Hyungseok Jang1, Yajun Ma1, Michael Carl2, Saeed Jerban1, Roland Lee1, Eric Y Chang1,3, Jody Corey-Bloom1, and Jiang Du1
1University of California, San Diego, San Diego, CA, United States, 2GE Healthcare, San Diego, CA, United States, 3Veterans Affairs San Diego Healthcare System, San Diego, CA, United States
Synopsis
It
is challenging to directly image myelin due to its extremely short T2*
(<300µs at 3T). Adiabatic Inversion Recovery prepared Ultrashort Echo Time
(IR-UTE) imaging has been proposed for direct myelin imaging in human brain. More
recently, Short Repetition Time Adiabatic Inversion Recovery (STAIR) has been
proposed as a novel contrast mechanism for myelin imaging with improved
suppression of long T2 signal. In this study, we explored feasibility of STAIR
based Zero Echo Time (STAIR-ZTE) combined with an amplitude- and phase-modulated
RF pulse and interleaved Water- and Fat-Suppressed Proton Projection MRI
(WASPI) for myelin imaging in human brain.
Introduction
It
is challenging to directly image myelin due to its extremely short T2*
(<300µs at 3T). Inversion Recovery prepared Ultrashort Echo Time (IR-UTE)
imaging has been proposed for direct myelin imaging in human brain1–3. More recently, Short Repetition Time
Adiabatic Inversion Recovery (STAIR) has been proposed as a novel contrast
mechanism for myelin imaging with improved suppression of long T2 signal4. In this study, we explored feasibility
of STAIR based Zero Echo Time (STAIR-ZTE) combined with an amplitude- and phase-modulated
RF pulse and interleaved Water- and Fat-Suppressed Proton Projection MRI
(WASPI) for myelin imaging in human brain. Methods
Figure
1A illustrates adiabatic inversion recovery (IR) of myelin and long T2 tissues
(i.e., white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF)). Due
to the short T2* decay and relatively long duration of the adiabatic IR pulse
(~8ms), the longitudinal magnetization of myelin is not inverted but partially
saturated (red line in Figure 1A). The longitudinal magnetizations of long T2 tissues
are inverted and undergo inversion recovery (black dotted line in Figure 1A). By
acquiring UTE data at the nulling point of WM, we can selectively image myelin with
long T2 WM signal suppressed. With a short TR, long T2 tissues with a wide
range of T1s can be suppressed regardless of regional T1 variations4.
We
utilized a multi-spoke STAIR-ZTE sequence where several spokes are continuously
acquired after each IR preparation (Figure 1B). ZTE imaging is typically
performed using an excitation scheme where the RF pulse is applied with a fully
ramped-up readout gradient to shorten the effective TE (Figure 1C)5. In this scheme, data during RF excitation
and deadtime are missed, leaving a hole in the center of k-space (black dots in
Figure 1D). WASPI encoding is used to fill the hole using a derated readout
gradient (Figures 1C and 1D). Typically, the WASPI encoding is performed sequentially
before or after ZTE encoding, but this scheme is not optimal for STAIR-ZTE since
the WASPI data in the center of k-space can be largely contaminated by T1
variations. To address this issue, we proposed an interleaved encoding scheme
where WASPI is interleaved at a desired inversion time (TI) (Figure 1E) in a
manner similar to our previous development, interleaved hybrid encoding6. Another technical issue is unwanted
slice selectivity caused by the RF excitation scheme of ZTE which could impact
image quality in STAIR-ZTE imaging (e.g., blurring)7–9. To address this, we incorporated an
amplitude- and phase-modulated hyperbolic secant (HSn) pulse as proposed by
Schieban et al.10
The STAIR-ZTE sequence was
implemented on a clinical 3T MRI scanner (MR750, GE Healthcare). Five different
HSn pulses with 12, 24, 36, 48, and 60µs duration were designed. Figures 2A and
2B show the resulting ZTE images with hard pulses and HSn pulses, respectively.
The 12-36µs HSn pulse yielded reasonable image quality, and the 36µs HSn pulse was
subsequently selected for in vivo study (Figures 2C, D). For in vivo investigation,
three healthy volunteers (33-, 38-, and 47-year-old males) and two multiple
sclerosis (MS) patients (22- and 54-year-old females) were recruited in
compliance with the institutional review board and underwent an MRI exam using
a GE 12-ch receive-only head-coil with the following parameters: 1) STAIR-ZTE:
TR/TI/TE=142ms/65ms/12µs, spoke-to-spoke timing=1684µs, flip angle (FA)=7°,
field of view (FOV)=300x300x300mm3, matrix=160x160x60, readout
bandwidth (rBW)=62.5kHz, WASPI factor=8, #-of-spokes-per-IR=12, and scan time=9min
54sec, 2) MP-RAGE: TR/TI/TE=8.3/450/3.2ms, FA=12°, FOV=256x256x158mm3,
matrix=256x256x132, rBW=83.4kHz, scan time=5min 51sec. 3) FLAIR: TR/TI/TE=6502/1925/127.8ms,
FA=90°, FOV=256x256x317mm3, matrix=256x256x264, rBW=83.4kHz, scan time=5min
40sec. For healthy volunteers, STAIR-ZTE was repeated with four different
configurations: 36µs HSn pulse or 24µs hard pulse paired with sequential or interleaved
WASPI encoding. For MS patients, STAIR-ZTE was performed with a 36µs HSn pulse
and interleaved WASPI encoding. Results
Regular
ZTE imaging showed no dramatic improvement with an HSn pulse (Figure 3A). In
contrast, STAIR-ZTE yielded significantly improved myelin contrast with an HSn
pulse compared to with a hard pulse, where low-frequency bias (yellow arrows) impairing
myelin contrast (red arrows) was well-suppressed. Figure 4 shows STAIR-ZTE with
four configurations of RF types and encoding schemes. With a hard pulse, low
frequency signal was propagated from outer FOV, causing strong signal bias regardless
of the WASPI encoding schemes (yellow arrows in Figures 4A,B). STAIR-ZTE with an
HSn pulse and sequential WASPI encoding showed improved myelin contrast but
exhibited streaking artifacts across the image (red arrows in Figure 4C).
STAIR-ZTE with an HSn pulse and interleaved WASPI encoding showed the best myelin
contrast with suppression of both signal bias and streaking artifacts (Figure
4D). Figure 5 shows results with an MS patient (22F). The proposed STAIR-ZTE detected
a demyelinated lesion and yielded myelin-specific images, corresponding well
with clinical MRI. Discussion and Conclusion
As
myelin signal intensity is very low compared to the surrounding long T2
tissues, even very subtle imaging artifacts (e.g., low-frequency bias or
streaking/aliasing) may be critical factors, as shown in Figures 3 and 4. In
this study, we showed efficacy of an HSn pulse and interleaved WASPI encoding for
STAIR-ZTE in reducing such artifacts, improving myelin contrast. In future works, we will further optimize
the protocol to achieve improved spatial resolution and SNR and evaluate it in
a large number of subjects. Acknowledgements
The authors
acknowledge grant support from the NIH (R01NS092650), Veterans Affairs
(I01RX002604 and I01CX001388), and GE Healthcare.References
1. Sheth
V, Shao H, Chen J, et al. Magnetic resonance imaging of myelin using ultrashort
Echo time (UTE) pulse sequences: Phantom, specimen, volunteer and multiple
sclerosis patient studies. Neuroimage 2016;136:37–44 doi:
10.1016/j.neuroimage.2016.05.012.
2. Ma Y, Jang H, Chang EY, et al. Ultrashort echo time (UTE) magnetic
resonance imaging of myelin: Technical developments and challenges. Quant.
Imaging Med. Surg. 2020;10:1186–1203 doi: 10.21037/QIMS-20-541.
3. Jang H, Wei Z, Wu M, et al. Improved volumetric myelin imaging in human
brain using 3D dual echo inversion recovery‐prepared UTE with complex echo
subtraction. Magn. Reson. Med. 2019:mrm.28082 doi: 10.1002/mrm.28082.
4. Ma Y, Jang H, Wei Z, et al. Myelin Imaging in Human Brain Using a
Short Repetition Time Adiabatic Inversion Recovery Prepared Ultrashort Echo
Time (STAIR-UTE) MRI Sequence in Multiple Sclerosis. Radiology 2020;297 doi:
10.1148/radiol.2020200425.
5. Jang H, Carl M, Ma Y, et al. Inversion recovery zero echo time (IR-ZTE)
imaging for direct myelin detection in human brain: a feasibility study. Quant.
Imaging Med. Surg. 2020;10:895–906 doi: 10.21037/qims.2020.04.13.
6. Jang H, Ma Y, Searleman AC, et al. Inversion recovery UTE based
volumetric myelin imaging in human brain using interleaved hybrid encoding.
Magn. Reson. Med. 2020;83:950–961 doi: 10.1002/mrm.27986.
7. Grodzki DM, Jakob PM, Heismann B. Correcting slice selectivity in hard
pulse sequences. J. Magn. Reson. 2012;214:61–7 doi: 10.1016/j.jmr.2011.10.005.
8. Jang H, Wiens CN, McMillan AB. Ramped hybrid encoding for improved
ultrashort echo time imaging. Magn. Reson. Med. 2016;76:814–825 doi:
10.1002/mrm.25977.
9. Cheng Li, Magland JF, Seifert AC, Wehrli FW. Correction of Excitation
Profile in Zero Echo Time (ZTE) Imaging Using Quadratic Phase-Modulated RF
Pulse Excitation and Iterative Reconstruction. IEEE Trans. Med. Imaging
2014;33:961–969 doi: 10.1109/TMI.2014.2300500.
10. Schieban K, Weiger M, Hennel F, Boss A, Pruessmann KP. ZTE imaging
with enhanced flip angle using modulated excitation. Magn. Reson. Med.
2015;74:684–693 doi: 10.1002/mrm.25464.