Michael Mullen1 and Michael Garwood1
1Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, MN, United States
Synopsis
Clinical MRI sequences for imaging near metallic
implants are mainly multi-spectral approaches, with fast spin-echo acquisitions
and large acceleration factors to achieve clinically relevant scan times. The
authors previously reported a broadband, low flip angle method at 1.5T to image
with large field inhomogeneity, such as near metallic implants, quickly
relative to non-spatially selective multispectral approaches. Herein it is demonstrated
that this approach, dual polarity missing pulse steady-state free precession,
can achieve high spatial resolutions at 3T with a large 3D FOV in a clinically
relevant scan time, ~8.74 minutes.
Purpose
A small number of sequences are currently available
for MR imaging near metallic implants in a clinically relevant scan time (1-3) due to the large induced field inhomogeneities. While the accuracy and
efficiency of dual polarity missing pulse steady-state free precession
(MP-SSFP) has previously been demonstrated at 1.5T, the spatial resolution was
2mm isotropic, the scan time was 14 minutes, and the signal-to-noise ratio (SNR) was approximately 12 (4). Here, a high spatial resolution of 1mm isotropic is
achieved with a large, 256mm3 FOV. The MP-SSFP sequence parameters
have been tuned and experiments were performed at 3T to boost SNR. Parallel
imaging was employed to attain clinically relevant scan times, using each gradient
polarity acquisition as a set of virtual coils to improve the reconstruction.Methods
The Missing Pulse Steady-State Free Precession (MP-SSFP)
(5) sequence was used, whereby a train of n small-flip angle pulses are applied to
achieve a steady-state. Data are acquired in place of the n+1 pulse, where an echo forms. This sequence achieves radiofrequency
(RF) refocusing with low peak amplitude. The dual polarity encoding was implemented
such that acquisition of the same phase encoding line for each gradient
polarity was separated by a single TR to limit motion artifacts. The
displacement field was estimated by aligning the intensity-corrected, resampled
magnitude images resulting from the opposing gradient polarities,
parameterizing the displacement in terms of 3D B-splines (6).
Phantom experiments were performed on a 3T Siemens Prisma scanner using a
32-channel head coil. The phantom
consisted of agar with one stainless steel and one titanium screw embedded
inside. Acquisitions were fully sampled experimentally and retroactively
undersampled in the phase encoding dimensions (R=3x2) for this demonstration. The
MP-SSFP sequence used n=4 pulses, TR=24ms, and 650 Hz/pixel receiver bandwidth.
Hyperbolic secant (7) pulses were used, with time-bandwidth product 10, 500µs duration, 20kHz bandwidth, and 20˚ flip angle. The field-of-view (FOV) was 256x256x256
mm3 with 1mm isotropic resolution, and total scan time of 52.43
minutes. The retroactively undersampled acquisition would have taken 8.74
minutes.
GRAPPA (8) was used for parallel imaging, with a 24x24
autocalibrating signal (ACS) region in the phase encoded dimensions. The ACS
data were not included in the final reconstruction. The 2D GRAPPA
reconstructions were performed per plane in the frequency encoding direction
following Fourier Transform in that dimension of the data. The reconstruction
was performed with data from each polarity concatenated along the coil dimension to increase the
number of effective coils, similar to the virtual coil technique introduced by
Blaimer et al. (9). A noise-only scan was used to correct for noise correlations
between coils. The g-factor was determined using the analytical method introduced
by Breuer et al. (10). The GRAPPA-reconstructed images were used as inputs
for the distortion correction algorithm, which used 3mm spline knot spacing to
estimate the displacement field.Results
A comparison between the fully sampled and
GRAPPA-reconstructed images is shown in Fig. 1 as one cross section along the
frequency encoded dimension. The residual aliasing artifacts are not
significant, as seen in the difference image of Fig. 1. Noise amplification is
pronounced, (g-factor in Fig. 2), but the SNR remains acceptable in the
reconstructed image, with a peak value of ~36 in the cross section shown in
Fig. 1, as determined using an SNR unit reconstruction (11).
The distortion corrected image is
shown in Fig. 3 with a B0 estimate in voxel units. The geometric and intensity distortions
are clearly corrected, but there is too little spatial encoding closer
to the metal to accurately reconstruct the image, due to large intrinsic field
gradients.Discussion
The primary disadvantage of the MP-SSFP technique described
here is the absence of slab selection, which would permit a shorter scan time.
While slab selection is possible, phase encoding remains necessary in the
spatially-selected dimension (1,3) to resolve slab distortions. On the contrary, the short TR of MP-SSFP alleviates
the need for spatial selection.
A common problem with Partial Fourier (PF) occurs in
regions with rapid image phase variation, due to the inability to correct such
phase variation with the symmetrically sampled portion of k-space. In contrast
to MSI sequences, PF acceleration is suboptimal with dual polarity MP-SSFP,
which simultaneously excites a broad bandwidth, leading to rapid phase
variations near metallic implants. MSI sequences avoid this issue by only
imaging a narrow bandwidth at any time. Hence, a small PF factor is acceptable
but is not useful for substantial acquisition time reduction when using MP-SSFP
as performed herein.
Compared to 1.5T, imaging near metallic implants at 3T
requires higher bandwidth pulses to excite spins in close proximity to the metal. Even
if excited, a large frequency encoding gradient is required to overcome the
intrinsic field inhomogeneity sufficiently for the distortion correction
technique presented here to successfully recover the undistorted image.
Future work will examine 2D CAIPIRINHA (12) in conjunction with dual polarity MP-SSFP to decrease
noise amplification in GRAPPA. A unique feature which will permit a wider range
of possible undersampling schemes is the sequential acquisition of each
polarity image, so that the k-space sampling patterns for each set of virtual
coils need not be the same.Acknowledgements
This
work was supported by the National Institutes of Health grants U01 EB025153 and
P41 EB025144-01.References
1. Koch
KM, Brau AC, Chen W, Gold GE, Hargreaves BA, Koff M, McKinnon GC, Potter HG,
King KF. Imaging near metal with a MAVRIC-SEMAC hybrid. Magn Reson Med
2011;65(1):71-82.
2. Koch KM, Lorbiecki JE, Hinks RS,
King KF. A multispectral three-dimensional acquisition technique for imaging
near metal implants. Magn Reson Med 2009;61(2):381-390.
3. Lu W, Pauly KB, Gold GE, Pauly JM,
Hargreaves BA. SEMAC: Slice Encoding for Metal Artifact Correction in MRI. Magn Reson Med 2009;62(1):66-76.
4. Mullen M, Garwood M.
Reverse-Encoding MRI Near Metallic Implants Using Missing Pulse Steady-State
Free Precession. In: ISMRM & SMRT Virtual Conference and
Exhibition, 2020. p 3404.
5. Patz S, Wong ST, Roos MS. Missing
pulse steady-state free precession. Magn Reson Med 1989;10(2):194-209.
6. Skare S, Andersson JL. Correction of
MR image distortions induced by metallic objects using a 3D cubic B-spline
basis set: application to stereotactic surgical planning. Magn Reson Med
2005;54(1):169-181.
7. Silver MS, Joseph RI, Hoult DI.
Highly selective π/2 and π pulse generation. Journal of Magnetic Resonance
(1969) 1984;59(2):347-351.
8. Griswold MA, Jakob PM, Heidemann RM,
Nittka M, Jellus V, Wang J, Kiefer B, Haase A. Generalized autocalibrating
partially parallel acquisitions (GRAPPA). Magn Reson Med 2002;47(6):1202-1210.
9. Blaimer M, Gutberlet M, Kellman P,
Breuer FA, Kostler H, Griswold MA. Virtual coil concept for improved parallel
MRI employing conjugate symmetric signals. Magn Reson Med 2009;61(1):93-102.
10. Breuer FA, Kannengiesser SA, Blaimer
M, Seiberlich N, Jakob PM, Griswold MA. General formulation for quantitative
G-factor calculation in GRAPPA reconstructions. Magn Reson Med
2009;62(3):739-746.
11. Kellman P, McVeigh ER. Image
reconstruction in SNR units: a general method for SNR measurement. Magn Reson
Med 2005;54(6):1439-1447.
12. Breuer FA, Blaimer M, Mueller MF,
Seiberlich N, Heidemann RM, Griswold MA, Jakob PM. Controlled aliasing in
volumetric parallel imaging (2D CAIPIRINHA). Magn Reson Med 2006;55(3):549-556.