unmet clinical need to allow for
the adoption of a completely motion compensated cervical MRI protocol. This approach can be applied at every field
strength providing even higher applicability given that motion artifacts are
more pronounced at 3T compared to 1.5T7. Based
on the prevalence of MRI spine exams and the degradation of image quality from
motion induced artifacts, adoption of a complete radial approach to imaging of the
cervical spine can increase overall IQ, enhance workflow, and diagnostic
confidence. Furthermore, with the
development of newer acceleration techniques like compressed sensing8 and deep learning based reconstruction9 the ability to speed up radial scanning and
implement this type of comprehensive motion-compensated imaging strategy across
all exams may become more clinically feasible. Acknowledgements
No acknowledgement found.References
1. Fellner
C, Menzel C, Fellner F, et al. BLADE in sagittal T2-weighted MR imaging of the
cervical spine. American journal of
neuroradiology. 2010;31(4):674-681.
2. Taber KH, Herrick RC, Weathers SW,
Kumar AJ, Schomer DF, Hayman LA. Pitfalls and artifacts encountered in clinical
MR imaging of the spine. Radiographics. 1998;18(6):1499-1521.
3. Anand PK, Shin DR, Saxena N, Memon
ML. Accelerated Reliability Growth Test for Magnetic Resonance Imaging System
Using Time-of-Flight Three-Dimensional Pulse Sequence. Diagnostics. 2019;9(4):164.
4. Pipe JG. Motion correction with
PROPELLER MRI: application to head motion and free‐breathing cardiac imaging. Magnetic Resonance in Medicine: An Official
Journal of the International Society for Magnetic Resonance in Medicine. 1999;42(5):963-969.
5. Radiology ACo. ACR-ASNR-SBCT-MR
Practice Parameter for the Performance of Magnetic Resonance Imaging (MRI) of
the Adult Spine. Res19-2012, Amended. 2014.
6. Chandarana H, Block TK, Rosenkrantz
AB, et al. Free-breathing radial 3D fat-suppressed T1-weighted gradient echo
sequence: a viable alternative for contrast-enhanced liver imaging in patients
unable to suspend respiration. Investigative
radiology. 2011;46(10):648-653.
7. Soher BJ, Dale BM, Merkle EM. A
review of MR physics: 3T versus 1.5 T. Magnetic
resonance imaging clinics of North America. 2007;15(3):277-290.
8. Lustig M, Donoho DL, Santos JM,
Pauly JM. Compressed sensing MRI. IEEE
signal processing magazine. 2008;25(2):72-82.
9. Lebel RM. Performance
characterization of a novel deep learning-based MR image reconstruction
pipeline. arXiv preprint arXiv:200806559.
2020.