Mao Sheng1, ZhongChang Miao2, Jian Bao3, Renjie Zong3, SiSeung Kim3, Huiyao Zhang3, and Bing Keong Li3
1Department of Radiology, Children’s Hospital of Soochow University, Suzhou, China, 2Department of Radiology, The First People’s Hospital of Lianyungang, Jiangsu Province, China, 3Jiangsu LiCi Medical Device Co., Ltd, Lianyungang, China
Synopsis
A dedicated 0.35T Neonatal-Infant Brain MRI system is
developed to investigate the effectiveness and safeness for neonatal and infant
patients. 66 volunteers from 0-12 months are recruited to undergo a clinical
trial and it is found that low field brain images displayed high grey-white
tissue contrast, which is notably helpful for clinical diagnosis. The newly
developed system also has very low acoustic noise and that most of the volunteer
showed no sign of temperature raise. Low field MRI system therefore has the
potential to be a effective and safer alternative MRI system for neonatal and infant
patients.
Introduction
In the past many years, R&D in MRI technology has
been concentrated in increasing the magnetic field strength. Although higher
SNR and thinner slices with higher anatomical spatial details can be obtained
with high field system, imaging sequences are mainly optimized for adult
subject. However, at high field, for neonatal and infant with higher water
content of the brain, notably the white matter, it makes the differential of
gray and white matter difficult, which is further complicated at 3.0T and above
because of the longer T1 relaxation times that already decrease T1 tissue
contrast1. In view of this, we hypostasized that low field MRI
system can be a effective and safer alternative system for neonatal and infant
brain imaging and an investigation is undertaken. In this work, we designed and
built a dedicated 0.35T Neonatal-Infant Brain MRI (NIB-MRI) system and
performed a clinical trial with 66 volunteers from 0-12 months old. The results
reported herein, showed that the acquired low field MR brain images can provide
high clinical information and can be effectively used for medical diagnosis
purposes. In addition, from the safety aspects it is shown to be a much safer system
for neonatal and infant patients.Methods
Shown in Fig 1 is the developed NIB-MRI system. A
0.35T open permanent magnet is used and positioned in a “U” shaped
configuration. For minimizing SAR, a dedicated two channels transceive RF head
coil is developed and optimized T1W-SE, T2W-FSE, T2W-FLAIR, DWI and 3D-T1-GRE
pulse sequences for neonatal and infant brain imaging are also developed and
tested. The NIB-MR system is firstly tested according to IEC 60601-1, 60601-1-2
and 60601-2-33 and passed all safety tests. 66 volunteers consisting of 33 male
and 33 female subjects ranging from 0-12 months are recruited. A minimum 4 and
maximum 7 subjects are allocated to each different month group (according to 12
different month groups). Ethical approvals for the clinical trial are given by
the Children’s Hospital of SooChow University and The First People’s Hospital
of Lianyungang ethics committee, and parental consent is obtained in all cases.
During the clinical trial, 62 volunteers are administrated with chloral hydrate,
while 4 volunteers are not given any form of sedation. All acquired images are independently
scored based on MRI‐based
five-point Likert scoring system by four experienced MR radiologists with above
10 years of experience in diagnosing low and high field MR images.Results
Shown in Fig 1(a) is the developed NIB-MRI system with
the patient bed and the RF coil out. Fig 1(b) shows how the volunteer is
positioned. The top part of the RF coil is first dislodged and the volunteer
head is placed in the center of the RF coil, thereafter the top part of the RF
coil is locked back into position and the bed is then pushed into the magnet.
Shown in Figs 2(a-f) are the acquired brain images of the volunteers using our
developed pulse sequence. Except for 3D-T1-GRE (Figs 2(d,e)), which used slice
thickness of 2mm, the rest of the sequence used 5mm slice thickness.Discussion
Based on MRI‐based
five-point Likert scoring system, it is found that 41 volunteers achieved a
score of 5 (Maximum), 21 volunteers have a score of 4 and 4 volunteers are
given a score of 3. Volunteers with score of 3 is due to movement of the
subjects during the scan, which resulted in motion related artifacts. T1W and
T2W images (Figs 2(a-e)) displayed high grey-white tissue contrast, which
correlate well with the shorter T1 relaxation times that can be achieved with
low field MRI system and is notably helpful for clinical diagnosis. The DWI is
designed based on SE method without fat suppression and the B value is set as
600. Hence, high fat signal around the skull can still be observed as depicted
in Fig 2(f). Acoustic noise of the NIB-MRI system is measured and the highest
recorded noise level is ≤70dB(A). With ear protection, volunteer is subjected
to ≤55dB(A) acoustic noise, which is close to the American Academy of
Pediatrics recommended noise level of 45-50 dB(A)2. It should be
noted that the 4 volunteers who are not sedated slept through their entire scan
procedure and successfully completed their trial in a single attempt. For each volunteer,
a pre and post scan body temperatures are measured and it is found that 47 volunteers
showed no sign of temperature raise, 15 volunteers have <0.3°C raise and 4 volunteers
have 0.3°C to 0.5°C raise, which are all well within ICNIRP recommendation of not
exceeding 0.5°C for infant3.Conclusion
In this work, we have shown that brain images acquired
with the developed 0.35T NIB-MRI system have high clinical value and can be effectively
used for medical diagnosis. Initial findings on the safety aspects of the
system has also been established and that the low-field NIB-MRI system has the
potential to be a much safer system for neonatal-infant patients. Future work
will concentrated on developing more efficient low field specific neonatal-infant
brain imaging pulse sequence and further reducing the acoustic noise, dB/dt and
SAR levels so as to provide a safer environment for neonatal and infant
patients.Acknowledgements
No acknowledgement found.References
1. Tocchio S, Kline-Fath B, Kanal E, et al. MRI evaluation and
safety in the developing brain. Seminars in Perinatology, 2015, 39(2):73-104.
2. American Academy of Pediatrics, Committee on Environmental
Health. Noise: A Hazard for the Fetus and Newborn. Pediatrics, 1997,
100(4):724-727.
3. International Non-Ionizing Radiation Protection (ICNIRP).
Statement of Medical Magnetic Resonance Procedures: Protection of Patients.
Health Phys. 2004; 87(2):197–216.