Michael Bock1, Thomas Lottner1, and Ali Caglar Özen1,2
1Radiology - Medical Physics, University Medical Center, Medical Faculty, University Freiburg, Freiburg, Germany, 2German Consortium for Translational Cancer Research (DKTK), Freiburg, Germany
Synopsis
Energy harvesting of the RF fields has been
proposed to power small milliwatt RF amplifiers. In this work we demonstrate
that the time-varying gradient fields can deliver much higher power levels up
to a kW which we used to energize a conventional drip coffee machine. A large
pick-up coil was designed in which voltages of 100-200V were induced by rapidly
oscillating gradients. The coil was connected to a commercial coffee machine in
the magnet, and coffee was prepared in and with the MRI in just 5min.
Introduction
In many
MRI applications devices are operated in the MRI environment which
need to be supplied with electric energy. For example, in radio-frequency (RF)
coils low noise amplifiers are used that have a power consumption of up to
several hundred milliwatts, or ECG systems utilize power circuits to detect,
amplify and transmit the electric surface potentials of the heart for ECG
gating. Typically, the energy for these devices is either delivered directly
via a galvanic connection to a power source (e.g., via the RF cables), or a
nonmagnetic battery is integrated into the device. Direct galvanic connections
can act as antennae that are prone to RF heating [1], and the multiple
connections in high-density coil arrays can lead to bulky and inflexible cables
[2]. Batteries on the other hand are often limited in the available power that
they can deliver, and they require space in the device which is often very
limited.
To overcome this
limitation wireless energy harvesting has been suggested where the required
power is taken from the existing electromagnetic fields in the MRI. For RF
amplifiers energy harvesting has been realized by extracting energy from the RF
transmit field [3,4], and also first attempts to utilize the gradient power
have been suggested [5,6]. However, so far in all these realizations the
harvested power was always less than a Watt. In this work we demonstrate that
the gradient system of a clinical MRI alone can provide enough power on the
order of several hundred Watts – for this, we constructed a simple induction
coil, connected it to a coffee machine and cooked the first-ever MR-generated
coffee.Materials and Methods
Energy harvesting from a time-varying magnetic
field $$$B(t) = z_0 G_z (t)$$$ can be
realized using Faraday’s law of induction and a gradient $$$G_z (t)$$$:
$$U_{ind} = - \frac{d\Phi}{dt} = - N \frac{d(A(t) B(t))}{dt} = - N A \frac{d B(t)}{dt} = = - N A z_0 \frac{d G_z(t)}{dt}$$
Here, $$$U_{ind}$$$ is the
voltage induced in a coil with $$$N$$$ windings
and an area $$$A$$$, which is placed at the position $$$z_0$$$ away from
the iso-center of the magnet. The slew rate $$$s = \frac{d G_z(t)}{dt}$$$ is
limited by the maximum slew rate of the gradient system.
To demonstrate that $$$U_{ind}$$$
can be as high as typical line voltages
(110-220 V), a dedicated rectangular pick-up coil with the following parameters
was built from insulated copper wire: $$$N = $$$ 25 windings, $$$A = $$$ 34 cm x 34 cm = 0.12 m². In a first test
this coil was placed at $$$z_0 = $$$ 25 cm in a clinical 3 T MRI system (Siemens
PRISMA), which is equipped with a gradient system that offers a maximum slew
rate of $$$s_{max} = $$$ 200 T/m s - with these values a $$$U_{ind}$$$ of 144 V can be realized.
For energy harvesting the coil was placed on the
patient table of the MRI system at
$$$z_0 = $$$ = 25 cm, and a conventional drip coffee
machine (Siemens, nominal power consumption at 220 V: 1 kW) was connected
to the coil (Fig. 1). Before the experiment the magnetic parts of the
machine were removed (e.g., the over-temperature fuse) or replaced with
non-magnetic elements (copper ground plate), and the machine was filled with
water and ground espresso coffee (Intermezzo, Segafredo Zanetti S.P.A.,
Pianolo, Italy). Then, a gradient-only pulse sequence with a sinusoidal
gradient (amplitude: 60 mT/m, frequency: 3 kHz,
$$$s = $$$ 180 T/m s was executed for about 5 min until
the coffee was ready.Results and Discussion
With the simple energy harvesting setup a peak voltage of about 130 V could be created with a
clinical gradient system. At the relative high gradient oscillation frequency
of 3 kHz no vibration of the pick-up coil could be detected. The applied slew
rate was intentionally set slightly below the attainable hardware maximum to
ensure that the heating element of the coffee machine does not overheat, as the
magnetic over-temperature fuse had to be removed. Even though the applied power
is in this experiment about 2.9 times lower than nominal power consumption of 1
kW, substantially higher power levels could be produced than with previously
described harvesting setups.
Obviously, a major drawback of this
demonstration setup is its large size and the need for long cables. If the
pick-up coil is used to energize a separate unit such as a TMS generator, a
defibrillator or a disinfection device, such a coil setup could be integrated
advantageously into the device housing. Conclusion
High
quality coffee can be prepared with and in MRI systems.Acknowledgements
No acknowledgement found.References
[1] Ladd
ME, Quick HH. Reduction of resonant RF heating in intravascular catheters using
coaxial chokes. Magn Reson Med. 2000; 43(4):615-9.
[2] Nahova
L, et al. Perspectives in Wireless Radio
Frequency Coil Development for Magnetic Resonance Imaging. Front Phys 8:11,
2020
[3] Turner
W, et al. Nuclear magnetic resonance
energy harvesting for ultra-low power biomedical implants. 12th Ann
Wireless Microwave Tech Conf, WAMICON 2011
[4] Byron
K, et al. An RF-gated wireless power
transfer system for wireless MRI receive arrays. Concepts Magn Reson Part B
Magn Reson Eng. 2017 Oct;47B(4):e21360.
[5] Höfflin J, et al. Energy Harvesting with a figure-8 coil - towards energy autonomous MRI
detection. IMTEK, Univ. of Freiburg.
[6] Mandal
R,
et al. MRI
Powered and Triggered Current Stimulator for Concurrent Stimulation and MRI.
bioRxiv, p. 715805,
Jan. 2019