Ehud Jeruham Schmidt1, Marc Morcos2, Anthony Gunderman3, Junichi Tokuda4, Ravi Teja Seethamraju5, Luca Neri1, Carmen Kut2, Henry R. Halperin1, Akila Ninette Viswanathan2, and Yue Chen3
1Medicine (Cardiology), Johns Hopkins School of Medicine, Baltimore, MD, United States, 2Radiation Oncology, Johns Hopkins School of Medicine, Baltimore, MD, United States, 3Mechanical Engineering, University of Arkansas, Fayetteville, AR, United States, 4Radiology, Brigham and Women's Hospital, Boston, MA, United States, 5Siemens Medical Solutions, Boston, MA, United States
Synopsis
A metallic actively-tracked injection needle
was constructed for purposes of reducing dose to normal tissues surrounding
irradiated tumors in cervical cancer and prostate cancer radiation therapy. Hydrogel is
injected into tissue or anatomic cavities between the tumor and normal tissues,
increasing the distance from the radiation source. The needle was tested in a gynecological
phantom and in swine. It provided 1.2x1.2x1.2mm3 targeting precision
@16 frames-per-second navigation, supporting rapid navigation speeds currently
possible only under X-ray or Ultrasound guidance. The injected topology over
time was visualized during injection, allowing creation of more uniform dose-shielding
regions.
Abstract
Introduction: Injection
needles are used in multiple therapies (local sedation, edema reduction,
chemotherapy delivery, biomarker insertion). In cancer treatment with radiation
delivery or thermal ablation, injection of liquids or gels serves to distance
sensitive tissues away from regions of energy delivery, reducing therapy-related
complications. It is advantageous to inject under MR-guidance since the tumor
and organs at risk are better visualized. Passively-tracked MR-conditional
metallic needles exist, although these are only used in a restricted number of
procedures, primarily due to longer navigation times relative to X-ray or
Ultrasound. After using MR-tracked (MRT) metallic stylets for
radiation-oncology brachytherapy (1) source deployment and elongated metallic
catheters for cardiac electrophysiology ablations (2), where navigation speeds approached
those under ultrasound and X-ray guidance (3), we explored use of actively-tracked metallic
injection needles to displace the rectum from irradiated tissues for both
cervical cancer and prostate cancer radiation therapy, based on injection of a biocompatible
hydrogel (4,5). Once the needle is guided
to the correct location, MR imaging can potentially monitor the topology
of the hydrogel pocket during filling which allows further changes in needle location
(5), thus providing a uniform (Anterior-Posterior) separation of 10-15 mm which
can significantly (~80%) reduce the rectal dose (6). Efficient MRI-guided injection requires (a)
MRI-conditional injection needles with large (Young’s, Torsional) elastic
moduli for effective steering, and (b) sufficient SNR MRT to allow needle-tip tracking
at ~1x1x1 mm3 resolution and at speeds of 12-14 frames-per-second (Ultrasound).
Methods: The 1.5T actively-tracked injection
needle construct: The injection needle (Figure 1) was formed from two
concentric titanium tubes (0.81 mm OD, 0.1mm wall and 1.6mm OD, 0.2 mm wall,
respectively) bonded to each other. The inner tube served as the liquid/gel
lumen, while the space between the inner and outer tubes served as the location of the two flexible printed circuit MRT coils with embedded 63.8 MHz tuning &
matching capacitors (1,2) and 46-gage micro-coaxial cables running up the shaft.
To enable the MRT radio-frequency (RF) lobes to see the surrounding media, two
14-mm long groves were cut in the external titanium tube. At the distal end of
the device, a 30° bevel cutting tip was created. At the proximal end of the 24
cm long needle (required for prostate cancer trans-perineal approach (4)),
the micro-coax were channeled via a tuning, matching and active decoupling
circuit in the handle to a quick-disconnect Redel (Lemo, Switzerland) connector.
The fluid lumen was connected to a standard Luer adaptor, enabling connection of
syringes. Further up the MRI front-end, signals propagated on half-wavelength coaxial
cables, overlaid with 30-cm periodic resonant RF traps (Baluns), to an
8-channel preamplifier. Experiments were
conducted using a dedicated MR-tracking sequence that reconstructed the needle-tip
location and orientation in <0.1ms on the scanner’s reconstruction processor.
For navigational display of the instantaneous needle-tip location and
orientation, the tip locations were sent to a workstation equipped with a 3D
Slicer MR-Tracking module (Slicer version 4.11.20200930), overlaid on MR navigational
roadmaps, and displayed (1,3) on an in-room monitor.
Experiments:
A. Phantom
experiments were performed in a 70-cm bore MRI using a custom female-pelvis
phantom (Figure 2). A T1-weighted Turbo-Spin-Echo (TSE) image dataset (TR\TE\Ɵ=500ms\3ms\900, ETL=6, 0.6x0.6mmx3.0mm3, 64 slice, 2min acquisition) was acquired covering the phantom using an array combining the scanner body and spine arrays along with the MRT coils,
and used as a navigational roadmap. MR-Tracking (1.1x1.1x1.1mm3
resolution, Hadamard encoding, 5 phase-dithering-directions/projection, 7
averages, 16 fps) was used during navigation. To show dynamic changes in the
injected volume during water injection at the desired phantom location, 2D GRE images
(TR\TE\Ɵ=10ms\3ms\500, 2.0x2.0x4.5mm3, 5 slice/sec) were
acquired for 20sec.
B. Swine experiment:
A vaginal obturator, which spatially-anchors the brachytherapy radiation sources,
was placed in the vaginal-canal. The injection-needle was used with a T2-weighted
(TR\TE\Ɵ=2000ms\101ms\1800, 0.8x0.8x3mm3, ETL=12, 32 slice,
1:30min acquisition) roadmap, to actively navigate and inject Gd-DTPA-doped water into a
pocket between the vaginal-canal and rectum.
Results: Navigational Precision and speed: Comparing MRI images of the beveled tip at
the target location to its MR-Tracked location (Figure 2A1), the tracking precision was 1.2x1.2x1.2mm3.
GRE and TSE images of the deflectable needle tip (Figure 1B, 2A2) showed hyperintense regions at the tip (in GRE-only) as well as at the MRT coils. MRT SNR supported 16 fps navigation with robust (<1mm, <5o degree deviation) visualization of the tip location and orientation. Imaging of
water-filling in phantom: MR-guided injection (Figure 2B) displayed the water-filling
direction and volume over time. Swine imaging of Gd-DTPA-filled pocket: MR-guided navigation, followed by T2-w monitoring (Figure 3), was used to separate the
vaginal wall and rectum, visualizing the pocket topology.
Conclusions: The MR-Tracked metallic injection needle
allowed rapid navigation to the desired targets, as well as accurate real-time assessment
of the injected topology.Acknowledgements
NIH R01CA237005,
R01HL094610References
[1] Wang W, Dumoulin
CL, Viswanathan AN, Tse ZTH, Mehrtash A, Loew W, Norton I, Tokuda J,
Seethamraju RT, Kapur T, Damato AL, Cormack RA, Schmidt EJ. Real-time active
MR-tracking of metallic stylets in MR-guided radiation therapy. Mag Reson Med.
2015; 73: 1803-11.
[2] Alipour A, Meyer ES, Tau S, Guttman M, Kolandaivelu A, Dumoulin
C, Watkins R, Elahi H, Loew W, Schweitzer J, Olson G, Chen Y, Halperin HR,
Schmidt EJ. MRI Conditional Actively Tracked
Metallic Electrophysiology Catheters and Guidewires with Miniature Tethered
Radio-Frequency Traps: Theory, Design and Validation. IEEE transactions on
Biomedical engineering 2020;67(6),1616-1627.
[3] de Arcos J, Schmidt EJ, Wang W, Tokuda J, Vij K, Seethamraju
RT, Damato AL, Dumoulin CL, Cormack RA, Viswanathan AN. Prospective Clinical
Implementation of a Novel Magnetic Resonance Tracking Device for Real-Time
Brachytherapy Catheter Positioning. International journal of radiation
oncology, biology, physics 2017;99(3):618-626.
[4] Hatiboglu G, Pinkawa M, Vall J-P, Hadaschik B, Hohenfellner
M. Application technique: placement of a
prostate – rectum spacer in men undergoing prostate radiation therapy, BJUI
2012;110: E647-E652
[5] Damato
AL, Kassick M, Viswanathan AN.
Rectum and
bladder spacing in cervical cancer brachytherapy using a novel injectable
hydrogel compound (TraceIT).
Brachytherapy. 2017;16(5):949-955.
[6] Marnitz
S, Budach V, Weisser F, Burova E, Gebauer B, Vercellino FG, Köhler C. Rectum
separation in patients with cervical cancer for treatment planning in primary
chemo-radiation. Radiat Oncol. 2012;7: 109-116.