Hao Chen1, Shiwei Yang1, Sijie Zhong1, and Zhiyong Zhang1
1School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China
Synopsis
EPI is a versatile MRI acquisition method, which
can ultrafast yield an image and has been widely applied. However, EPI is
well-known sensitive to B0
inhomogeneities and susceptibility effects, which leading to geometrical image
distortions. In order to reduce the geometric distortion on EPI images, this
work presents a new interleaved single-shot EPI method, in which ky-t line
can be sampled with higher slope rate such that the geometric distortion can be
alleviated. The experimental results of phantom and human brain at 3T
demonstrate the capability of the new method.
Introduction
Echo planar imaging (EPI) is a powerful
technique to acquire MRI image rapidly, which can produce a 2D image within a
single shot. EPI has been widely used in the investigation of diffusion,
perfusion and functional imaging. However, as EPI takes a few tens of
milliseconds to complete k-space coverage, the nuisance phase caused by B0
inhomogeneities and susceptibility effects accumulates during this process, thus
giving rise to geometric distortion on images. Several methods based on
multi-shot acquisition, such Top-up, PSF-EPI and EPTI, have been proposed to
address this issue.1-3 However, multi-shot acquisition is sensitive
to the shot-to-shot phase variation, which may be induced by motion or pulsation.
Here we present a new single-shot EPI method with interleaved phase encoding
lines (interleaved single-shot EPI, isEPI), to reduce the phase error accumulation
during spatial encoding, therefore improve the distortion on images.Methods
Conventional gradient-echo EPI (Fig. 1a) uses an RF
for excitation and prolonged echo train to sample the k space (Fig. 1b, left),
resulting in a consecutive ky-t space (Fig. 1b, right). Here, we use
two RF pulses to stimulate two echo trains (Virtual Shot #1 and Virtual Shot
#2) to sample k space (Fig. 1c). Different from single-shot multi-echo EPI
methods for multiple-contrast acquisition,4 the two echo trains can
be interleaved in the PE direction to cover the full k space (Fig. 1d), because
the blip gradients are double-sized to produce a larger leap on ky space
and the echo time make T2* decay consecutive along the PE direction. As shown
in Fig. 1d, the interleaved ky-t line is 2-fold shortened in t
direction compared to the conventional EPI. To make the amplitude of two visual
shots equal, the flip angle of two RF pulses (α
and β) should satisfy the
condition $$$\cos^2\frac{\beta}{2}\sin\alpha=\sin\beta\cos\alpha$$$.
All
experiments were executed on a UIH uMR790 3T MRI scanner with a 24-channel
head-neck coil (United Imaging Healthcare, Shanghai, China). The American
College of Radiology (ACR) MRI phantom (J.M. Specialty Parts, San Diego,
California) and a healthy human subject were scanned. The acquisition
parameters for ACR phantom are as follows: FOV=220x220 mm2; slice
thickness = 3 mm; TR/TE = 4000/74 ms; echo spacing = 0.61 ms; matrix size = 128
x 128 (64 PE lines for each echo train). The flip angle of α and β are set to 45° and
53°, respectively. To demonstrate the
performance of isEPI, the
currents in shimming coils were altered to yield an inhomogeneous magnetic
field with 31 Hz line width on water resonance, while the line width is 10 Hz
on shimmed water resonance. In human brain experiments, we use
double-sized blip gradient for 2-fold acceleration and set TE = 44 ms. The
slice thickness is 4 mm for human brain, and the other settings are same to ACR
phantom. Conventional gradient-echo single shot EPI and 2D FSE data were also
obtained on the same phantom and human brain for the comparison. Results and Discussion
The experimental results of phantom in inhomogeneous
field are listed in Fig. 2. The conventional EPI image is subjected to severe
geometric distortion and localization shift along phase encoding direction. Compared
to conventional EPI, the outlines of isEPI image are much closer to the
boundaries of distortion-free FSE image, and much less geometric distortion. In
the human brain experiments, the isEPI image (Fig. 3, left column) exhibits
less geometric distortion than conventional EPI (Fig. 3, right column). The
isEPI method shows the ability to reduce the geometric distortion due to T2*
effects in inhomogeneous field, since the T2* phase accumulation is reduced.
To
obtain two echo trains with identical amplitude, we applied two RF pulses for
stimulation in isEPI. The delay between two RF pulses is equal to the length of
one echo train, leading to restriction on minimum TE. As the optimal flip angle
of the two RF pulses is around 45° and 53° degree, the SNR in the isEPI is about
half of that in conventional EPI which uses 90° RF pulse for excitation. Partial
k-space technique or center-out k space trajectory may help to reduce the
minimum TE and increase the signal intensity.5-6Conclusions
In this work, we present a new EPI method to improve
the geometric distortion in EPI image. With two echo trains to cover the
k space, the nuisance phase accumulation due to B0 inhomogeneities
and susceptibility effects is reduced in ky-t space. The experimental
results validate its ability to obtain images with reduced distortion by a
single shot. It has the potential to be expanded to acquire diffusion weighting
images or perfusion images.Acknowledgements
This work is supported by National Science
Foundation of China (No. 62001290) and Sponsored by Shanghai Sailing Program
(20YF1420900).References
1. Andersson JL, Skare S, Ashburner J. How to
correct susceptibility distortions in spin-echo echo-planar images: application
to diffusion tensor imaging. NeuroImage. 2003;20:870-888.
2. Zeng H, Constable RT. Image distortion
correction in EPI: comparison of field mapping with point spread function
mapping. Magn Reson Med. 2002;48:137-146.
3. Wang F, Dong Z, Reese T, et al. Echo planar
time-resolved imaging (EPTI). Magn Reson Med. 2019;81:3599-3615.
4. Posse S, Wiese S, Gembris D, et al. Enhancement
of BOLD contrast sensitivity by single-shot multi-echo functional MR imaging.
Magn Reson Med. 1999;42:87-97.
5. Jesmanowicz A, Bandettini PA, Hyde JS.
Single-shot half k-space high-resolution gradient-recalled EPI for fMRI at 3
Tesla. Magn Reson Med. 1998;40:754-762.
6. Chen X, Zhu A, Du Y. Centre-out (COEPI): A fast
single-shot imaging technique with a short TE. Magn Reson Med. 2020;84:787-799.