UTE Fat Suppression Technique for MSK
Hyungseok Jang1
1University of California, San Diego, La Jolla, CA, United States

Synopsis

Fat suppression can improve detection of lesions or reduce imaging artifacts, and therefore is often utilized in musculoskeletal (MSK) imaging. Ultrashort echo time MSK (UTE-MSK) imaging can also benefit from fat suppression to better visualize short T2 tissues and lesions. However, the conventional fat suppression approaches such as two-point Dixon or fat-saturation may not be effective in UTE imaging due to the signal attenuation in short T2 tissues. In this talk, we will review alternative approaches for fat suppression in UTE imaging, including single-point Dixon, UTE-IDEAL, single or dual IR based methods, multi-component analysis, and soft-hard pulse based water excitation.