Zhao Wei1,2,3, Hyungseok Jang1, Zubiad Ibrahim1, Mohammadamin Cheraghi1, Graeme M. Bydder1, Wenhui Yang2,3, and Ya-Jun Ma1
1Department of Radiology, UC San Diego, San Diego, CA, United States, 2Institute of Electrical Engineering, Chinese Academy of Sciences, Beijing, China, 3University of Chinese Academy of Sciences, Beijing, China
Synopsis
To improve the time efficiency for cortical bone
T1 measurement on clinical scanners, we proposed a new method which
fits the datasets of 3D ultrashort echo time actual flip angle imaging
(UTE-AFI) and UTE with a single TR (UTE-STR) simultaneously (UTE-AFI-STR). The
results show that the UTE-AFI-STR can measure the T1 of cortical
bone accurately and more efficiently than the UTE-AFI-VTR method.
Introduction
Ultrashort echo time (UTE) sequences have been used for cortical bone water imaging
and quantification in order to acquire useful magnetic resonance (MR) signals1. The T1 is a useful biomarker for the cortical bone porosity and age related deterioration
assessment and is also a sensitive biomarker for monitoring temperature change
in cortical bone2,3. Recently, we proposed a 3D UTE based actual flip angle imaging (AFI)-variable TR (VTR)
(UTE-AFI-VTR) method for accurate cortical bone T1 measurement4.
The UTE-AFI-VTR method can achieve accurate measurement of the T1 of cortical
bone without degradation from the inefficient RF excitation of clinical
scanners4. However, the scan time of the UTE-AFI-VTR sequence was
relatively long for clinical practice because the UTE-VTR dataset with multiple
TRs were required. To improve the time efficiency for accurate cortical bone T1
measurement, we proposed a generalized optimization framework which simultaneously
fits the UTE-AFI and the UTE with a single TR (UTE-STR) data sets to estimate T1.
This method would significantly reduce the total scan time for T1
measurement. Methods
The
features of the 3D UTE-AFI and UTE-VTR sequences are shown in Figure 14.
A short rectangular RF pulse (duration 150 μs) is used for signal excitation
and this is followed by a spiral trajectory data acquisition with 3D conical
view ordering in these two
sequences.
For
ultrashort T2 tissue excitation, the steady-state signals in TR1
and TR2 of the 3D UTE-AFI sequence conform to the signal models
described by Eqs. [1] and [2] in reference 4. The signal detected
by the 3D UTE sequence is described by Eq. [9] in reference 4. In addition, since
the equilibrium magnetization M0 and transverse magnetization
mapping function generated by the RF pulse Fxy
are independent of TR, they can be combined into a single unknown parameter
(e.g., κ).
As a result, when the RF
pulse and flip angle (FA) used in the AFI and VTR sequences are identical,
there are only three unknown parameters (i.e., κ, T1, Fz)
in these signal models to fit. Therefore, we
can use the optimization framework below to simultaneously fit (sf) UTE-AFI and
UTE-VTR (sf-UTE-AFI-VTR) data:
$$\left[\kappa~T_{1}~F_{z}\right]=\text{arg}\min\limits_{\kappa,T_{1},F_{z}}\left\{\sum_{i=1}^2[I_{AFI,i}-S_{AFI,i}]^2+\sum_{j=1}^N[I_{VTR,j}-S_{VTR,j}]^2\right\}~~~~~~~[1]$$
where $$$\text{arg}\min\limits_{x}\left(y(x)\right)$$$ means
finding a $$$x$$$ value
which can make $$$y(x)$$$ attain
its minimum value. $$$I_{AFI,i}$$$ ($$$i$$$ = 1 and 2) and $$$I_{VTR,j}$$$ ($$$j$$$ = 1,2,..., N) are
the sampled signal intensities from the UTE-AFI and UTE-VTR sequences
respectively. $$$S_{AFI,i}$$$ and $$$S_{VTR,j}$$$ are the signal models of the UTE-AFI and UTE-VTR
sequences respectively4. N is the total number of TRs. $$$\kappa=M_{0}F_{xy}\left(\alpha,\tau,T_{2}\right)$$$. This
framework simultaneously yields a solution for the three unknown parameters: κ, T1, and
Fz. Since
there are only three unknown parameters in Eq. [1], a single regular UTE
acquisition together with the other two UTE-AFI acquisitions is sufficient to
estimate them using the optimization framework. To save scan time, we therefore
propose a UTE-AFI-STR method (i.e., a combination of the UTE-AFI and UTE-STR
methods) to provide accurate measurement of T1.
Numerical simulations, ex vivo and in vivo human
cortical bone experiments were conducted to demonstrate the accuracy of the
proposed fast 3D UTE-AFI-STR method. In simulations, T1 maps
generated with the UTE-AFI-VTR, sf-UTE-AFI-VTR, and UTE-AFI-STR methods were
compared at different SNR levels and the error ratios were calculated. The
UTE-AFI and UTE-VTR sequences were performed on nine human cortical bone samples
(aged from 38 to 95 years, 5 females) and four healthy volunteers (35±16 years, 3 males) and
their T1 maps were calculated with the aforementioned three methods.
The T1 value difference errors between the UTE-AFI-STR and the UTE-AFI-VTR/sf-UTE-AFI-VTR
methods were calculated as well. The parameters used in the simulations were
the same as in vivo experiments and the sequence parameters of ex vivo and in
vivo experiments are shown in Table 1. The total scan times of the UTE-AFI-STR method were
about 54 minutes and 17 minutes less than the UTE-AFI-VTR/sf-UTE-AFI-VTR methods in the ex
vivo and in vivo experiments, respectively.Results and Discussion
As shown in
Figure 2, when the signal to noise ratio (SNR) of UTE-STR image was higher than
40, the T1 value derived from UTE-AFI-STR method was close to the
standard T1 (i.e., 220 ms) (error ratios ranged from -5% to 5%). This
observation shows that the fast UTE-AFI-STR method is accurate
for T1 measurement when the SNR of the UTE-STR image is higher than
40.
In Figure 3, the T1 maps of ex vivo
and in vivo cortical bones obtained with three methods looked similar. In Table
2, the T1 value difference ratios between range between -5.0% and
0.4%, and -5.0% to 7.6% for ex vivo and in vivo measurements, respectively. The
average cortical bone T1 for the four volunteers calculated using
the UTE-AFI-STR method was 246 ms, which is similar to previously reported values5-7. These observations demonstrate the accuracy and good agreement
of the T1 measurement of the UTE-AFI-STR method as compared with the
other two methods which require much more scan time.Conclusion
The proposed 3D UTE-AFI-STR method
provides accurate T1 mapping of cortical bone with improved time
efficiency compared with the UTE-AFI-VTR/sf-UTE-AFI-VTR methods.Acknowledgements
The
authors acknowledge grant support from GE Healthcare, NIH (R01AR068987,
R01NS092650, and R21AR075851) and scholarship support from the Joint Ph.D.
Training Program of the University of Chinese Academy of Sciences (UCAS).References
1. Du J, Bydder GM. Qualitative and
quantitative ultrashort-TE MRI of cortical bone. NMR Biomed. 2013;26:489–506.
2. Akbari A, Abbasi-Rad S, Rad HS. T1
correlates age: A short-TE MR relaxometry study in vivo on human cortical bone
free water at 1.5T. Bone 2016;83:17–22.
3. Han M, Rieke V, Scott SJ, et al.
Quantifying temperature-dependent T1 changes in cortical bone using ultrashort
echo-time MRI. Magn. Reson. Med. 2015;74:1548–1555.
4. Ma YJ, Lu X, Carl M, et al. Accurate
T1 mapping of short T2 tissues using a three-dimensional ultrashort echo time
cones actual flip angle imaging-variable repetition time (3D UTE-Cones AFI-VTR)
method. Magn. Reson. Med. 2018;80:598–608.
5. Yarnykh VL. Actual flip-angle imaging
in the pulsed steady state: A method for rapid three-dimensional mapping of the
transmitted radiofrequency field. Magn. Reson. Med. 2007;57:192–200.
6. Du J, Carl M, Bydder M, Takahashi A,
Chung CB, Bydder GM. Qualitative and quantitative ultrashort echo time (UTE)
imaging of cortical bone. J. Magn. Reson. 2010;207:304–311.
7. Chen J, Chang EY, Carl M, et al.
Measurement of bound and pore water T1 relaxation times in cortical bone using
three-dimensional ultrashort echo time cones sequences. Magn. Reson. Med.
2017;77:2136–2145.