Nathanael Kim1, Kathryn Tringale2, Christopher Crane2, Neelam Tyagi1, and Ricardo Otazo1,3
1Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 2Radiation Oncology, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 3Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, United States
Synopsis
MRSIGMA
is a promising real-time volumetric imaging for MRI-guided adaptive
radiotherapy using a MR-Linac system. However, the lack of a real-time 3D
reference image acquired with similar temporal resolution introduces
significant challenges for in vivo validation. This work proposes a
retrospective self-validation for MRSIGMA, where the same data used for
real-time imaging are used to create a non-real-time reference. MRSIGMA with
self-validation is tested in patients with liver tumors using quantitative
metrics defined on the tumor and nearby organs-at-risk structures.
Introduction
The MR-linac, which combines an MRI scanner
and a linear accelerator, provides a platform for MRI-guided adaptive
radiotherapy of moving organs1,2. For example, the shape of the
radiation beam can be adapted in real-time to the tumor motion using a
multi-leaf collimator(MLC)3,4. However, current real-time MRI
technology in the MR-linac is limited to 2D imaging. MR SIGnature MAtching (MRSIGMA)
was recently introduced for real-time 3D imaging in the context of the MR-linac,
consisting of a non-real-time learning step to compute pairs of 3D motion
states and signatures and a real-time signature matching step to rapidly
acquire signature data only and perform signature matching. MRSIGMA was able to
achieve approximately 250ms of total imaging latency (including acquisition and
reconstruction)5. However, MRSIGMA was only validated qualitatively
without a proper reference. This work presents a novel version of MRSIGMA with
retrospective self-validation, where the same data used for real-time signature
matching are used to reconstruct a non-real-time 4D reference for comparison.
MRSIGMA with retrospective self-validation was tested on patients with liver
tumors referred for radiotherapy.Methods
MRSIGMA:
The offline learning step (non-real-time) uses the XD-GRASP approach6
to create the database of 3D motion states and corresponding motion signatures
(Figure 1.a). First, a motion signal is estimated from the center of k-space.
Second, the motion signal is binned into motion signatures, where the signature
for each motion state is the motion amplitude range of each respiratory bin.
Third, data is sorted into undersampled 3D motion states. Fourth, temporal
compressed sensing reconstruction is performed to obtain unaliased 3D motion
states. A total of 10 motion states were computed in this work. The online matching step (real-time) uses the same
acquisition trajectory, but each radial line is immediately processed to
compute a motion signature and to find a match in the pre-computed database of 10
3D motion states (Figure 1.b).
Retrospective
self-validation: Since data acquisition during offline training and online
matching are similar, this works proposes to use a XD-GRASP reconstruction using
all the data acquired during online matching to retrospectively form a 4D reference
(10 motion states in this work) with similar temporal resolution to MRSIGMA.
The 4D reference image was computed by retrospectively assigning one of the
XD-GRASP motion states to each spoke during online matching (Figure 2).
Data
Acquisition: Data was acquired on a 3T MRI scanner (Philips Ingenia,Best,The
Netherlands) using the 3DVANE pulse sequence in research mode to enable
golden-angle stack-of-stars acquisition. Three patients with liver tumors participated in the study.
Image
Reconstruction: Real-time MRSIGMA and non-real-time XD-GRASP reference images
were reconstructed in MATLAB (MathWorks,Natick,MA) using in-house algorithms
and raw data imported from the scanner. Images were converted to dicoms for
evaluation.
Image
contouring: Tumors and nearby organs-of-interest (liver, kidneys, and pancreas)
in each motion state of the MRSIGMA database and reference database were manually
contoured by a radiation oncologist. The contours were performed in MIM (MIM
Software Inc.) on individual axial slices, which were later interpolated to
obtain 3D contours.
Quantitative
evaluation: The dice coefficient7 was employed to quantitatively
compare the contours corresponding to MRSIGMA and the reference for each time
point. The dice coefficient is given by $$$d(A,B)=\frac{2\times Area(Intersection(A,B))}{Area(A)+Area(B)}$$$, where A and B are the contours being compared and is 1 if they completely overlap.Results
Figure 3 shows real-time 3D motion imaging of the liver tumor
and organs of interest for the first patient and Figure 4 shows the dice
coefficient over time for the same patient. The mean value ± standard deviation of the dice coefficient
for the tumor, liver, kidneys, and pancreas were 0.89±0.02, 0.96±0.003, 0.93±0.03, and 0.73±0.11 respectively. The mean value of the dice coefficient was
close to 1 and the standard deviation was lower than 0.03 for the liver tumor,
liver and kidneys, which demonstrate high performance real-time 3D motion
imaging in these anatomical regions. The lower performance in the pancreas is
mainly due to the reduced T1-weighting contrast which limited ability to
contour. Table 1 presents the mean±standard deviation of the dice coefficient
for all three patients, where patients 2 and 3 show similar trends in organs-at-risk contours. The dice coefficient corresponding to the GTV for
patient 3 is the lowest because the tumor was not clearly visible even on the conventional
NUFFT reconstruction (no motion sorting). Future work will consider the
utilization of a gadolinium-based contrast agent to enhance GTV visualization
in T1w images. Discussion
The
retrospective self-validation approach for MRSIGMA provides a mechanism to quantitatively
evaluate real-time volumetric motion imaging performance in each study. The
mean value and standard deviation of the dice coefficient over time can be used to
assess motion imaging accuracy and uncertainty during the real-time imaging
period. The utilization of the same data increases robustness of the validation
and removes potential inconsistencies in using a separate reference. In
addition, real-time adaptation to 3D shapes enables real-time dose
reconstruction of mobile organs on the MR-linac.Conclusion
This
study demonstrated that real-time 3D imaging results of MRSIGMA can be
validated using a reference estimated from the same data used for signature
matching. The quantitative self-validation can be used as a metric for
real-time motion imaging performance, a current need in clinical real-time
imaging.Acknowledgements
No acknowledgement found.References
1.
Mutic
S, D.J., The ViewRay system: magnetic
resonance‐guided
and controlled radiotherapy.
Semin Radiat Oncol, 2014. 24: p.
196-199.
2.
Raaymakers
BW, L.J., Overweg J, et al, Integrating
a 1.5 T MRI scanner with a 6 MV
accelerator: proof of concept. Phys Med Biol, 2009. 54: p. N229-N237
3.
Keall
PJ, K.V., Vedam SS, Mohan R, Motion
adaptive x-ray therapy: a feasibility study. Phys Med Biol, 2001. 46(1).
4.
Poulsen
PR, C.B., Sawant A, Keall PJ, Implementation
of a new method for dynamic multileaf collimator tracking of prostate motion in
arc radiotherapy using a single kV imager. Int J Radiation Oncol Biol Phys,
2010. 76(3): p. 914-923.
5.
Feng
L, T.N., Otazo R, MRSIGMA: Magnetic
Resonance SIGnature MAtching for real-time volumetric imaging. Magnetic
Resonance in Medicine, 2020. 84(3):
p. 1280-1292.
6.
Feng
L, A.L., Chandarana H, Block KT, Sodickson DK, Otazo R, XD-GRASP: Golden-angle radial MRI with reconstruction of extra
motion-state dimensions using compressed sensing. Magn Reson Med, 2016. 75: p. 775-788.
7.
Taha
AA, H.A., Metrics for evaluating 3D
medical image segmentation: analysis, selection, and tool. BMC Med Imaging,
2015. 15.