Javad Parsa1, Thomas O'Reilly1, Bart de Vos1, and Andrew Webb1
1Leiden University Medical Center, Leiden, Netherlands
Synopsis
An integrated transmit coil and four-element receive
array has been simulated, constructed, characterized and tested on a low field
MRI system operating at 2.15 MHz. Using a combination of loops and butterfly
coils neighbouring inter-element coupling is below -18 dB, with directly
opposite coils ~-9 dB, and <-17 dB coupling for all receive coils to the
transmit coil. Images of a phantom have been acquired with a simple
sum-of-squares reconstruction.
Introduction
Phased
array receive coils1 are ubiquitous on clinical MRI systems,
providing improved signal-to-noise (SNR) over larger volume coils and the capability of undersampling k-space for more rapid image acquisition2.
Phased array coils would also be very useful for very low field systems for
reducing imaging time and enabling lossless receive bandwidth amplification via
preamplifier impedance mismatch1, even though coil-dominated noise
means that SNR is not increased as it is at higher fields. However, implementation of
phased arrays receivers on very low field systems is more challenging since
coil/sample coupling is much lower, loaded Q-values are higher, and so
inter-coil coupling is higher. In addition, different geometries are required
due to the transverse nature of the B0 field in Halbach arrays, the
most common geometry for low field MRI. One example has been presented
previously3, but this used only surface loops, many of which have
very low sensitivity since B0 and B1 directions are
coincident. In this work, we design a four coil receive array constructed by a
combination of loop and butterfly coils and demonstrate performance on a 50 mT
(2.15 MHz) Halbach-based MRI system. Methods
Two
loop coils and two butterfly coils were constructed from flexible PCB with 0.3
mm thickness and 3 mm trace widths, and were fixed on a
plexiglass cylinder with 15 cm diameter and 25 cm length (Figure1): loop and
butterfly coils have 4 and 2 turns, respectively, with a total conductor length
<< λ/20. The spacing between the coils was optimized to minimize coupling
between the adjacent butterfly and loop coils (Figure 1). A solenoid transmit coil
was designed with MATLAB and CST (CST Microwave Studio, Darmstadt, Germany) to
create a 20 cm axially-uniform B1+ field (Figure 2). The Tx coil was constructed with 48 turns of
1 mm copper wire on a plexiglass former with 22 cm diameter and 30 cm length. Impedance
matching to 50 W was
performed for each coil loaded with a saline phantom. Images were obtained
using a 50 mT Halbach-magnet based MRI system4 using a Magritek Kea2
spectrometer (Aachen, Germany). A turbo spin-echo sequence was used with TR/TE:
3000 ms/ 15 ms, Echo train length: 40, FOV: 180x180x201 mm, 3x3x3 mm resolution,
and acquisition bandwidth: 20 kHz.Results
The
simulated B1+ field of the solenoid coil is shown in
figure 2: the field is homogeneous to within 4% over an axial region of 20 cm.
Reflection coefficients (S11) for each of the four individual Rx
coils and the transmit coil under loaded conditions were lower than -22 dB at
2.15 MHz. Figure 3 shows the inter-element coupling, as well as the coupling of
each element with the transmit coil. Turbo spin-echo (TSE) images were obtained
from a melon using the array and compared to the image from the solenoid coil
in transmit and receive mode (Figure 4 and Figure 5). Individual images are
shown, as well as a simple sum-of-squares reconstruction (Figure 4). There is
almost no coupling between adjacent elements because of the low mutual inductance, but
signal coupling between opposite elements is apparent due to the very low
loading of the sample.Conclusion
A prototype
construction four-element array has been constructed and tested on a low field
Halbach-based MRI system. The relatively large coils used in the array mean
that opposite elements have significant coupling, although this can potentially
be reduced to ~-20 dB by additional preamplifier decoupling, as well as by
using a larger number of smaller RF coils. Images have been acquired and show
slightly higher SNR than from a transmit/receive solenoidal coil, primarily due
to the slightly smaller diameter.Acknowledgements
This work was funded by H2020-MSCA-ITN-ETN-2019 and
Horizon 2020 ERC Advanced NOMA-MRI 670629.References
- Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array,
Magn Reson Med. 1990; 2:192-225.
- Hamilton J, Franson D, Seiberlich N. Recent advances in
parallel imaging for MRI. Progress NMR Spectroscopy. 2017; 101:71-95.
- Cooley C Z,
Stockmann J, Armstrong B D, et al. Two-Dimensional Imaging in a
Lightweight Portable MRI Scanner without Gradient Coils. Magn
Reson Med. 2015; 73:872–883.
- O’Reilly
T, Teeuwisse W M, and Webb A G, In vivo 3D brain and extremity MRI at 50 mT
using a permanent magnet Halbach array. Magn Reson Med. 2020;00:1-11.