Emilia Palmér1, Fredrik Nordström1,2, Anna Karlsson1,2, Karin Petruson3, Maria Ljungberg1,2, and Maja Sohlin1,2
1Department of Radiation Physics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden, 2Department of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Gothenburg, Sweden, 3Department of Oncology and Radiotherapy, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden
Synopsis
In an MRI-only
radiotherapy workflow, in addition to enabling absorbed dose calculation the
generated synthetic CT (sCT) must also be valid for patient positioning at
treatment course. To evaluate MRI-only patient positioning for head and neck
cancer patients, fourteen 3D cone beam CTs were retrospectively registered to
CT and sCT. Further, original Digital Reconstructed radiographs originating
from the CT and synthetic Digital Reconstructed radiographs originating from
the sCT, were retrospectively registered to orthogonal projections. The small mean
difference between the registrations showed that sCT could replace the CT
for both 3D and 2D head and neck radiotherapy patient positioning.
Introduction
In a traditional
external radiotherapy workflow, CT data is used for absorbed dose calculation and
delineation of target and organs at risk (OAR). In addition, CT data are used
for patient positioning by registration of daily orthogonal 2D projections to Digital
Reconstructed Radiographs (DRR) originating from CT, or registration of a 3D cone
beam CT (CBCT) directly to the CT data. MRI, with its superior soft tissue
contrast, can be incorporated into the radiotherapy workflow via a
co-registration to the CT data, and contributes with a more consistent
estimation of tumor volumes1,2. Lately,
interest have been directed towards MRI-only radiotherapy workflows, i.e. a
radiotherapy workflow that uses MRI as the only imaging modality. Since the MRI
does not provide the electron densities necessary for absorbed dose calculation
and cannot be directly reconstructed to DRRs with the same contrast as the CT
data, additional steps are required. These generally includes a conversion of
the MRI data to synthetic CT data (sCT) using e.g. machine learning. In an MRI-only
workflow, sCT and synthetic DRR (sDRR) originating from the sCT data replaces
the original CT and DRR data. The accuracy of sCT based absorbed dose
calculations has previously been evaluated for various sCT generation methods
and treatment sites3,4, however the impact on the
patient positioning has not yet been thoroughly investigated5 and to our knowledge
never in the head and neck (H&N) region. The aim of this study was to
evaluate the use of sCT and sDRR for patient positioning in an MRI-only H&N
radiotherapy workflow. Methods
In this study,
14 data sets with MRI, sCT, CT, CBCT (figure 1) and orthogonal projections (figure
2) in the H&N region were used. A T1 weighted Dixon Vibe (3D spoiled GRE)
acquisition was used for generation of sCT data (MRI Planner software, Spectronic
Medical AB). The original CT was deformably pre-registered to the MRI (Elastix,
MICE Toolkit) to mitigate differences due to inconsistent positioning of the
patient during the two scanning sessions and/or anatomical movement in the
body. To validate the physical behavior of the deformation, the
determinant of the Jacobian of the deformation field was calculated within the
body contour of the deformed CT (dCT)
data.
For positioning using
3D data, the CBCT was registered to the dCT and sCT in six degrees of freedom
with a rigid auto-registration algorithm and a threshold of 200-1700 HU
(Eclipse Image Registration, Varian Medical Systems).
For positioning
using 2D data, the deformed DRRs (dDRR) and sDRRs were retrospectively and manually
registered to orthogonal projections in five degrees of freedom by six blinded
observers (3 physicists, 3 radiotherapy
technologists), resulting in a total of 168 registrations (Eclipse Offline
Review, Varian Medical Systems). None of the observers were informed of the
type of DRRs that were registered to the orthogonal projections.
The difference between
patient positioning in all directions (frontal, longitudinal and sagittal axis,
and rotation around these axis) were evaluated for dCT and sCT, as well as for dDRR
and sDRR.Results
All determinants
of the Jacobin of the deformation fields were >0 within the body contours
for thirteen of the dCT data sets.
The differences
in translation and rotation for 3D and 2D patient positioning in all directions
are presented in figure 3. The mean difference (± 1sd) for 3D and 2D
patient positioning is shown in table 1. Discussion
The small mean differences
found indicated that both sCT and sDRRs contributed with an overall comparable patient
positioning as for CT and DRRs. Similar results have previously been presented
for prostate and brain5.
Larger
inter-observer variations were seen for some cases. These deviations could have
arisen from different strategies of using translation and rotation when registering
the images. The different registration strategies could provide relatively
large differences in mm or degree between observers, but the outcome of patient
positioning would be similar. These cases however need to be more deeply
analyzed.
The
pre-registration used to mitigate differences between CT and MRI data is a critical
step in this study, since an insufficient pre-registration would affect the
results in the patient positioning evaluation. The
quality of the pre-registration was estimated by calculating the determinant of
the Jacobian of the deformation field, where a
determinant >1 would indicate expansion and a determinant between 0 and 1 would
indicate a contraction of the voxel. One of the 14 dCT
data set had voxels <0 within
the body contour which would indicate a
physically unrealistic deformation. The cause was identified to be the skin on
the back which had been folded double at the MRI acquisition causing a non-invertible
deformation. To further investigate this, an evaluation of
the anatomical correspondence between the dCT and MRI
using anatomical landmarks will be added to the evaluation of
the Jacobian of the deformation field. Conclusion
The study showed that sCT could replace the CT for both 3D and 2D patient positioning and are therefore
promising to be utilized in MRI-only head and neck radiotherapy workflow. Acknowledgements
No acknowledgement found.References
1. Coen R, Ronald K, Frank AP, et al.
The potential impact of CT-MRI matching on tumor volume delineation in advanced
head and neck cancer. International
Journal of Radiation Oncology Biology Physics. 1997;39:841-848.
2. Anuradha T, Nicola C, Heiko S, et
al. Target volume delineation in oropharyngeal cancer: Impact of PET, MRI, and
physical examination. International
Journal of Radiation Oncology Biology Physics. 2012;83:220-227.
3. Anna MD, Jelmer MW, Matteo M, et al.
MR-Only Brain Radiation Therapy: Dosimetric Evaluation of Synthetic CTs
Generated by a Dilated Convolutional Neural Network. International Journal of Radiation Oncology Biology Physics. 2018;102:801-812.
4. Neelam T, Sandra F, Jing Z, et al.
Dosimetric and workflow evaluation of first commercial synthetic CT software
for clinical use in pelvis. Physics in
Medicine and Biology. 2017;62:2961-2975.
5. Amir
MO, Peter BG, Carri KG-H. MRI-only treatment planning: Benefits and challenges.
Physics in Medicine and Biology. 2018;63.