Haonan Zhang1, Qingwei Song1, Jiazheng Wang2, Peng Sun2, Renwang Wang1, Nan Zhang1, and Ailian Liu1
1Department of Radiology, the First Affiliated Hospital of Dalian Medical University, Dalian, China, 2PHILIPS——Philips Healthcare, beijing, China
Synopsis
Traditional
3D-TOF Magnetic Resonance Angiography (MRA) in carotid imaging needs a relatively
long scan time and is prone to be disturbed by motion artifacts such as vascular
pulsation, respiratory movement, and some other physiological factors. A
combination of compressed sensing (CS) can achieve high
acceleration factors and thus lead to a significant reduction of scan time. Here,
we investigated the impact of acceleration factors of compressed sensing on
the image quality of 3D-TOF-MRA for cervical vessels. CS acceleration
factor of 6 is recommended for clinical 3D-TOF carotid MRA to achieve an
optimal balance between imaging time and image quality.
Introduction
3D-TOF
magnetic resonance angiography (MRA)1 enables non-invasive visualization
of the anatomical structure of blood vessels. It has been widely used for the measurement
of vascular morphology and detection of carotid artery stenosis 2. However,
the traditional 3D-TOF MRA often needs a relatively long scan time, during which
the patient's swallowing, respiratory movement, neck movement, etc. might result
in motion artifacts and thus hamper the follow-up diagnosis3. For
patients with carotid artery stenosis, it is difficult to maintain a long time immobile
state due to cerebral hemodynamic insufficiency. The CS technique 4
was proven a powerful tool for MR imaging acceleration 5-6. The purpose
of this study was to explore the impact of acceleration factors of CS on the image
quality of 3D-TOF-MRA for cervical vessels and to find an optimal acceleration
factor to reduce the scan time without degeneration of image quality.Materials and methods
22
healthy volunteers (15 males, 54.41±18.18
years) were recruited and underwent the 3D-TOF MRA scan of neck vessels on a
3.0 T MR scanner (Ingenia CX, Philips Healthcare, Best, the Netherlands). Four groups
with different acceleration schemes were set up in our study, group A with a
routine clinical setup of SENSE acceleration
factor 3, and groups B, C, and D with CS factors of 4, 6, and 8. The scan time for
the 4 groups were 258s, 206s, 138s, and 106 s, respectively. Other scan parameters
were shown in Table 1. Regions
of interest were placed manually at both sides of the carotid artery and nearby
sternocleidomastoid
muscle by experienced radiologists for
measurement of signal intensity, standard deviation, and diameter of carotid
artery vessels
(Figure 1). Signal to Noise Ratio (SNR) and Contrast
to Noise Ratio (CNR) were calculated from the labeled ROIs for all volunteers. The
two observers used a four-point scoring method to evaluate the quality of the
four groups of images: 3, excellent (high quality, high intensity of arterial
vessels, clear contour, no artifacts, and clear display of primary and
secondary branch vessels); 2, good (strong signal of arterial vessels, clear contour,
and clear first-grade branch vessels); 1, fair (strong signal of the artery, moderately
clear contour, and unclear branch of the artery, with a few artifacts which did
not hamper the diagnosis); 0, poor (shallow arterial angiography, unclear
contour which can not meet the diagnostic requirements).
The SNR and CNR on the left and right sides of the carotid
artery and nearby sternocleidomastoid
muscle were compared with a Wilcoxon-rank sum test.
The Kappa statistics were calculated for determining the interobserver
agreement. The assessment of intermethod agreement was based on the evaluation
of the senior physicians. Kruskal-Wallis test was employed to assess the difference
of SNR, CNR, the largest, smallest diameters of the common carotid arteries and
image scores between the 4 groups. Mann-Whitney U test was used to make a
pairwise comparison. This study has been approved by the local IRB.Results
SNR
and CNR from two sides of the carotid artery are in good agreement (Table. 2). Scores
measured by two observers are in good agreement (p=0.880). There
were no statistically significant differences in SNR, CNR, the largest and
smallest diameters of the common carotid arteries between the four groups.
However, if CS acceleration factor of 8 was used, the subjective scores decreased
obviously (p < 0.05, Table. 3). And no significant differences of
image quality were detected between conventional SENSE acceleration with a factor of 3 and CS acceleration with factors of 4 and 6.Discussion and Conclusions
Scan
time for the cervical vascular 3D-TOF MRA could be reduced significantly by CS
technique. However, if a large CS acceleration factor (≥ 8) is used, the image quality
degrades evidently. CS acceleration factor of 6 is recommended for clinical 3D-TOF
carotid MRA to achieve an optimal balance between imaging time and image
quality.Acknowledgements
No acknowledgement found.References
1.
Haifeng
Liu, Yongsheng Xu, Yangqin Xun, et al. Diagnostic value of 3D time-of-flight magnetic
resonance angiography for detecting intracranial aneurysm: a meta-analysis.
Neuroradiology. 2017;59(11):1083-1092.
2.
Yamada
K, Yoshimura
S, Shirakawa
M,
et al. High intensity signal in the plaque on routine 3D-TOF MRA is associated
with ischemic stroke in the patients with low-grade carotid stenosis. Neurol Sci. 2018;385:164-167.
3.
Bo
Li, Hao Li, Li Dong, Guofu Huang, Fast carotid artery MR angiography with
compressed sensing based three-dimensional time-of-fight sequence. Magn Reson Imaging. 2017;43:129-135.
4.
Bratke
G, et al. Accelerated MRI of the lumbar spine using compressed sensing: quality
and efficiency. Magn Reson Imaging. 2018;49: 164-175.
5.
Seung
Hyun Leea, Young Han Lee, Jin-Suck Suh. Accelerating knee MR imaging:
Compressed sensing in isotropic three-dimensional fast spin-echo sequence. Magn
Reson Imaging. 2018;46:90-97.
6. Alessandro
Furlana, Ersin Bayram, Senthur Thangasamy, Dale Barleya, Anil Dasyam.
Application of compressed sensing to 3D magnetic resonance
cholangiopancreatography for the evaluation of pancreatic cystic lesions. Magn Reson
Imaging. 2018;52:131-136.