Jeff Snyder1 and Alan H Wilman1
1Biomedical Engineering, University of Alberta, Edmonton, AB, Canada
Synopsis
A 3D FSE method using variable flip angle trains is proposed for quantification of T2 in brain. Two images are acquired at different echo times to allow for decay curve fitting using Bloch equation and Echo Phase Graph (EPG) simulations. Echo train lengths of 96 and 192 were investigated (total scan times of 8:12 and 2:38) in phantom and healthy subjects at 3 T, with isotropic resolutions of 0.9 and 1.3 mm3, respectively. RF was optimized to reduce blurring, sustain signal and allow T2 resolution. Comparison with previous methods was excellent, with good resolution and contrast in the 96 case.
Introduction
Three-dimensional
(3D) fast spin echo (FSE, [1, 2]) techniques are seeing increas use for
brain imaging over 2D methods mainly due to higher spatial and
isotropic resolutions. Quantification of relaxation parameters in 2D and 3D
via FSE techniques is often time-consuming due to the acquisition of images at
several echoes along the train to model the decay curve. Slice-selective 2D methods result in
sub-optimal resolution in the slice direction and confounding incidental magnetization transfer
effects [3]. While 3D methods mitigate these issues, long echo trains are
desired to reduce scan time, which can adversely affect image quality (point spread function blurring) and patient safety (high SAR deposition). Additionally, 3D FSE techniques are not
optimized for T2 quantification which requires significant decay curve
resolution between different T2 species, i.e. adequate T2 decay
separation across the T2 spectrum within the brain (~20 – 80 ms) must be
achieved to delineate T2 components. We propose a 3D FSE technique based
on a dual-echo approach (via two separate acquisitions) with decay curve
modeling accomplished by Bloch equation simulations [4]. This technique uses variable flip angle trains optimized for sustained signal, PSF blurring reduction and T2
quantification. The method is
illustrated using echo train lengths of 96 and 192.Methods
A modified version of the Siemens 3D FSE sequence (SPACE - Sampling
Perfection with Application optimized Contrasts using different flip angle
Evolution) was used to acquire two different images - one with proton density
(PD) weighting and one with T2 weighting – to provide two points
on the decay curve for each image pixel. Two echo train lengths
were explored (96 and 192) in phantom measurements and healthy volunteers using
a 3 T Siemens Prisma. The phantom
consisted of 6 cylindrical tubes with different concentrations of MnCl2
to produce a different T2 value for each tube.
An isotropic resolution of 0.9 mm3 and 1.3 mm3 was
used for the 96 and 192 trains, respectively.
Other parameters included 5.8 ms inter-echo spacing, 2x GRAPPA acceleration and a TR of 1400 ms. A rapid B1 map was also acquired for input
into the fitting routine.
An in-house signal response program developed in MATLAB using
Echo Phase Graph (EPG, [5]) and Bloch equation techniques to simulate flip angle
trains provided maximization of T2 resolution (comparing ratios and differences
of amplitudes between different T2 species) while keeping signal as high.
The proposed train for the 96 echo case
was based on achieving 20% of the maximal amplitude for each echo of
the 80 ms T2 isochromat by ramping the flip angles from 17° to 180°. When greater than 180° was required to keep signal near 20%,
the flip angle was dropped to 120° for the remainder of the train to reduce SAR. Further, the amplitude of the first echoes were chosen to move the
system to the pseudo-steady state (PSS, [6]) while keeping the 1st echo
amplitude as high (see Figure 1).
The echo times chosen for the proposed train were 6 ms (1st
echo) and 132 ms (23rd echo).
This train was compared to a constant echo train using flip angles of
120°. Results
Figure 2 shows phantom experiments for the constant flip angle
(right) and proposed 96 and 192 variable flip angle trains (left and middle)
for the two different TEs (6 ms top, 132 ms bottom). While the constant case shows good signal at TE = 6 ms,
substantial blurring is evident compared to the variable cases. Signal for shorter T2 components has completely decayed in the constant case at TE = 132 ms (see missing tubes in bottom right
image), while the variable cases retain signal required for T2 quantification. There is more artifact present in the 192
cases as expected, as well as lower signal overall.
Images for volunteer studies using
the variable trains are shown in Figure 3.
Acquisition time for the 96 case (left four images) was 4:06 for each 3D
image (8:12 total) and 1:17 for each (2:34 total) for the 192 case (right four
images). Reference lines show the
location of the axial slice used for T2 quantification. T2 maps are shown in Figure 4 for the 96 and 192 cases,
and quantified values from 3 regions are shown in Figure 5 compared to previous
work from our lab (2D methods using MESE and PD-T2, [3]). Computation time for each map slice was 1 min 38 s.
T2 values are reported as an average of a manually segmented
region entirely contained in the structure (~roughly 50 pixels total). The 192
case shows significantly more noise but has values similar to the
96 case and previous methods. Further, it
should be noted that the slice position is slightly offset from the 96 case as
the resolution differed. The largest
variance occurred in the short T2 region Globus Pallidus. Discussion
We have introduced variable flip angle approaches to enable T2 quantification using long train 3D FSE sequences. This technique can provide T2
quantification at high resolution with minimal impact on total scan time (2:38 for 192 echo trains). Further
work will focus on simultaneous optimization of train length and signal, with
additional emphasis on T2 precision. Acknowledgements
Operating grant funding from Canadian Institutes of Health.References
1. Hennig J, Scheffler, K. Easy improvement of signal-to-noise in RARE-sequences with low refocusing flip angles. Rapid acquisition with relaxation enhancement. Magn Reson Med. 2000;44(6):983-985.
2. Mugler JP. Optimized three-dimensional fast-spin-echo MRI. J Magn Reson Imaging. 2014;39(4):745-767.
3. McPhee KC, Wilman AH. T1 and T2 quantification from standard turbo spin echo images. Magn Reson Med. 2019; 81:2052-2063.
4. McPhee KC, Wilman AH. Transverse relaxation and flip angle mapping: Evaluation of simultaneous and independent methods using multiple spin echoes. Magn Reson Med. 2017; 77(5):2057-2065
5. Weigel M. Extended phase graphs: Dephasing, RF pulses, and echoes - Pure and simple. J Magn Reson Imaging. 2015;41(2):266-295.
6. Hennig J, Weigel M, Scheffler K. Multiecho sequences with variable refocusing flip angles: optimization of signal behavior using smooth transitions between pseudo steady states (TRAPS). Magn Reson Med. 2003; 49(3):527-535.