Niranjan Balu1, Wenjin Liu1, Zhensen Chen1, Anders Gould1, Dan S Hippe1, Li Chen1, Binbin Sui2, Mi Shen2, Peiyi Gao2, Thomas S Hatsukami1, and Chun Yuan1
1Radiology, University of Washington, Seattle, WA, United States, 2Beijing Tiantan Hospital, Beijing, China
Synopsis
Distal intracranial artery length
measurement on TOF-MRA is a biomarker of age-related vascular changes but its measurement
reproducibility in multi-center setting is unknown. We studied the dependence
of measurement reproducibility on protocol and scanner platform variation. Reproducibility
increased with increasing superior-inferior TOF-MRA coverage and consistency of
imaging parameters. After correcting for these factors, high reproducibility
was achieved between Philips 3T and Siemens 3T scanner platforms suggesting
feasibility of quantitative intracranial vessel length measurements for
multi-platform serial MRI studies.
Introduction
Artery length measurements on MRA
have been suggested as biomarkers of age-related vascular changes over time (1).
Intracranial artery feature extraction (iCafe) is one method to quantitatively
measure proximal and distal artery lengths on TOF-MRA (2).
A prior study of iCafe measurements performed on TOF-MRA on a Philips 3T
scanner proved to have high inter-reader and inter-scan measurement
reproducibility (3).
However, reproducibility of iCafe artery length measurements may be scan
protocol and scanner platform dependent. Therefore, the applicability of iCafe
in multi-center, multi-platform situations where scan protocols may differ is
unknown.
Furthermore, distal intracranial artery
length on iCafe was shown in prior studies to be the quantitative iCafe measure
most associated with age-related vascular changes. However, distal artery
length measurement is also more likely to be affected by variations in superior-inferior
(S/I) FOV and variations in positioning of the S/I FOV between scan
time-points. Visualization of distal arteries can also vary based on TOF-MRA
protocol parameters. Such variations are more likely across scanner platforms
in multi-center studies. Hence it is critical to study the reproducibility of
distal artery length measurements across scan protocols in a multi-center,
multi-platform setting.Aim
To investigate the factors determining inter-scan
reproducibility of iCafe distal artery length namely, 1) scanner platform
differences, 2) scan parameter differences, 3) S/I FOV differences. Methods
TTen subjects were scanned on a
Siemens Trio 3T scanner. TOF-MRA was obtained with the parameters shown in
Table 1. At a different institution, fifteen subjects were scanned on a Philips
Ingenia CX 3T scanner and TOF MRA was obtained with parameters shown in Table
1. All subjects were rescanned within a two-week time interval with the same
protocol as the first scan. Artery centerlines were drawn using the iCafe
procedure and quantitative artery length and branching measurements were
obtained as listed in tables 2 and 3. The coefficient of variation (CV) and intraclass correlation coefficient (ICC) were computed to determine the
inter-scan reproducibility. All study procedures were reviewed and approved by
local Institutional Review Boards, and all subjects provided written informed
consent.
Factors determining
reproducibility: The improvement in reproducibility by reducing variation
in protocol and FOV was investigated by comparing CV and ICC for subsets of the
Siemens 3T scans. Three subsets: 1) complete set (N=10), 2) subset with
consistent imaging parameters between the two scans (N=9); and 3) subset with
consistent imaging parameters and matched FOV between the two scans (N=7) were
compared (Table 2).
The Philips 3T TOF S/I FOV (7.2cm)
was reduced to match the Siemens 3T S/I FOV (4.9cm). CV and ICC were compared
between the two 4.9cm FOV TOF-MRA iCafe measures to determine multi-platform
reproducibility.
Finally, the impact of reducing
the S/I FOV on measurement reproducibility was determined by comparing the CV
and ICC for full S/I FOV (7.2cm) and reduced S/I FOV (4.9cm) on the Philips 3T
TOF-MRA.Results
iCafe centerlines on the first and
second time points (TP1 and TP2) were consistent both visually (figure 1) and
quantitatively (tables 2 and 3). The primary measurement of distal vessel
length had a consistently high ICC (0.9 or higher) across the subsets and CV of
9%. Other vessel length measurements (total length and proximal length) had
similar CV (9-11%) but lower ICC. Other measures such as number of branches and
normalized signal intensity had lower levels of reproducibility suggesting that
these measures are not suitable as primary biomarkers in serial MRI studies.
When S/I FOV was matched between
Siemens 3T and Philips 3T, distal vessel length had comparable reproducibility
on the two scanner platforms (table 3).
Compared to the reduced S/I FOV of 4.9cm,
measurements using the S/I FOV of 7.2cm proved more reproducible across all
length and branching measurements (compare CV and ICC across tables 3 and 4). Conclusions
When the same FOV and TOF MRA
protocol was used for reproducibility assessment, distal artery length
measurements were highly reproducible (ICC = 0.9). Distal artery length
measurement had a similar high reproducibility on both Philips and Siemens 3T
(ICC > 0.9) when imaging parameters and S/I coverage were consistent between
scans. This suggests that iCafe measured distal artery length measurements are
reproducible in a multi-platform setting and therefore usable to monitor
changes in multi-center studies using serial iCafe measurements.
Mis-registration between
time-points did not play a major role in determining reproducibility suggesting
that scan positioning was consistent between the time-points. While the S/I FOV
of 4.9cm centered on the circle of Willis and in the plane of the MCA was
sufficient, lower S/I FOV is expected to increase distal artery measurement
variability.
Corroborating this, we found that
measurement reproducibility was improved when S/I FOV was increased from 4.9cm
to 7.2cm suggesting that full S/I FOV coverage of the brain is ideal for serial
iCafe TOF-MRA measurements. Whole brain coverage will also negate variability
due to mis-positioning of scans between two time-points.
In summary, we identified the factors affecting
iCafe distal artery length measurement inter-scan reproducibility. While both
Siemens and Philips 3T TOF-MRA provided highly reproducible measurements, our
results suggest that whole brain S/I FOV coverage of TOF-MRA can further
improve inter-scan reproducibility in multi-center settings.Acknowledgements
This work was supported in part by NIH grants R01HL103609 and R01NS092207References
1. Chen
L, Sun J, Hippe DS, et al. Quantitative assessment of the intracranial vasculature
in an older adult population using iCafe. Neurobiol Aging 2019;79:59-65.
2. Chen
L, Mossa-Basha M, Balu N, et al. Development of a quantitative intracranial
vascular features extraction tool on 3D MRA using semiautomated open-curve
active contour vessel tracing. Magn Reson Med 2017.
3. Chen L, Mossa-Basha M, Sun J, et al.
Quantification of morphometry and intensity features of intracranial arteries
from 3D TOF MRA using the intracranial artery feature extraction (iCafe): A
reproducibility study. Magn Reson Imaging 2019;57:293-302.