Kaibao Sun1, Guangyu Dan1,2, Zheng Zhong1,2, and Xiaohong Joe Zhou1,2,3
1Center for MR Research, University of Illinois at Chicago, Chicago, IL, United States, 2Department of Bioengineering, University of Illinois at Chicago, Chicago, IL, United States, 3Departments of Radiology and Neurosurgery, University of Illinois at Chicago, Chicago, IL, United States
Synopsis
The dependence of apparent diffusion coefficient (ADC) on echo time (TE) has
been observed in a number of biological tissues. Such coupling can be exploited
to extract tissue microstructural information. We herein introduce a
time-efficient sequence that is capable of studying the interplay between ADC
and TE. The sequence incorporated
multiple echo-train readouts into diffusion-weighted echo planar imaging,
together with a 2D RF excitation pulse to reduce the FOV, thereby allowing
multiple effective TEs in an acquisition. With this sequence, ADCs were
evaluated at three TEs and T2*s at three b-values. The proposed technique
has been demonstrated in the prostate.
Introduction:
It has been observed
that the measured apparent diffusion coefficient (ADC) depends on the TE in a
number of biological tissues1–4. Such coupling can be exploited to extract
tissue microstructural information for improved characterization of lesions,
such as prostate cancer3,4. In a method known as hybrid multidimensional imaging4, images with multiple b-values and multiple
TEs are acquired to study the interplay between ADC and T2, which
typically require multiple acquisitions. The approach not only lengthens the
scan times but also increases susceptibility to motion. We herein report a time-efficient pulse sequence to
acquire diffusion-weighted signals at multiple TEs following a single
excitation. This sequence, which we call multi-readout DWI, is capable of
obtaining ADC values at different TEs, as well as generating T2*
maps at different b-values. This ability has been demonstrated in both healthy prostate
tissues and prostatic carcinoma. Methods:
Sequence design: The multi-readout
DWI sequence was built upon a spin-echo
diffusion-weighted EPI sequence in which
multiple (e.g., 3) EPI readout echo-trains were placed after a Stejskal-Tanner
diffusion preparation module5 (Figure 1A). Each EPI readout echo-train corresponded
to a distinct effective TE, where
the effective TE was defined as the TE when k-space center is sampled. The first echo-train coincided with the
nominal TE (TE0). Large gradient blips (Figure 1A) along the
phase-encoding direction were applied between adjacent echo-trains to re-position
the start point of the k-space trajectory for the next echo-train acquisition. To
reduce the duration of each echo-train without compromising the spatial
resolution so that multiple echo-trains could be acquired, a 2D RF pulse
(Figure 1B) was used to limit
the FOV6. The 2D RF pulse was designed by employing a fly-back
EPI-like excitation k-space trajectory. Eleven sub-pulses with a time-bandwidth
product (TBP) of 3.01 were modulated by an envelope pulse whose TBP was 3.53 and
pulse width was 14.7 ms. A tilted excitation k-space strategy was employed to
enable multi-slice imaging6.
Image analysis: Diffusion-weighted images acquired
at different TEs were separately reconstructed. The image from the first
echo-train (TE0) corresponded to a conventional diffusion-weighted image.
Images from the subsequent echo-trains were subject to T2* decay as described
by7:
$$S(TE_n) = S_0exp(-(TE_n-TE_0)/T_2^{*})$$
where S0 is the
diffusion-weighted signal at TE0.
T2* maps at each b-value were obtained from images acquired with
three TEs. Similarly, ADC maps were calculated based on three b-values at each
TE. From the T2* and ADC maps, the mean values of T2* and
ADC were calculated over an ROI and used for comparisons among different b-values
and TEs.
Experiments: The multi-readout DWI sequence was implemented
on a GE MR750 3T scanner. To demonstrate the sequence, axial prostate MRI scans
were performed using a 32‐channel cardiac coil on healthy volunteers and a
patient with confirmed prostate cancer with the following parameters: FOV = 180x90 mm2, reconstruction
matrix = 128x64, number of slices = 18, slice thickness = 3 mm, slice spacing =
1 mm, TR = 4000 ms, number of echo-trains = 3, TEs = 63/78.8/94.6 ms, and b = 0/500/1000
s/mm2, NEX = 6, and the scan time = 3 min and 40 s.Results:
Figures 2 and 3 display a set of
multi-dimensional diffusion-weighted prostate images, with different b-values
and TEs, for a representative healthy subject and the patient with prostate
cancer, respectively. The ADC maps at each TE values for the two subjects are
shown in Figure 4, whereas the T2* maps at each b-value are
illustrated in Figure 5. In both healthy and cancerous prostate tissues, ADC
increased with TEs (Figure 4). However, the ADC values in the prostate cancer
exhibited a higher degree of heterogeneity and a lower value than those in the
healthy prostate gland. Figure 5 indicates that T2* values decreased
as b-valued increased for both subjects. Additionally, the prostate cancer
exhibited a lower T2* value. Discussion and conclusion:
In this study, we have demonstrated a
novel sequence – multi-readout DWI – that was capable of producing
diffusion-weighted images at different TEs following an excitation without extending the scan time. This approach also
resulted in a set of spatially co-reregistered images across multiple TEs,
reducing or eliminating the problems caused by motion. In normal prostate tissue, ADC and T2*
values changed with TE and b-values, respectively, consistent with the
observations reported in previous studies3,4. These coupling was likely a reflection of the
underlying tissue microstructures related to the amounts of stroma, epithelium, and lumen in the prostate3. For example, fluid in the glandular lumina has a
long T2* and a high ADC, while stroma and epithelium have shorter T2*
and lower ADC values. DWI with high b-values suppressed signals from rapidly
diffusing spins in the glandular lumen, and thus highlighted shorter T2*
signals from stroma and epithelium. Alternatively, long TEs highlighted long T2*
signals in lumen, which are characterized by a high ADC. Although this study
employed prostate imaging as an example, the multi-readout DWI sequence is
expected to find other applications in tissues where the coupling of ADC and
TEs may provide useful information to aid in diagnosis. Acknowledgements
This work was supported in part by
the National Institutes of Health (5R01EB026716-01 and 1S10RR028898-01). The content is solely the responsibility of
the authors and does not necessarily represent the official views of the National
Institutes of Health. The authors are grateful to Drs. Muge Karaman, Qingfei
Luo, Kezhou Wang, and Cui Feng for helpful discussions.References
1. Stanisz GJ, Henkelman RM. Diffusional
anisotropy of T2 components in bovine optic nerve. Magn Reson Med.
1998.
2. Does MD, Gore JC. Compartmental study of
diffusion and relaxation measured in vivo in normal and ischemic rat brain and
trigeminal nerve. Magn Reson Med. 2000.
3. Wang S, Peng Y, Medved M, et al. Hybrid
multidimensional T2 and diffusion-weighted MRI for prostate cancer detection. J
Magn Reson Imaging. 2014.
4. Chatterjee A, Bourne RM, Wang S, et al.
Diagnosis of prostate cancer with noninvasive estimation of prostate tissue
composition by using hybrid multidimensional MR imaging: A feasibility study. Radiology.
2018.
5. Stejskal EO, Tanner JE. Spin diffusion
measurements: Spin echoes in the presence of a time-dependent field gradient. J
Chem Phys. 1965.
6. Finsterbusch J. Fast-spin-echo imaging of
inner fields-of-view with 2D-selective RF excitations. J Magn Reson Imaging.
2010.
7. Eichner C, Paquette M, Mildner T, et al.
Increased sensitivity and signal-to-noise ratio in diffusion-weighted MRI using
multi-echo acquisitions. Neuroimage. 2020.