Austen Curcuru1, Deshan Yang1,2, and Michael Gach1,2,3
1Biomedical Engineering, Washinton University in Saint Louis, Saint Louis, MO, United States, 2Radiation Oncology, Washinton University in Saint Louis, Saint Louis, MO, United States, 3Radiology, Washinton University in Saint Louis, Saint Louis, MO, United States
Synopsis
MRI
linear accelerator hybrid systems use balanced steady state free precession sequences
(bSSFP) during radiotherapy due to the sequence’s high signal to noise ratio
and rapid acquisition times. However, bSSFP sequences are highly sensitive to off-resonance
effects.1 As a result, the position and velocity
of the radiotherapy gantry can cause imaging artifacts that can impact tumor
tracking and cause fluctuations in the main magnetic field that can result in
radiotherapy/MRI isocenter misalignments. Sequence parameters were adjusted
during the sequence run time to reduce these issues by integrating a navigator
into the bSSFP sequence and measuring B0 variations in real-time.
Purpose
Gantry rotation in MRI-guided radiotherapy (MR-IGRT) systems results in image
artifacts and spatial shifts due to electromagnetic interference (EMI) between
the MRI and linear accelerator (linac) subsystems (Fig. 1).2 Additionally, B0 in MR-IGRT systems
displays gantry angle position dependence due to the metal (e.g.,
ferromagnetic) components located in the gantry.3–5 Changes in B0 can result in: 1) imaging
artifacts that can impact tumor tracking; and 2) isocenter shifts that can
cause misalignments between the radiotherapy and MRI subsystems. The purpose of
this study was to measure and correct the B0 field variations in
real time associated with the gantry position in a balanced steady state free
precession (bSSFP) sequence.Methods
A hard pulse navigator (FA:70,
dwell time = 8 µs, complex points = 64) was added to a 2D Cartesian bSSFP
sequence (TE/TR: 1.7/3.4, FA: 70°, rBW: 558 Hz/pixel, single average) preceding
each image acquisition on a ViewRay 0.35 T MRI-Linac. ViewRay uses the Siemens IDEA/ICE
programming environment (VB19) and electronics. Measurements were performed using a
spherical 24 cm DSV phantom doped with 5 mM NiSO4 (T1/T2:
330/260 ms) using phased
array torso coils designed for use in the MR-IGRT system. Phase data from the first navigator was used
as a reference, and the central frequency offset was determined using phase
data from each subsequent navigator. The frequency offset was then passed back
to the sequence from the image calculation computer as a real time feedback
object and used to adjust the resonant frequency for the bSSFP image pulses and
receiver. B0 fluctuations as a result of EMI were measured by
allowing the gantry to rotate uninterrupted counterclockwise from 30° to 33°. In
addition, the B0 variations due to the stationary gantry from 30° to
33° in 15° increments Shimming for both experiments was performed at gantry
angle 30° prior to gantry rotation. The calculated central frequency offsets
from the bSSFP navigators were compared against shifts in the peak frequency
from a pair of FID sequences (rBW = 4Hz/point for EMI induced B0
offsets and rBW=1Hz/point for stationary gantry related B0 offsets).
Linear interpolation was used to align the time bases of the navigator and of
the FID data for comparison.Results and Discussion
Central frequency shifts calculated using the inserted
bSSFP navigators showed similar results as the reference FID data both while
the gantry was rotating (RMSE = 20.9 Hz, Fig. 2) and while the gantry was
stationary (RMSE = 15.5 Hz, Fig. 3). The EMI from gantry rotation resulted in B0
offsets about 10 times larger than the B0 shifts due to the stationary
gantry position. Both sets of B0 offsets displayed sinusoidal
behavior with a period of 60°. This behavior corresponds to the six 227 kg mu-metal
shield buckets on the MRI-Linac that are spaced 60° apart and house accelerator
components (Fig. 4). In 2020, ViewRay offered a gantry angle dependent B0
correction that is designed for their Step and Shoot radiation therapy delivery
in which the gantry is stationary during radiation delivery. Currently, faster radiation
delivery techniques are not possible for MR-IGRT systems due to gantry angle and time-dependent
EMI between the linac and MRI subsystems. The B0 navigator is being
developed to permit real-time B0 correction for arc therapy in which
radiation could be delivered during gantry rotation. Currently, the method is
restricted to Cartesian acquisitions. However, the navigator can be adapted to
radial acquisitions. ViewRay currently offers 4 frames/s Cartesian and 8
frames/s radial bSSFP (TrueFISP) sequences for real-time target tracking and
beam gating during radiotherapy.Conclusion
Navigators inserted into MR-IGRT sequences offer a promising way to
rapidly account for gantry angle related variations in B0 and to compensate
by adjusting sequence parameters in real time.Acknowledgements
No acknowledgement found.References
1. Bieri O, Scheffler K.
Fundamentals of balanced steady state free precession MRI. J Magn Reson
Imaging. 2013;38(1):2-11. doi:10.1002/jmri.24163
2. Gach HM, Curcuru AN,
Wittland EJ, et al. MRI quality control for low-field MR-IGRT systems: Lessons
learned. J Appl Clin Med Phys. 2019;20(10):53-66.
doi:10.1002/acm2.12713
3. Gach HM, Curcuru AN, Mutic
S, Kim T. B0 field homogeneity recommendations, specifications, and
measurement units for MRI in radiation therapy. Med Phys. 2020.
doi:10.1002/mp.14306
4. Kim T, Gu B, Maraghechi B,
et al. Characterizing MR Imaging isocenter variation in MRgRT. Biomed Phys
Eng Express. 2020;6(3). doi:10.1088/2057-1976/ab7bc6
5. Latifi K, Moros EG, Zhang
G, Harrison L, Feygelman V. A Method to Determine the Coincidence of
MRI-Guided Linac Radiation and Magnetic Isocenters. Technol Cancer Res
Treat. 2019;18:1-6. doi:10.1177/1533033819877986