Norio Hayashi1, Akio Ogura1, Atsuya Fuju2, Tomokazu Takeuchi3, Yusuke Sato4, Maiko Hashimoto5, and Masahiko Takahashi5
1Radiological Technology, Gunma Prefectural College of Health Sciences, Maebashi, Japan, 2Radiology, Kiryu Kosei General Hospital, Kiryu, Japan, 3Graduate School of Radiological Technology, Gunma Prefectural College of Health Sciences, Maebashi, Japan, 4Radiology, Gunma University Hospital, Maebashi, Japan, 5Isesaki Municipal Hospital, Isesaki, Japan
Synopsis
The purpose of this study was to evaluate the attraction force and image artifacts of 17 different leave-on powdered hair thickeners from 7 vendors. Displacement force was measured by the ASTM F2052, and MRI artifact was estimated by the ASTM F2052. There were some types of leave-on powdered hair thickener that do contain magnetic materials. They were highly susceptible to mechanical effects and will cause more image artifacts, so it must be removed before the examination.
Background
In recent years, more and more people are
using leave-on powdered hair thickeners instead of wigs to treat thinning hair.
The number of vendors and types of leave-on powdered hair thickeners has been
increasing rapidly. Sneha et al. reported a case in which sprinkled a leave-on
powdered hair thickener caused significant image artifacts [1], but there is no
detailed study on the mechanical effects and image artifacts of various leave-on
powdered hair thickeners. The purpose of this study was to evaluate the
attraction force and image artifacts of 17 different leave-on powdered hair thickeners
from 7 vendors.Methods
Seventeen types of leave-on powdered hair
thickeners from seven vendors were included in the study. Nine of them were
hair thickener types, and 8 were foundation types (Fig. 1). We used a 1.5 Tesla
MRI system (Ingenia 1.5T,Philips electronics) and a 3
Tesla MRI system (MAGNETOM Skyra, Siemens) to measure the displacement force
using the method described in ASTM F2052 [2]. After measuring 10 g of each
powder, the magnetically induced deflection force was calculated using the mean
deflection angle. MRI image artifacts were evaluated based on the method
described in ASTM F2119 [3]. MATLAB 2020b (Mathworks) was used to analyze the
image artifacts. All samples and a reference water sample were sealed in a
plastic container. The containers were sealed in an MRI sample measurement
phantom (90–401 type, NIKKO FINES). The phantom obtained T1-weighted images by
spin-echo sequence and T2 star-weighted images by gradient-echo sequence using
a 1.5-tesla MRI system and 3-tesla MRI system. From the MR images of each
sample obtained by each sequence, the area where the presence of the sample
caused a signal change of 30% or more compared to the image without the sample
was measured (ΔS[%]).Results
The deflection angle of all eight hair
thickener types was 0 degrees, and there was no mechanical effect on the change
force measured by ASTM F2052. On the other hand, for the foundation type, the
deflection angle of five samples was 90 degrees except for samples o, p, and q,
and was strongly affected by mechanical effects. No artifacts appeared in the
MR images of the hair thickener type samples even when all the samples were
imaged on the 1.5 Tesla MRI system and the 3 Tesla MRI system. On the other
hand, significant artifacts were observed on both MRI systems and sequences for
all foundation type samples except for samples o, p, and q. The ΔS according to
ASTM F2119 was less than 13% in the images obtained from the 1.5 Tesla MRI
system and less than 7% in the images obtained from the 3 Tesla MRI system for
the hair thickener type. On the other hand, for samples j, k, l, m, and r, the
percentage was more than 79% in the images obtained from the 1.5 Tesla MRI
system and more than 63% in the images obtained from the 3 Tesla MRI system.Conclusions
Leave-on powdered hair thickener looks similar,
but there are several suppliers and two types of hair thickening type and
foundation type. There are many types of leave-on powdered hair thickener that
do not contain magnetic materials. At the same time, there are also some types
of leave-on powdered hair thickener that do contain magnetic materials. The
leave-on powdered hair thickener that contains magnetic material is highly
susceptible to mechanical effects and will cause more image artifacts, so it
must be removed before the examination. The MRI technologist should check the
type of powdered hair thickener before the examination and take appropriate
action if necessary.Acknowledgements
No acknowledgement found.References
[1] Chenji S, Wilman AH, Mah D, Seres P,
Genge A, Kalra S. Hair product artifact in magnetic resonance imaging. Magn
Reson Imaging 2017;35:1-3.
[2] ASTM F2052-15, Standard Test Method for
Measurement of Magnetically Induced Displacement Force on Medical Devices in
the Magnetic Resonance Environment, ASTM International, West Conshohocken, PA,
2015, www.astm.org.
[3] ASTM F2119-07(2013), Standard Test
Method for Evaluation of MR Image Artifacts from Passive Implants, ASTM
International, West Conshohocken, PA, 2013, www.astm.org.