Alecio F. Lombardi1,2, Zhao Wei1, Hyungseok Jang1, Saeed Jerban1, Lillian Gong1, Jiang Du1, Eric Y. Chang1,2, and Ya-Jun Ma1
1Radiology, University of California, San Diego, CA, United States, 2Radiology Service, Veterans Affairs, San Diego, CA, United States
Synopsis
To overcome the limitations of
imaging short T2/T2* tissues with conventional MRI and
increase contrast between the osteochondral junction and adjacent tissues, we developed
a 3D dual adiabatic inversion recovery prepared ultrashort echo times cones (3D
DIR-UTE-Cones) sequence for volumetric imaging at a 3T scanner. We expected the proposed DIR-UTE-Cones sequence to generate
higher OCJ contrast than the IR-FS-UTE Cones, especially in between OCJ and fat.
Introduction
The osteochondral junction (OCJ) may be related to early changes
of knee osteoarthritis, but tissues are invisible with clinical sequences due
to their short T2/T2*. A few UTE sequences were developed to study this region,
including UTE dual-echo subtraction and inversion recovery (IR)-prepared UTE
techniques (1-5). UTE dual-echo acquisition is time-efficient but suffers from
a low signal-to-noise ratio (SNR) (2). Further, the high proton density and
relatively short T2* from fat reduce the contrast between the cartilage and
subchondral bone (6). A single IR-prepared UTE sequence, incorporated with a
fat saturation module (IR-FS-UTE), was proposed to suppress signals from
superficial cartilage (SC) and subchondral bone fat to highlight the OCJ region
(4). When combined with 3D sequences, a multi-spoke acquisition was employed to
reduce total scan time (4). However, the fat suppression efficiency is
decreased as the number of spokes (Nsp) in each TR increases.
A decade ago, a 2D UTE sequence with a dual inversion recovery preparation
(DIR-UTE) was developed for imaging of short T2 tissues. The two IR pulses
centered on water and fat are used to suppress the signals from
superficial/middle layers of cartilage and subchondral bone fat (3). However,
this sequence has mainly been used for ex vivo small sample imaging due to its
sensitivity to eddy currents and out-of-slice signal excitation artifacts (7).
This study combined the DIR scheme with a 3D UTE Cones sequence (DIR-UTE-Cones)
for high-resolution and high-contrast imaging of OCJ. Multi-spoke acquisition
was employed to reduce total scan time. The 3D IR-FS-UTE Cones sequence was
scanned for comparison. We expected the proposed DIR-UTE-Cones sequence to
generate higher OCJ contrast than the IR-FS-UTE Cones, especially in between
OCJ and fat.Methods
The DIR-UTE-Cones sequence combined a 3D UTE-Cones sequence with
two adiabatic IR pulses. A series of spokes (Nsp) were acquired after each DIR
preparation to accelerate the scan (Figure 1). As seen in Figure 1, the
magnetization of the superficial/middle layers of cartilage and fat were inverted
by two adiabatic IR pulses. These pulses were centered on the spectrum
frequencies of water (0 Hz) and fat (-440 Hz) and were applied at different
times so that they could achieve the nulling points at the same time after
magnetization recovery. In contrast, the short T2 magnetization of OCJ is not
inverted by the adiabatic IR pulses but saturated. When the DIR-UTE-Cones
acquisition starts around the nulling point of the superficial/middle layers of
cartilage and fat, the OCJ tissue presents with a high signal.
The 3D DIR-UTE-Cones sequence was applied to a phantom (Figure 2) comprised of
a tube with vegetable oil (T1 ~ 260 ms; simulating fat), a tube of distilled
water containing MnCl2 (0.015g/L, T1 ~ 900ms; T2 ~ 120ms, simulating
superficial and middle layers of cartilage), and a tube with distilled water
containing MnCl2 in a different concentration (3.6 g/L, T1 ~ 30ms; T2 ~ 0.5ms,
simulating OCJ), all in a container filled with agarose gel (T1 ~ 2600ms). The
T1s of the MnCl2 solutions, fat, and agarose gel were assessed using 3D
UTE-Cones actual flip angle (FA), and variable FA sequence described previously
(8, 9). For comparison, fast spoiled gradient echo (FSPGR) and four additional
UTE protocols were applied: 3D IR-UTE-Cones, a proton density-weighted UTE
(PDw-UTE) with fat saturation (FatSat), a PDw-UTE without FatSat, and a
T1-weighted UTE (T1w-UTE, high FA with short TR). Finally, 3D DIR-UTE-Cones,
IR-FS-UTE Cones, FSPGR, and clinical 2D T2-weighted fast spin-echo (T2w-FSE)
sequences were performed in four knee joints from four healthy male volunteers
and four patients with osteoarthritis (OA).Results and Discussion
Figure 2 shows the 3D DIR-UTE-Cones sequence (E) on the phantom,
with simultaneous suppression of the signal from vegetable oil and the solution
representing the superficial cartilage (SC), highlighting the signal from the
solution representing the OCJ region more efficiently than the 3D IR-UTE-UTE
Cones sequence (F). In (A, FSPGR), we observe low signal from the solution
representing SC due to its limitations in imaging short T2 tissue. The PDw- and
T1w- UTE sequences capture signal from the solutions with short T2 but cannot
differentiate between the SC and OCJ region.
Figure 3 shows knee images from two healthy volunteers. On T2w-FSE and FSPGR,
the signal in the OCJ is not detected due to its fast signal decay. Using the
IR-FS-UTE Cones sequence, suppressing the superficial layer of cartilage
highlights the OCJ region, but the subchondral bone marrow fat is still present
with a relatively high signal. DIR-UTE-Cones suppress signals from both SC and
subchondral bone, increasing the contrast between these tissues and the OCJ.
Figure 4 shows the knee images from two patients with OA, demonstrating areas
of deep cartilage erosions and subchondral bone exposure where the high signal
from deep and calcified layers of cartilage is interrupted on the DIR-UTE
sequence.Conclusion
The 3D-DIR-UTE-Cones sequence can produce high-resolution and
high-contrast imaging of the OCJ region of the knee in vivo.Acknowledgements
The authors acknowledge grant support from the NIH (R01 AR062581, R01 AR068987, R21AR075851), Veterans Affairs (I01RX002604 and I01CX001388), and GE Healthcare. References
1. Du
J, Takahashi AM, Bae WC, Chung CB, Bydder GM. Dual inversion recovery,
ultrashort echo time (DIR UTE) imaging: creating high contrast for short‐T2 species.
Magnetic Resonance in Medicine: An Official Journal of the International
Society for Magnetic Resonance in Medicine. 2010;63(2):447-55.
2. Lee
YH, Kim S, Song H-T, Kim I, Suh J-S. Weighted subtraction in 3D ultrashort echo
time (UTE) imaging for visualization of short T2 tissues of the knee. Acta
radiologica. 2014;55(4):454-61.
3. Du
J, Carl M, Bae WC, Statum S, Chang EY, Bydder GM, et al. Dual inversion
recovery ultrashort echo time (DIR-UTE) imaging and quantification of the zone
of calcified cartilage (ZCC). Osteoarthritis Cartilage. 2013;21(1):77-85. Epub
2012/10/03. doi: 10.1016/j.joca.2012.09.009. PubMed PMID: 23025927; PubMed
Central PMCID: PMCPMC4051156.
4. Ma
YJ, Jerban S, Carl M, Wan L, Guo T, Jang H, et al. Imaging of the region of the
osteochondral junction (OCJ) using a 3D adiabatic inversion recovery prepared
ultrashort echo time cones (3D IR‐UTE‐cones) sequence at 3 T. NMR in Biomedicine.
2019;32(5):e4080.
5. Mackay
JW, Low SBL, Houston GC, Toms AP. Ultrashort TE evaluation of the osteochondral
junction in vivo: a feasibility study. The British Journal of Radiology.
2016;89(1059):20150493. doi: 10.1259/bjr.20150493. PubMed PMID: 26781345.
6. Kühn
J-P, Hernando D, Meffert PJ, Reeder S, Hosten N, Laqua R, et al. Proton-density
fat fraction and simultaneous R2* estimation as an MRI tool for assessment of
osteoporosis. European radiology. 2013;23(12):3432-9.
7. Lu
A, Daniel BL, Pauly JM, Butts Pauly K. Improved slice selection for R2* mapping
during cryoablation with eddy current compensation. Journal of Magnetic
Resonance Imaging: An Official Journal of the International Society for
Magnetic Resonance in Medicine. 2008;28(1):190-8.
8. Ma
Y-J, Lu X, Carl M, Zhu Y, Szeverenyi NM, Bydder GM, et al. Accurate T1 mapping
of short T2 tissues using a three-dimensional ultrashort echo time cones actual
flip angle imaging-variable repetition time (3D UTE-Cones AFI-VTR) method.
Magnetic Resonance in Medicine. 2018;80(2):598-608. doi: https://doi.org/10.1002/mrm.27066.
9. Ma
YJ, Zhao W, Wan L, Guo T, Searleman A, Jang H, et al. Whole knee joint T1
values measured in vivo at 3T by combined 3D ultrashort echo time cones actual
flip angle and variable flip angle methods. Magnetic resonance in medicine.
2019;81(3):1634-44.