Evangelia Kaza1, Jeffrey P Guenette2, Christian V Guthier1, Steven Hatch3, Alexander Marques3, Lisa Singer1, and Jonathan D Schoenfeld1
1Radiation Oncology, Brigham and Women’s Hospital, Dana-Farber Cancer Institute, Harvard Medical School, Boston, MA, United States, 2Division of Neuroradiology, Brigham and Women’s Hospital, Dana-Farber Cancer Institute, Harvard Medical School, Boston, MA, United States, 3Radiation Oncology, Brigham and Women’s Hospital, Dana-Farber Cancer Institute, Boston, MA, United States
Synopsis
Quality characteristics of healthy volunteer
head images (SNR, CNR, artifact size) were compared between two flexible coil
arrangements encompassing radiotherapy immobilization masks relative to a gold standard diagnostic
Head/Neck20 coil that cannot accommodate the masks. The novel arrangement of
two UltraFlexLarge18 coils provided higher SNR ratios and was more spacious
than the commercially recommended arrangement of two FlexLarge4 coils. Artifact
size and CNR ratios were similar for the two coil setups. Clinical application
of the UltraFlexLarge18 coil arrangement would be advantageous for head
and neck radiotherapy MRI simulations due to higher SNR and increased patient
comfort.
Introduction
Imaging for radiotherapy treatment planning is performed in the treatment
position to accurately localize tumor
and organs at risk. Head and neck treatment
planning would benefit from the superior soft tissue contrast provided by MRI,
but standard MR receive coils cannot accommodate thermoplastic masks used for
immobilization during irradiation. Flexible array coils can be arranged to encompass
these immobilization devices; however, compromises in image quality compared to
standard diagnostic coils should be minimized. The present healthy volunteer study
compared a novel arrangement of two
UltraFlexLarge18
(UFL18) coils against a commercially recommended (Qfix Insight) arrangement of
two FlexLarge4 (FL4) coils relative to a diagnostic Head/Neck20 (HN20) coil
for signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and artifact size
(AS).
Methods
Three consented healthy volunteers
(two females, one male) were imaged successively in the UFL18 and FL4 coil
setups with immobilization mask and in the HN20 coil without a mask in a 3T
Siemens Vida (Figure 1). Three axial
clinical sequences (T1-weighted Turbo Spin Echo (TSE), T2-weighted SPectral
Attenuated Inversion Recovery (SPAIR), Short-TI Inversion Recovery (STIR);
TR/TE/TI: 669/9.7, 4970/108, 3000/65/220 ms; pixel size: 0.68x0.68, 0.83x0.83,
0.94x0.94 mm2) covering the head from the chin to the top of the
brain using 50 4-mm thick slices were acquired for each coil setup.
Applying custom MATLAB software on each
data series, circular regions of interest (ROIs) were placed on a single slice
for the tongue base, parotids, pterygoid muscles, brainstem and thalamus;
rectangular ROIs were placed on the cerebellum and brain, and on background
areas presenting flow or eye motion artifacts (one, four, and three slices at
different levels, respectively). ROI size was equal between the matching slices
of different sequences and coils for the same subject. Example ROIs are presented
on Figure 2. The histogram spread of rectangular ROIs was calculated as a
measure of contrast in these ROIs [1] and represented artifact size (AS) for
the artifact ROIs. The mean signal of each circular or rectangular tissue ROI
was extracted. Signal- and contrast-to-noise ratios (SNR, CNR) of tissue ROIs
were computed using as noise estimate the average signal standard deviation
from four background ROIs at the corners of selected slices. Ratios of SNR, CNR
and AS parameters obtained with the UFL18 or FL4 arrangement relative to the
HN20 coil (parameter_UFL18/parameter_HN20, parameter_FL4/parameter_HN20) were
formed and compared using Wilcoxon rank-sum tests.Results
With the applied method AS was calculated for 6,
SNR for 18, and CNR for 10 ROIs for each sequence and coil. A total of 54 AS,
162 SNR and 90 CNR data points were obtained for each coil over all three
sequences and three volunteers. Consequently, an equal number of AS, SNR and CNR parameter ratios of UFL18 and FL4
relative to HN20 were produced. Graphs of the calculated parameter ratios over
all ROIs and summary box-and-whisker plots are shown on Figure 3. Wilcoxon
rank-sum tests indicated that SNR_UFL18/SNR_HN20 was significantly higher (Z=5.22,
p=1.7E-7) than SNR_FL4/SNR_HN20, while AS_UFL18/AS_HN20 and AS_FL4/AS_HN20 as
well as CNR_UFL18/CNR_HN20 and CNR_FL4/CNR_HN20 ratios had equal medians (Z=-1.87,
p=0.06 and Z=-0.22, p=0.82, respectively).Discussion
This work compared a new
UFL18 coil arrangement to a commercially
recommended FL4 coil arrangement using radiotherapy immobilization
equipment and clinical sequences covering the whole head. Multiple ROIs on
different slices allowed for thorough image quality assessments at different
areas inside the coils and provided a sufficient number of data points for
statistical analysis. As mean signal and contrast varies for different ROIs and
imaging sequences, the assessed image
quality characteristics were not compared directly between the two coil
setups but relative to a diagnostic HN20 coil considered as gold standard. A
newly developed MATLAB code allowed for expedited and equitable data processing.
The combination of two phased array coils
with 18 elements each demonstrated a significantly
higher SNR ratio than the combination of phased array coils having 4
elements. These
results agree with expectations since SNR theoretically increases with the
square root of coil element number [2],
although gains are less high in practice depending on geometrical and noise
factors [3]. Application of the UFL18 coil arrangement in MRI head and neck
radiotherapy treatment planning would allow for improved tumor and organ at
risk delineation thanks to higher SNR. Moreover, artifact size ratio was
overall smaller for UFL18, albeit not statistically
significant. Statistical tests revealed no significant CNR ratio differences between
the assessed coil setups.
In addition, the combination of the longer
UFL18 coils provided more room to better accommodate heads of larger size and volunteers
reported that it felt more comfortable and less claustrophobic than the tighter
FL4 coil setup. Conclusion
The new UFL18 arrangement was more spacious and yielded higher SNR
relative to a diagnostic head/neck coil than a commercially recommended FL4 coil arrangement, suggesting that its clinical application
would be advantageous for
head and neck
radiotherapy MRI simulations.
Acknowledgements
Kayes Techonology
Grant 2019References
1. Tripathi AK, Mukhopadhyay S, Dhara AK. Performance Metrics for Image Contrast. Proceedings of the 2011 International Conference on Image Information Processing (ICIIP 2011).
2. Wright SM, Wald LL. Theory and application of array coils in MR spectroscopy. NMR Biomed. 1997;10(8):394-410.
3. Fleischer CC, Zhong X, Mao H. Effects of proximity and noise level of phased array coil elements on overall signal-to-noise in parallel MR spectroscopy. Magn Reson Imaging. 2018;47:125-130.